EXPEDIUM 286200240Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM TORQUE WRENCH HANDLE · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10705034210995), received between Jun 11, 2019 and Sep 5, 2024 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Calibration Problem” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Calibration Problem8
- Material Twisted/Bent1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient1
- Not Applicable1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON AUG 9, 2024, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED THAT ITEM WAS FOUND TO BE OUT OF DRAWING SPECIFICATION, THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. ALL INFORMATION REGARDING THIS EVENT HAS BEEN REPORTED. THIS REPORT IS FOR ONE (1) TORQUE WRENCH HANDLE FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20150762 ↗MalfunctionSep 5, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON (B)(6) 2024 DURING ROUTINE INCOMING INSPECTION OF A LOANER SET AT THE HOSPITAL, IT WAS OBSERVED THAT 2862-00-240, 7 NM T-HANDLE, WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE DRAWING SPECIFICATION TORQUE TESTED LOW. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THE PRODUCT HAS BEEN QUARANTINED AND IS AVAILABLE AND IS BEING RETURNED TO THE REPAIR SITE. ALL INFORMATION REGARDING THIS EVENT HAS BEEN REPORTED. THIS REPORT IS FOR ONE (1) TORQUE WRENCH HANDLE FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20145993 ↗MalfunctionJul 11, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2024, THE TORQUE TEST FAILED DURING INSPECTION AT THE LOCAL DISTRIBUTION CENTER. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM TORQUE WRENCH HANDLE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19719919 ↗MalfunctionNov 3, 2023Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING ROUTINE INSPECTION OF THE LOANER SET, IT WAS IDENTIFIED THAT THE TORQUE WRENCH HANDLE WAS OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED LOW. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM TORQUE WRENCH HANDLE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18067087 ↗MalfunctionMay 24, 2023Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON (B)(6) 2023 DURING ROUTINE INCOMING INSPECTION OF A LOANER SET WAS OBSERVED THAT MATERIAL TORQUE WRENCH HANDLE WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THE PRODUCT HAS BEEN QUARANTINED AND IS AVAILABLE AND IS BEING RETURNED TO THE REPAIR SITE. A REPLACEMENT DEVICE HAS BEEN ORDERED. NO REPLACEMENT PRODUCT IS REQUIRED, AND NO SHIPPING LABEL IS REQUIRED FOR THE PRODUCT RETURN. ALL INFORMATION REGARDING THIS EVENT HAS BEEN REPORTED. THIS REPORT IS FOR ONE (1) TORQUE WRENCH HANDLE THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16991613 ↗MalfunctionApr 26, 2021Calibration Problem; Material Twisted/Bent; Calibration Problem; Material Twisted/Bent
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT UNDERWENT FOR A REVISION SURGERY. DURING THE REVISION SURGERY, THE SURGEON WAS DOING AN EXPEDIUM 4.5 GROWING ROD REVISION. IT WAS NOTICED THAT HANDLES HAVE NOT BEEN CHANGED OVER 4 YEARS. THE SURGEON SUGGESTED TO CHANGE SFX TORQUE AND TORQUE WRENCH HANDLE. THEY¿RE NEED TO RECALIBRATE. A FEW SCREWDRIVERS HAVE BEEN STRIPPED IN LAST COUPLE OF MONTHS. THERE WERE NO PATIENT CONSEQUENCES. CONCOMITANT DEVICE REPORTED: UNKNOWN ROD (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR (1) TORQUE WRENCH HANDLE. THIS REPORT IS 1 OF 1 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11725553 ↗MalfunctionDec 22, 2020Calibration Problem; Calibration Problem
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED AN UNKNOWN DATE THAT THE ITEM WAS FOUND DAMAGED DURING REGULAR LOANER INSPECTION PROCESS FROM TORQUE TESTING PROCEDURE. THERE WAS NO PROCEDURE AND PATIENT INVOLVEMENT ARE REPORTED. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1)TORQUE WRENCH HANDLE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11057322 ↗MalfunctionDec 22, 2020Calibration Problem; Calibration Problem
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED AN UNKNOWN DATE THAT THE ITEM WAS FOUND DAMAGED DURING REGULAR LOANER INSPECTION PROCESS FROM TORQUE TESTING PROCEDURE. THERE WAS NO PROCEDURE AND PATIENT INVOLVEMENT ARE REPORTED. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) TORQUE WRENCH HANDLE THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11057243 ↗MalfunctionJun 11, 2019Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT ON AN (B)(6) 2019, THAT THE POINT OF TORQUE WRENCH SHAFT WAS SPINNING IN SET SCREW HEAD, THEY HAD TO CHANGE TO ALTERNATIVE INSTRUMENTS.. THERE WAS EVIDENCE OF WEAR AT THE HEX OF THE TORQUE SHAFT. IT IS UNKNOWN IF THERE WERE PROCEDURE AND PATIENT INVOLVEMENT. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREW (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES ONE (1) DEVICE.
Patient problems: No Consequences Or Impact To Patient; Not Applicable
MAUDE report 8689737 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page