Skip to content

MONARCH 277040700Recalls & reported problems

Depuy Spine, LLC · MONARCH SPINE SYSTEM CAP INSERTER · Class 1

All models of MONARCH (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034196879), received between Aug 13, 2018 and Feb 16, 2023 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break4

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 16, 2023Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN ITALY AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2023, THE DEVICES WERE BROKEN. THERE WAS NO PATIENT CONSEQUENCE. NO BROKEN FRAGMENTS WERE RETAINED IN THE PATIENT. THIS REPORT INVOLVES ONE MONARCH SPINE SYSTEM CAP INSERTER. THIS IS REPORT 1 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16384836 ↗
  • MalfunctionFeb 6, 2023Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2023 DURING A SPINE PROCEDURE TYPHOON TIP SNAPPED OFF WHILE REMOVING SCREW. SURGERY WAS COMPLETED WITHOUT ANY SURGICAL DELAY. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) CAP INSERTER THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16305503 ↗
  • MalfunctionNov 15, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2021, THE CAP INSERTER WAS BROKEN DURING REVERSE LOGISTICS AUDIT OF RETURNED DEVICE AT A THIRD PARTY SERVICER. THERE WAS NO PATIENT INVOLVEMENT AND NO ADDITIONAL INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) CAP INSERTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12814911 ↗
  • MalfunctionAug 13, 2018Break; Break

    REMOVAL OF THE MONARCH SYSTEM WAS REPORTED: DR (B)(6), (B)(6) HOSPITAL (B)(6) 2018. PRIMARY IMPLANT DATE 2011. REVISION: REMOVING IMPLANTS FOR SCOLIOSIS CORRECTION. BOTH INSTRUMENTS BROKE AT THE DISTAL TIPS ON TRYING TO REMOVE THE TYPHOON SET CAP - THIS WAS DUE TO BONE FORMATION AROUND THE SCREW HEAD. NO ADVERSE AFFECT TO THE PATIENT AND NO TIME WASTED AS WE COULD USE OTHER ONES ON THE SET AND REMOVED THE BONE PRIOR TO THE THIRD AND FORTH ATTEMPT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7776709 ↗

FDA clearance and classification

The FDA regulates product code HWR (driver, prosthesis) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the MONARCH 277040700

Has the MONARCH 277040700 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034196879), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the MONARCH 277040700?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Aug 13, 2018 and Feb 16, 2023. In those reports, the most common specific device problem is “Break” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the MONARCH 277040700?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the MONARCH 277040700 FDA approved?

Devices of this type (product code HWR) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy Spine, LLC.

Who makes the MONARCH 277040700, and is it still sold?

The FDA's device database lists Depuy Spine, LLC as the company responsible for it (brand MONARCH, model 277040700). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other DRIVER, PROSTHESIS devices

All DRIVER, PROSTHESIS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page