VIPER 286705220Recalls & reported problems
Depuy Spine, LLC · VIPER SYSTEM GUIDEWIRE, BLUNT AND DISPOSABLE 1.45mm · Class 1
All models of VIPER (60)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10705034213682), received between Apr 2, 2018 and May 6, 2026 — classified as 3 injuries, 13 malfunctions.
The most frequently reported specific device problem is “Break” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break10
- Device-Device Incompatibility8
- Entrapment of Device5
Patient problems
- Foreign Body In Patient10
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information1
- No Code Available1
- Not Applicable1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 6, 2026Device-Device Incompatibility
IT WAS REPORTED THAT THIS WAS A POSTERIOR FIXATION PERFORMED ON AN UNKNOWN DATE. WHEN A PVP AND A POSTERIOR FIXATION WERE PERFORMED, THE STENT OF THE VBS WAS RETAINED IN THE BODY, AND THE GUIDE WIRE WAS INSERTED AFTER THE CEMENT WAS INJECTED. DURING THE SCREW INSERTION THROUGH THE GUIDE WIRE, THE CEMENT HAD ALREADY HARDENED OVER TIME, AND THE GUIDE WIRE COULD NOT BE REMOVED SMOOTHLY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES SURGICAL DELAY. THE SURGEON SAID THAT IT WAS NEEDED TO BE MORE CAREFUL IN THE FUTURE AND PROCEED OPERATION IN COOPERATION WITH ENGINEER. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25092906 ↗MalfunctionJan 13, 2026Device-Device Incompatibility
IT WAS REPORTED THAT DURING A VBS PERFORMED ON (B)(6) 2025, VIPERPRIME WAS USED FOR VBS AND POSTERIOR FIXATION. AFTER VBS WAS PERFORMED, THE GUIDE WIRE WAS INSERTED WHEN INSERTING A SCREW INTO THE SAME VERTEBRAL BODY. AFTER INSERTING A SCREW OVER THE GUIDE WIRE, THE CEMENT ADHERED WHEN THE GUIDE WIRE WAS REMOVED, MAKING IT DIFFICULT TO REMOVE. WHEN THE SURGEON TRIED TO REMOVE IT, THE VBS FILLED WITH CEMENT WAS BACKED OUT. THIS OCCURRED ON THE LEFT SIDE. THE RIGHT SIDE WAS COMPLETED WITHOUT INCIDENT. VBS WAS PERFORMED AGAIN DURING RECOVERY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES OF SURGICAL DELAY. UPON CHECKING, THERE WAS NO IMPACT TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24055641 ↗MalfunctionJul 2, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING THE SURGERY, THE SURGEON APPROACHED THE VERTEBRAL BODY WITH A PROBE AND PASSED A GUIDE WIRE THROUGH THE HOLLOW PART, BUT THEN HAD DIFFICULTY REMOVING THE PROBE FROM THE BODY. BECAUSE NO DEFORMATION WAS NOTED IN THE GUIDE WIRE OR PROBE, IT WAS SUSPECTED THAT THE DIAMETER OF THE GUIDE WIRE MAY HAVE BEEN LARGER THAN THE SIZE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22387257 ↗MalfunctionApr 1, 2025Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT PATIENT UNDERWENT POSTERIOR SPINAL FUSION ON (B)(6) 2025. DURING SCREW INSERTION, A PIECE OF THE GUIDE-WIRE BROKE UNDER THE SCREW AND REMAINED IN THE PATIENT. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO ADDITIONAL INTERVENTION. PATIENT OUTCOME WAS POSITIVE.
