EXPEDIUM 279703140Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM ELLIPTICAL HANDLE - RATCHETING ADAPTER · Class 1
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10705034198743), received between Jul 20, 2023 and Jun 9, 2026 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Break” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break15
- Mechanical Jam11
- Device-Device Incompatibility2
Patient problems
- No Clinical Signs, Symptoms or Conditions17
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2026Break
IT WAS REPORTED THAT RATCHETING PORTION OF DRIVER WAS BROKEN. THERE WAS NO SURGERY DELAYED DUE TO THE REPORTED EVENT. PROCEDURE SUCCESSFULLY COMPLETED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25444877 ↗MalfunctionJun 9, 2026Break
IT WAS REPORTED THAT RATCHETING PORTION OF DRIVER WAS BROKEN. THERE WAS NO SURGERY DELAYED DUE TO THE REPORTED EVENT. PROCEDURE SUCCESSFULLY COMPLETED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25444876 ↗MalfunctionAug 4, 2025Break; Break
IT WAS REPORTED THAT DURING A PROCEDURE; THE RATCHETING HANDLE WAS BROKEN' WON¿T RATCHET. THERE WAS NO SURGERY DELAY DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22708068 ↗MalfunctionDec 13, 2024Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE INSTRUMENT BROKE DURING SURGERY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCES. THERE WAS NO GENERATION OF FRAGMENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20926071 ↗MalfunctionSep 17, 2024Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING INSPECTION IT WAS NOTICED THAT THE RATCHING MECHANISM WAS BROKEN, COULD NOT TURN TO REVERSE. THERE WERE NO BROKEN FRAGMENTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237939 ↗MalfunctionSep 17, 2024Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING INSPECTION IT WAS NOTICED THAT THE RATCHING MECHANISM WAS BROKEN, COULD NOT TURN TO REVERSE. THERE WERE NO BROKEN FRAGMENTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237914 ↗MalfunctionSep 17, 2024Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING INSPECTION IT WAS NOTICED THAT THE RATCHING MECHANISM WAS BROKEN, COULD NOT TURN TO REVERSE. THERE WERE NO BROKEN FRAGMENTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237901 ↗MalfunctionSep 17, 2024Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING INSPECTION IT WAS NOTICED THAT THE RATCHING MECHANISM WAS BROKEN, COULD NOT TURN TO REVERSE. THERE WERE NO BROKEN FRAGMENTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237900 ↗MalfunctionSep 17, 2024Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING INSPECTION IT WAS NOTICED THAT THE RATCHING MECHANISM WAS BROKEN, COULD NOT TURN TO REVERSE. THERE WERE NO BROKEN FRAGMENTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237895 ↗MalfunctionSep 17, 2024Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING INSPECTION IT WAS NOTICED THAT THE RATCHING MECHANISM WAS BROKEN, COULD NOT TURN TO REVERSE. THERE WERE NO BROKEN FRAGMENTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237893 ↗MalfunctionSep 17, 2024Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING INSPECTION IT WAS NOTICED THAT THE RATCHING MECHANISM WAS BROKEN, COULD NOT TURN TO REVERSE. THERE WERE NO BROKEN FRAGMENTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237879 ↗MalfunctionSep 17, 2024Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING INSPECTION IT WAS NOTICED THAT THE RATCHING MECHANISM WAS BROKEN, COULD NOT TURN TO REVERSE. THERE WERE NO BROKEN FRAGMENTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237791 ↗MalfunctionSep 17, 2024Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING INSPECTION IT WAS NOTICED THAT THE RATCHING MECHANISM WAS BROKEN, COULD NOT TURN TO REVERSE. THERE WERE NO BROKEN FRAGMENTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237759 ↗MalfunctionSep 17, 2024Break; Mechanical Jam; Break; Mechanical Jam
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING INSPECTION IT WAS NOTICED THAT THE RATCHING MECHANISM WAS BROKEN, COULD NOT TURN TO REVERSE. THERE WERE NO BROKEN FRAGMENTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237627 ↗MalfunctionApr 4, 2024Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam
IT WAS REPORTED THAT ON (B)(6) 2024 THE HANDLE WAS PULLED FROM STOCK TO USE AS A "MAKE FROM" PART FOR A SPECIAL REQUEST TEAM DEVICE AND THE BUTTON WAS STICKING AND UNUSABLE. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) EXPEDIUM SPINE SYSTEM ELLIPTICAL HANDLE RATCHETING ADAPTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19039949 ↗MalfunctionFeb 12, 2024Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT ONE OF THE 1085 BASE SIEVES IS A DEFECTIVE HANDLE 279703140. THE HANDLE PART TURNS OFF FROM THE RATCHET PART WHEN YOU SCREW IT AROUND TO THE LEFT. NO PATIENT HARM. IT WAS NOTICED INTRAOPERATIVELY. THIS REPORT IS FOR ONE (1) ELLIPTICAL RATCHETING HANDLE THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18691312 ↗MalfunctionJul 20, 2023Break; Break
IT WAS REPORTED THAT DURING A POSTERIOR SPINAL FUSION PROCEDURE ON (B)(6) 2023, THE RATCHETING HANDLE WAS BROKEN. IT WAS INSUFFICIENT TO PLACE SCREWS. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY DELAY, AND NO FRAGMENTS WERE RETAINED IN THE PATIENT. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM ELLIPTICAL HANDLE - RATCHETING ADAPTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17360299 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page