MOUNTAINEER 188320336Recalls & reported problems

Medos International Sàrl · MOUNTAINEER OCT SPINAL SYSTEM MINI POLYAXIAL BONE SCREW 3.5 x 3.5 x 36mm · Class 2 · Implant

All models of MOUNTAINEER (51)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034158167), received between Mar 19, 2018 and Mar 19, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Torn Material” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Torn Material1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 19, 2018Torn Material; Torn Material

    I WAS COVERING A SPINE CASE AT (B)(6) HOSPITAL ((B)(4)) WITH DR. (B)(6). WE WERE PERFORMING A POSTERIOR SPINE FUSION OF C1-C2 USING MOUNTAINEER. THE CASE WENT WELL UNTIL THE SURGEON WENT TO USE THE FINAL TIGHTENER ON THE SET SCREW (1883-41-200) WHILE TIGHTENING, THE SET SCREW BEGAN TO SPIN FREELY INSIDE THE PILYAXIAL HEAD OF THE LONG SHANK SCREW 3.5MM X 36MM (1886-20-336). THE SURGEON ATTEMPTED TO USE A SEPARATE SET SCREW AND REALIZED THE POLYAXIAL HEAD MUST HAVE STRIPPED. AT THIS POINT WE DECIDED TO BACK OUT THE SCREW AND REPLACE IT WITH THE SAME ITEM. ONCE THIS WAS COMPLETED A NEW SET SCREW THREADED PERFECTLY AND COMPLETED THE CONSTRUCT. I SAVED THE STRIPPED SCREW AND DAMAGED SET SCREWS. PATIENT CONSEQUENCE? :NO IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? : YES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7350116 ↗

Other Orthosis, cervical pedicle screw spinal fixation devices

All Orthosis, cervical pedicle screw spinal fixation devices →

Sources and updates