Skip to content

MOUNTAINEER 188319525Recalls & reported problems

Medos International Sàrl · MOUNTAINEER OCT SPINAL SYSTEM MINI POLYAXIAL BONE SCREW 3.5 x 4.35 x 25mm · Class 2 · Implant

All models of MOUNTAINEER (51)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034158044), received between Sep 1, 2017 and Sep 1, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 1, 2017Break; Break

    ON (B)(6) 2017, SURGERY FOR SPONDYLOLISTHESIS WAS PERFORMED USING THE MOUNTAINEER SYSTEM. THE FIXED AREA WAS C7 ¿ T1. THE FOLLOWING ISSUES ARE CONFIRMED DURING THE SURGERY. - THE SURGEON MADE THE HOLE BY USING 4.0 MM TAP. - HE TRIED TO INSERT THE 4.35 MM SCREW USING THE SCREW DRIVER (PART#: 288304000), HOWEVER, TIP OF THE SCREW DRIVER GOT BROKEN AND THEN THE BROKEN TIP WAS BURIED INSIDE THE SCREW. - HE ATTACHED THE SCREW TO THE ROD AND SET SCREW TOGETHER, AND THEN PULLED THE SCREW OUT. - EVENTUALLY THE SCREW GOT BROKEN. THERE WERE NO BROKEN PARTS FOUND LEFT IN THE BODY AND THE SURGERY WAS SUCCESSFULLY COMPLETED WITH A 30-MINUTE DELAY. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6839766 ↗

Other Orthosis, cervical pedicle screw spinal fixation devices

All Orthosis, cervical pedicle screw spinal fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page