Patient problems: Foreign Body In Patient
MAUDE report 21741428 ↗MalfunctionDec 11, 2024Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT THIS WAS PVP AND PLF (L3) PERFORMED ON (B)(6) 2024. DURING THE SURGERY, AN UNKNOWN DEFECT OCCURRED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH 60 MINUTES DELAY. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 20903913 ↗MalfunctionOct 21, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE. THE DETAILS OF THE PRODUCTS USED IN THE INITIAL SURGERY ARE UNKNOWN. ON (B)(6) 2024, THE SECOND SURGERY WAS PERFORMED TO RE-FIX THE LUMBAR SPINE FOR DEGENERATIVE LUMBAR SPONDYLOLISTHESIS. IN THE SURGERY, A SCREW WAS INSERTED OVER THE GUIDE WIRE. THE SURGEON ATTEMPTED TO REMOVE THE GUIDE WIRE AFTER INSERTING THE SCREW BUT WAS UNABLE TO DO SO. THE SURGEON THEREFORE BACKED OUT THE SCREW SLIGHTLY AND ATTEMPTED TO REMOVE THE GUIDE WIRE AGAIN. IN DOING SO, THE TIP OF THE GUIDE WIRE BROKE OFF AND REMAINED INSIDE THE PATIENT. THE BROKEN PORTION OF THE GUIDE WIRE WAS THEN REMOVED FROM THE PATIENT'S BODY UNDER IMAGING, AND ULTIMATELY NO FRAGMENTS WERE FOUND IN THE PATIENT'S BODY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN THIRTY MINUTES SURGICAL DELAY. PATIENT WAS STABLE. THE SURGEON'S OPINION ON THE CAUSE OF THE BREAKAGE IS AS FOLLOWS. THE SCREW'S TRAJECTORY WAS MISALIGNED WITH THE TAPPED TRAJECTORY, AND AN ATTEMPT WAS MADE TO FORCIBLY REMOVE THE GUIDE WIRE WITH ITS DIRECTION BENT, WHICH LED TO THIS BREAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20493172 ↗MalfunctionMay 23, 2024Entrapment of Device; Device-Device Incompatibility; Entrapment of Device; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2024, DURING A VBS (TH12) FOR FRACTURE, FOR 1ABOVE1BELOW FUSION CASE, WHEN INSERTING A SCREW, THE GUIDEWIRE IN QUESTION WAS USED. THE CEMENT AND THE TIP OF THE GUIDEWIRE ADHERED, MAKING EXTRACTION DIFFICULT. THE PRIMARY SURGEON ATTEMPTED TO REMOVE THE CEMENT LUMP BY CUTTING IT OFF WITH THE TIP OF THE SCREW OR A TAP, BUT WAS UNABLE TO DO SO, SO THE 4 CM TIP OF THE GUIDEWIRE WAS CUT WITH A HOSPITAL INSTRUMENT AND THE RELEVANT PORTION WAS LEFT IN THE HOLLOW OF THE SCREW. THE CUT GUIDEWIRE WAS SANDWICHED BETWEEN THE VBS CEMENT MASS AND THE ROD AND WAS DETERMINED NOT TO MOVE WITHIN THE BODY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH 30 MINUTES SURGICAL DELAY. THE SURGEON COMMENTED THAT THE GUIDEWIRE WAS INSERTED INTO THE CEMENT MASS FOR A CERTAIN AMOUNT OF TIME, WHICH WAS THE CAUSE OF THIS EVENT. THIS REPORT INVOLVES ONE (1) VIPER SYSTEM GUIDEWIRE, BLUNT AND DISPOSABLE 1.45MM. THIS IS REPORT 1 OF 1 FOR (B)(4). THIS REPORT IS RELATED TO (B)(4), WHICH CAPTURES THE CEMENT INVOLVED IN THE INCIDENT.
