EXPEDIUM TOP NOTCH 279712525Recalls & reported problems

Medos International Sàrl · EXPEDIUM SPINE SYSTEM SQUEEZE-HANDLE ROD APPROXIMATOR FOR TOP NOTCH 5.5 · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034198972), received between Jun 30, 2021 and Aug 21, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Mechanical Jam2
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Unspecified Tissue Injury1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 21, 2024Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024. WHEN REDUCING THE ROD INTO THE SET SCREW, A REDUCTION TOOL WAS REQUIRED. THE KERRISON STYLE ONE FROM THE EXPEDIUM 5.5 CORE INSTRUMENT SET WAS USED. THIS BECAME JAMMED ON THE SCREW TULIP AND DID NOT PERFORM IT¿S FUNCTION AS A ROD REDUCER. IT WAS EXTREMELY DIFFICULT TO REMOVE FROM THE TULIP AND IN DOING SO THE PATIENTS SKIN WAS CAUGHT UP IN THE STRUGGLE TO REMOVE AND WAS SLIGHTLY CUT. THE DEVICE WAS EVENTUALLY REMOVED, AND THE CASE WAS COMPLETED WITHOUT FURTHER COMPLICATION. THIS REPORT IS FOR A ROD APPROX W/DOVETAIL. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Unspecified Tissue Injury

    MAUDE report 20012629 ↗
  • MalfunctionMar 23, 2023Mechanical Jam; Mechanical Jam

    DEVICE REPORT FROM DEPUY REPORTS AN EVENT IN SINGAPORE AS FOLLOWS: IT WAS REPORTED THAT ROD APPROXIMATOR FROM EXPEDIUM SET WAS JAMMED ON THE SCREWHEAD AND COULD NOT BE REMOVED. SURGEON HAD TO BREAK IT IN ORDER TO RELEASE IT. THERE WAS SURGICAL DELAY OF 120 MINUTES DUE TO THE REPORTED EVENT. THE SURGERY WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR ONE (1) ROD APPROX W/DOVETAIL. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16600917 ↗
  • MalfunctionFeb 15, 2023Break; Break

    DEVICE REPORT FROM SYNTHESE REPORTS AN EVENT IN ITALY AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, IT WAS DISCOVERED PREOPERATIVELY THAT THE DEVICES WERE BROKEN. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM SQUEEZE-HANDLE ROD APPROXIMATOR FOR TOP NOTCH 5.5. THIS IS REPORT 13 OF 13 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16378165 ↗
  • MalfunctionJun 30, 2021Device-Device Incompatibility; Device-Device Incompatibility

    THIS REPORT IS 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12095559 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices