MOSS 274579000Recalls & reported problems
Depuy Spine, LLC · MOSS MIAMI SPINE SYSTEM ROD CUTTER · Class 1
All models of MOSS (16)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034192475), received between Mar 26, 2020 and May 25, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Corroded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Corroded1
- Detachment of Device or Device Component1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 25, 2022Corroded; Material Deformation; Corroded; Material Deformation
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED BY THE SALES REP IN JAPAN THAT DURING A PSF (TH5-TH12) FOR TREATING THORACIC OPLL ON (B)(6) 2022, A ROD CUTTER DEVICE WAS USED TO TRY TO CUT A COBALT CHROMIUM ROD BUT THE ROD GOT STUCK IN THE ROD CUTTER AND BECAME IMMOVABLE. ACCORDING TO THE REPORT, THE ROD WAS REMOVED FROM THE ROD CUTTER, AND IT WAS CONFIRMED THAT THE CUTTING LINE ON THE ROD CUTTER HAD SOME BURRS. IT WAS FURTHER REPORTED THAT THE BLADE ON THE ROD CUTTER HAD TURNED TO RUST-LIKE COLOR. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH LESS THAN 30-MINUTE SURGICAL DELAY. THIS COMPLAINT INVOLVES ONE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14500989 ↗MalfunctionMar 26, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT THE PATIENT HAD A SURGERY FOR PLF (POSTERIOR LUMBAR FUSION) ON (B)(6) 2020, AND POSTOPERATIVELY THE ROD CUTTER WAS CLEANSED. THEY FOUND THAT THE SCREW, WHICH CONNECTS THE HANDLE AND THE UNIT, HAD COME OFF. THERE WAS NO SURGICAL DELAY REPORTED. THE PATIENT OUTCOME WAS STABLE. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9884500 ↗
Other INSTRUMENT, CUTTING, ORTHOPEDIC devices
- SIGMA LCS 9505-01-230Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC95 reports
- ATTUNE 2544-00-522Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC58 reports
- SIGMA LCS 9505-01-229Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC56 reports
- SIGMA 2024-40-100Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC43 reports
- ATTUNE 2545-00-156Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC41 reports
- ATTUNE 2545-00-158Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC41 reports
- SIGMA 2024-40-200Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC40 reports
- ATTUNE 2545-00-155Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC39 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page