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 19380459 ↗MalfunctionMar 12, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE STAFF COULD NOT THREAD THE GUIDE WIRE DOWN BACK OF TWO VIPER TAPS AS THERE WAS SOMETHING BLOCKING GUIDE WIRE FROM FUNNELING DOWN. THE SURGEON TRIED DIFFERENT GUIDEWIRES (X2) AND HANDLES AND THOSE WERE NOT THE ISSUE. THE GUIDE WIRES SNAPPED AFTER TRYING TO GET THEM THROUGH TAP. FINALLY THE THIRD TAP WAS TRIED AND WORKED WELL. THERE WAS NO SURGICAL DELAY AND NO PATIENT CONSEQUENCES/OUTCOME. THIS REPORT IS FOR ONE VIPER2 1.45 DIA GUIDEWIRE, BLU. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18885381 ↗MalfunctionMar 12, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE STAFF COULD NOT THREAD THE GUIDE WIRE DOWN BACK OF TWO VIPER TAPS AS THERE WAS SOMETHING BLOCKING GUIDE WIRE FROM FUNNELING DOWN. THE SURGEON TRIED DIFFERENT GUIDEWIRES (X2) AND HANDLES AND THOSE WERE NOT THE ISSUE. THE GUIDE WIRES SNAPPED AFTER TRYING TO GET THEM THROUGH TAP. FINALLY THE THIRD TAP WAS TRIED AND WORKED WELL. THERE WAS NO SURGICAL DELAY AND NO PATIENT CONSEQUENCES/OUTCOME. THIS REPORT IS FOR ONE (1) VIPER2 1.45 DIA GUIDEWIRE, BLU THIS IS REPORT 1 OF 4 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18885281 ↗InjuryMar 6, 2023Entrapment of Device; Device-Device Incompatibility; Entrapment of Device; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A POSTERIOR FUSION AND A PERCUTANEOUS VERTEBROPLASTY (TH12) FOR THE 12TH THORACIC COMPRESSION FRACTURE ON (B)(6) 2023. THE EVENT TOOK PLACE IN SEQUENCE AS FOLLOWS: THE EXPEDIUM VERSE FENESTRATED SCREWS WERE INSERTED INTO TH10, TH11, L1, AND L2. THEN, CEMENT AUGMENTATION WAS PERFORMED. AFTER CEMENT AUGMENTATION, A BKP WAS PERFORMED WITH SYNFLATE AT TH12. BOTH SIDE OF BALLOONS WERE INFLATED UP TO 4.0 ML. 3.7 CC OF THE CEMENT WAS INJECTED ON TH12 RIGHT, AND 3.8 CC OF THE CEMENT WAS INJECTED ON TH12 LEFT. AT THE END OF CEMENT INJECTION, 12 MINUTES HAVE ELAPSED SINCE THE START OF CEMENT MIXING. THE CEMENT NEEDLE AND THE SLEEVE WERE REMOVED. THE GUIDE WIRE WAS INJECTED, AND THEN THE SCREWS WERE IMMEDIATELY INJECTED WITHOUT TAP. ALTHOUGH THE SCREWS WERE INSERTED ON BOTH SIDES, RIGHT SIDE OF THE GUIDE WIRE COULD NOT BE REMOVED BECAUSE OF CEMENT HARDENING. AT THIS POINT, 15 MINUTES AND 30 SECONDS HAVE ELAPSED SINCE THE START OF CEMENT MIXING. THE SCREW WAS REMOVED BECAUSE THE GUIDE WIRE COULD NOT BE REMOVED. THE GUIDE WIRE WAS CUT BY THE PIN CUTTER. APPROXIMATELY 3 TO 4 CM OF THE GUIDE WIRE REMAIN
Patient problems: Foreign Body In Patient
MAUDE report 16484996 ↗InjuryFeb 17, 2023Break; Entrapment of Device; Break; Entrapment of Device
