CONFIDENCE SPINAL CEMENT SYSTEM 283902613Recalls & reported problems
Medos International Sàrl · CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, BEVELED TIP 13G x 6 Inches · Class 1
All models of CONFIDENCE SPINAL CEMENT SYSTEM (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034209449), received between Mar 11, 2019 and Aug 25, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- Injury1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 25, 2020Break; Break
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 DURING A VERTEBROPLASTY OF A METASTASIZED VERTEBRAL BODY IN A (B)(6) YEAR-OLD TUMOR PATIENT, THE HANDLE OF THE CONFIDENCE NEEDLE BROKE WHEN TRYING TO REMOVE IT AFTER THE CEMENT APPLICATION. THE SHAFT GOT STUCK IN THE PATIENT AND COULD ONLY BE REMOVED WITH THE HELP OF FORCEPS, THE ASSISTANCE OF AN ACCIDENT SURGEON, CONSIDERABLE ADDITIONAL EXPENDITURE OF TIME AND RADIATION EXPOSURE FOR THE PATIENT. THERE WAS A SURGICAL DELAY OF FORTY-FIVE (45) MINUTES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE (1) CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, BEVELED TIP 13G X 6 INCHES. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Injury; No Code Available
MAUDE report 10448117 ↗MalfunctionMar 11, 2019Break; Break
IT WAS REPORTED THAT THE END OF NEEDLE TORE DURING USE. NO PATIENT IMPACT. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? --> NO, ACTION TAKEN WHEN EVENT OCCURRED? --> NEW NEEDLE USED, WAS PROCEDURE SUCCESSFULLY COMPLETED? --> YES, WERE FRAGMENTS GENERATED? --> NO, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? --> UNKNOWN, PATIENT STATUS/ OUTCOME / CONSEQUENCES --> NO, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: --> UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY --> UNKNOWN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8408874 ↗
Other NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE devices
- N/A 305211Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE425 reports
- N/A 305180Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE300 reports
- N/A 305200Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE99 reports
- PneumoDart TM-317TyTek Medical Inc. · Needle, Aspiration And Injection, Disposable1 recall1 report
- TPAK TM-303TyTek Medical Inc. · Needle, Aspiration And Injection, Disposable1 recall0 reports
- N/A 305181Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE32 reports
- SOL-M™ Blunt Fill Needle with 5 Micron Filter 110022Sol-Millennium Medical, Inc. · Needle, Aspiration And Injection, Disposable17 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283903611Medos International Sàrl · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page