CONFIDENCE SPINAL CEMENT SYSTEM 283902611Recalls & reported problems
Medos International Sàrl · CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, BEVELED TIP 11G x 6 Inches · Class 1
All models of CONFIDENCE SPINAL CEMENT SYSTEM (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034209432), received between Oct 4, 2018 and Dec 31, 2024 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Adverse Event Without Identified Device or Use Problem2
- Entrapment of Device1
- Material Deformation1
Patient problems
- Foreign Body In Patient3
- Unspecified Infection2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryDec 31, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT UNDERWENT SPINAL FUSION ON (B)(4) 2024. OVER THE FOLLOWING WEEKEND, THE PATIENT PRESENTED WITH AN INFECTION. PATIENT HAD BEEN IMPLANTED WITH TRIALTIS THORACOLUMBAR SYSTEM. IN ADDITION TO IMPLANTING TRIALTIS, THE SURGEON USED CONFIDENCE BONE CEMENT.
Patient problems: Unspecified Infection
MAUDE report 21048365 ↗InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT UNDERWENT SPINAL FUSION ON (B)(6) 2024. OVER THE FOLLOWING WEEKEND, THE PATIENT PRESENTED WITH AN INFECTION. PATIENT HAD BEEN IMPLANTED WITH TRIALTIS THORACOLUMBAR SYSTEM. IN ADDITION TO IMPLANTING TRIALTIS, THE SURGEON USED CONFIDENCE BONE CEMENT.
Patient problems: Unspecified Infection
MAUDE report 21035837 ↗MalfunctionAug 22, 2024Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT DURING A REVISION SURGERY ON (B)(6) 2024, A CEMENTED SCREW WAS REMOVED AND THE TROCAR WAS USED TO INSERT ADDITIONAL CEMENT. THE TROCAR WAS LEFT INSIDE THE CEMENT TOO LONG AND WHEN THE SURGEON TRIED TO REMOVE IT, ABOUT 4 CM OF THE TROCAR REMAINED IN THE VERTEBRAL BODY, SURROUNDED BY CEMENT AND WITH CEMENT INSIDE. THE SURGERY WAS DELAYED BY 30 MINUTES DUE TO THE EVENT. THE PATIENT OUTCOME WAS REPORTED TO BE GOOD, WITHOUT COMPLICATIONS. IT WAS REPORTED THAT THE REVISION WAS DUE TO INFECTION OF THE PREVIOUS AUGMENTED SCREWS SYSTEM. A SCREW REMOVAL WAS PLANNED AND CEMENT WAS INSERTED INSTEAD OF SCREWS. THIS REPORT INVOLVES ONE CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, BEVELED TIP 11G X 6 INCHES THIS IS REPORT 1 OF 1 FOR (B)(4). THIS REPORT IS RELATED TO (B)(4), WHICH CAPTURES THE REPORTED INFECTION.
Patient problems: Foreign Body In Patient
MAUDE report 20031397 ↗MalfunctionMar 30, 2023Material Deformation; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A POSTERIOR SPINAL FUSION PERFORMED ON (B)(6) 2022. THE BONE QUALITY WAS VERY HARD, AND THE NEEDLES WERE FORCEFULLY INSERTED. WHEN THE NEEDLES WERE REMOVED, SIGNIFICANT FORCE WAS REQUIRED TO REMOVE THE NEEDLES. THE NEEDLES WERE REMOVED BY USING PLIERS AND HAMMERS. ONE WAS DISASSEMBLED AND HAD A T-HANDLE THAT CAME OFF. ONE WAS DEFORMED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) CONF INTRO NDLE, BEV, 11G 6". THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16645384 ↗InjuryOct 4, 2018Break; Break
IT WAS REPORTED THAT ON (B)(6) 2018 THE SURGEON WAS USING (B)(4) AND THE SURGEON TWISTED IT AND IT BECAME APPARENT THERE WAS A PROBLEM. THE NEEDLE THEN BROKE INSIDE THE PATIENT AND MY UNDERSTANDING IS THAT THE NEEDLE HAD TO BE LEFT THERE. THERE WERE TWO NEEDLES INVOLVED WITH THE SAME PRODUCT CODE BUT THEY HAD DIFFERENT LOT NUMBERS ¿ HTHBFK AND 184350.
Patient problems: Foreign Body In Patient
MAUDE report 7935871 ↗InjuryOct 4, 2018Break; Break
IT WAS REPORTED THAT ON (B)(6) 2018 THE SURGEON WAS USING (B)(4) AND THE SURGEON TWISTED IT AND IT BECAME APPARENT THERE WAS A PROBLEM. THE NEEDLE THEN BROKE INSIDE THE PATIENT AND MY UNDERSTANDING IS THAT THE NEEDLE HAD TO BE LEFT THERE. THERE WERE TWO NEEDLES INVOLVED WITH THE SAME PRODUCT CODE BUT THEY HAD DIFFERENT LOT NUMBERS ¿ HTHBFK AND 184350.
Patient problems: Foreign Body In Patient
MAUDE report 7935839 ↗
Other NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE devices
- N/A 305211Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE425 reports
- N/A 305180Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE300 reports
- N/A 305200Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE99 reports
- PneumoDart TM-317TyTek Medical Inc. · Needle, Aspiration And Injection, Disposable1 recall1 report
- TPAK TM-303TyTek Medical Inc. · Needle, Aspiration And Injection, Disposable1 recall0 reports
- N/A 305181Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE32 reports
- SOL-M™ Blunt Fill Needle with 5 Micron Filter 110022Sol-Millennium Medical, Inc. · Needle, Aspiration And Injection, Disposable17 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283903611Medos International Sàrl · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page