Skip to content

MOUNTAINEER 188319422Recalls & reported problems

Medos International Sàrl · MOUNTAINEER OCT SPINAL SYSTEM MINI POLYAXIAL BONE SCREW 3.5 x 4 x 22mm · Class 2 · Implant

All models of MOUNTAINEER (51)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034157887), received between May 3, 2018 and May 3, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Consequences Or Impact To Patient1
  • Not Applicable1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 3, 2018Break; Break

    ON (B)(6) 2018, THE SURGERY FOR UNKNOWN REASONS WAS PERFORMED USING THE MOUNTAINEER SPINE SYSTEM. THE FIXED AREA WAS C7 (RIGHT). DURING THE SURGERY, THE FOLLOWING EVENT WAS CONFIRMED BY THE SURGEON. HE INSERTED THE REPORTED SCREW ((B)(4)) TO C7 (RIGHT) AFTER ATTACHING THIS SCREW TO THE SCREWDRIVER. HE TEMPORARILY FELT DIFFICULTIES PULLING OUT THE SCREWDRIVER, BUT HE COULD DO IT. WHEN PLACING THE ROD, HE NOTICED THAT THE HEAD OF THE SCREW WAS BROKEN. HE EXTRACTED THE BROKEN SCREW AND REPLACED IT WITH A NEW ONE. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH A LESS-THAN-30-MINUTE DELAY, AND THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient; Not Applicable

    MAUDE report 7482843 ↗

Other Orthosis, cervical pedicle screw spinal fixation devices

All Orthosis, cervical pedicle screw spinal fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page