MOSS 274556000Recalls & reported problems
Depuy Spine, LLC · MOSS MIAMI SPINE SYSTEM X-RAY MARKER · Class 1
All models of MOSS (16)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034192444), received between Jul 1, 2025 and Jul 1, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING THE PROCEDURE WHEN USING PEDICLE MARKERS, WHEN REMOVING THEM FROM THE PATIENT, THE SPECIALIST RETURNED THEM BENT. FOR WHAT HAPPENED THERE WERE NO DISCOMFORT ON THE PART OF THE SPECIALIST SINCE BEFORE THEY WERE BENT HE COULD USE THEM WITHOUT ANY PROBLEM. THERE WERE NO AFFECTIONS TO THE PATIENT, NOR ALTERATIONS IN THE SURGICAL TIMES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22370677 ↗MalfunctionJul 1, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING THE PROCEDURE WHEN USING PEDICLE MARKERS, WHEN REMOVING THEM FROM THE PATIENT, THE SPECIALIST RETURNED THEM BENT. FOR WHAT HAPPENED THERE WERE NO DISCOMFORT ON THE PART OF THE SPECIALIST SINCE BEFORE THEY WERE BENT HE COULD USE THEM WITHOUT ANY PROBLEM. THERE WERE NO AFFECTIONS TO THE PATIENT, NOR ALTERATIONS IN THE SURGICAL TIMES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22370675 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page