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023 A PPS (L4) FOR LUMBAR DEGENERATIVE DISEASE WAS BEING PERFORMED. THE GUIDE WIRE WAS INSERTED THROUGH J-PROBE, AND DEPTH WAS ADJUSTED USING PLIERS AND A HAMMER. WHEN THE GUIDE WIRE WAS REMOVED, THE TIP WAS BROKEN. THE IMAGE SHOWED THAT APPROXIMATELY 2 MM OF THE BROKEN TIP REMAINED IN THE VERTEBRAL BODY. THE SURGEON DETERMINED THAT IT WAS DIFFICULT TO REMOVE AND THERE WOULD BE NO IMMEDIATE IMPACT ON THE PATIENT. THEN, THE WOUND WAS CLOSED WITH THE BROKEN TIP LEFT INSIDE THE BODY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. THIS REPORT IS FOR A VIPER SYSTEM GUIDEWIRE, BLUNT AND DISPOSABLE 1.45MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 16397932 ↗InjuryAug 6, 2020Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT FOR AN UNKNOWN SURGERY. DURING THE SURGERY, WHEN THE SCREW INSERTION WAS CHECKED WITH THE C-ARM, THE CHIPPED GUIDE WIRE (ABOUT 5 MM LONG) WAS FOUND NEAR THE SCREW TIP. WHEN THE REMOVED GUIDE WIRE WAS CHECKED, IT WAS CONFIRMED THAT THE GUIDE WIRE'S TIP HAD BEEN BROKE AND LOST. THE SURGEON DECIDED TO LEAVE THE FRAGMENT IN THE PATIENT'S BODY AND COMPLETED THE PROCEDURE. THE PATIENT OUTCOME WAS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREW (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN ). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) VIPER2 1.45 DIA GUIDEWIRE, BLU. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 10376255 ↗MalfunctionOct 29, 2018Break; Break
IT WAS REPORTED DOCTOR WAS PERFORMING A MIS LUMBOPELVIC STABILIZATION PROCEDURE FOR A TRAUMATIC SACRAL FRACTURE. HE DID THIS ON (B)(6) 2018 AND USED VIPER PRIME FOR THE L5 SCREW AND A VIPER SAI FULLY THREADED SCREW FOR THE ILIUM. THE CONSTRUCT WAS A UNILATERAL PROCEDURE AND WAS ON THE RIGHT SIDE. DOCTOR USED "JAMSHIDI" NEEDLES TO CREATE THE TRACTS FOR HIS SCREWS. HE THEN INSERTED 286705220 GUIDEWIRES AND REMOVED THE JAMSHIDI NEEDLES. HE PLACED THE L5 VIPER PRIME SCREW WITHOUT INCIDENT. HE THEN TAPPED THE ILIAC SCREW AND THEN PLACED THE VIPER SAI SCREW. ONCE THE SCREW WAS PLACED WHERE HE WANTED, HE THEN ATTEMPTED TO REMOVE THE GUIDWIRE. THE GUIDEWIRE WOULD NOT MOVE. HE USED VARIOUS STYLES OF FORCEPS FROM THE ACCOUNT GENERAL SPINE INSTRUMENTS. NOTHING WOULD WORK. HE THEN ASKED FOR A BATTERY POWERED DRILL AND USED A WIRE DRIVER AND PLACED IT ON THE "STUCK" WIRE. HE THEN USED THE DRILL TO ROTATE THE GUIDEWIRE AND PULLED. AFTER 10-15 SECONDS, THE WIRE CAME OUT, BUT ABOUT 8MM OF THE DISTAL TIP BROKE OFF. IT APPEARED AS THOUGH THE BROKEN PART OF THE WIRE WAS IN THE CANNULATION OF THE SCREW. DOCTOR ELECTED TO LEAVE THE SCREW IN PLACE. HE THEN SUCCESSFULLY COMPLETED THE PROCEDURE. SURGICAL
Patient problems: Foreign Body In Patient
MAUDE report 8016034 ↗MalfunctionApr 16, 2018Break; Break
TIP OF THE GUIDING WIRE BROKE OFF (2CM APPROX.) INTRA-OP, DURING REMOVAL - IT REMAINED INSIDE VERTEBRAL BODY.
Patient problems: Foreign Body In Patient; No Code Available
MAUDE report 7432531 ↗MalfunctionApr 9, 2018Break; Break
SURGEON: MR (B)(6). HOSPITAL TOWN: (B)(6). PRODUCT NAME: VIPER 2 GUIDEWIRE. PRODUCT CODE : 286705220 WHEN INSERTING VIPER SAI SCREW (179704880). LOT/BATCH/EXP: NOT KNOWN. DID THE EVENT HAPPEN DURING A PROCEDURE? YES. WERE YOU IN THE PROCEDURE AT THE TIME OF THE EVENT? YES. WAS THE PRODUCT BEING USED IN A CLINICAL TRIAL? NO. EVENT OUTCOME/HOW WAS IT MANAGED? THE GUIDE WIRE HAD BEEN REMOVED. THE SURGEONS CHECKED THE X RAYS AND CONFIRMED NO ACTION WOULD BE TAKEN. WAS THERE A PATIENT IMPACT OR WAS THE PROCEDURE EXTENDED GREATER THAN 30 MINUTES DUE TO THE FAILURE? NO. HAS THE REPORTER FACILITY INDICATED THERE MAY BE LEGAL ACTION? NO. IS THE PRODUCT AVAILABLE FOR RETURN? NO DISPOSED OF IN SHARPS BIN. PLEASE GIVE A DETAILED EXPLANATION OF THE EVENT: UPON INSERTION OF THE SAI SCREW AND SUBSEQUENT REMOVAL OF THE GUIDE WIRE. THE TIP OF THE GUIDEWIRE (~5 MM) SNAPPED REMAINING IN THE PATIENTS BONE. WHEN FIRST REMOVED THE GUIDEWIRE APPEARED SLIGHTLY BENT FOLLOWING SCREW INSERTION, BUT THIS DIDN¿T CONCERN MR (B)(6). THE PROCEDURE CONTINUED (INSERTION OF SAI SCREWS AND 2 NEW RODS WITH UNIVERSAL CONNECTOR TO CONNECT TO EXISTING CONSTRUCT T12 SACRUM) AND UPON TAKING THE FINAL AP TO IMAGE IT WAS NOT
Patient problems: Foreign Body In Patient
MAUDE report 7409921 ↗MalfunctionApr 2, 2018Break; Break
ON (B)(6) 2014, THE SURGERY FOR KYPHOSCOLIOSIS WAS PERFORMED BY USING THE VIPER SYSTEM. IN ORDER TO INSERT THE SAI SCREW, THE GUIDE WIRE WAS INSERTED FROM THE SACRUM TO THE ILIAC. WHEN THE TAP WAS CUT OVER THE GUIDE WIRE, THE TIP OF THE GUIDE WIRE WAS BROKEN. A BROKEN PIECE WAS LEFT IN THE PATIENT. THE BROKEN PIECE WAS NOT REMOVABLE. NO FURTHER MEDICAL INTERVENTION IS PLANNED. THE SURGERY WAS COMPLETED WITH A DELAY (LESS THAN 30 MINUTES).
Patient problems: Foreign Body In Patient; Not Applicable
MAUDE report 7390273 ↗
FDA clearance and classification
The FDA regulates product code LXH (orthopedic manual surgical instrument) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VIPER 286705220
Has the VIPER 286705220 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705034213682), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the VIPER 286705220?
The FDA's MAUDE database lists 16 adverse event reports for this device, received between Apr 2, 2018 and May 6, 2026. In those reports, the most common specific device problems are “Break” (10), “Device-Device Incompatibility” (8) and “Entrapment of Device” (5); the most common patient problems are “Foreign Body In Patient” (10) and “Not Applicable” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VIPER 286705220?
None of the reports are classified as deaths, 3 as injuries and 13 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the VIPER 286705220 FDA approved?
Devices of this type (product code LXH) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy Spine, LLC.
Who makes the VIPER 286705220, and is it still sold?
The FDA's device database lists Depuy Spine, LLC as the company responsible for it (brand VIPER, model 286705220). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,989 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,199 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,163 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT852 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument821 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument559 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT492 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page