VIPER 286715300Recalls & reported problems

Depuy Spine, LLC · VIPER 2 SYSTEM STRAIGHT HANDLE, CANNULATED AND RATCHETING · Class 1

All models of VIPER (60)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705034214023), received between Feb 13, 2020 and Apr 20, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 4 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Device-Device Incompatibility5
  • Contamination /Decontamination Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 20, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2026, WHILE IN FORWARD POSITION THE HANDLE SLIPS. THERE WAS NO PATIENT CONSEQUENCES. THERE WAS NO SURGICAL DELAY AND SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24926365 ↗
  • MalfunctionFeb 12, 2025Break; Break

    IT WAS REPORTED THAT THE RATCHETING PORTION WAS BROKEN ON THE INSTRUMENTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21364244 ↗
  • MalfunctionFeb 12, 2025Break; Break

    IT WAS REPORTED THAT THE RATCHETING PORTION WAS BROKEN ON THE INSTRUMENTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21364243 ↗
  • MalfunctionFeb 12, 2025Break; Break

    IT WAS REPORTED THAT THE RATCHETING PORTION WAS BROKEN ON THE INSTRUMENTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21364242 ↗
  • MalfunctionFeb 12, 2025Break; Break

    IT WAS REPORTED THAT THE RATCHETING PORTION WAS BROKEN ON THE INSTRUMENTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21364241 ↗
  • MalfunctionDec 2, 2024Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION RECEIVED STATES THAT NO OTHER MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20824515 ↗
  • MalfunctionJul 2, 2024Break; Break

    IT WAS REPORTED ON JUNE 5, 2024, SALES REPRESENTATIVE RECEIVED SET BACK FROM STERILE PROCESSING (SPD) AND THE DRIVER HANDLE WAS IN PIECES. IT CAME APART WHILE BEING CLEANED. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE VIPER2 RATCHTING MODULAR STRAI. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19655685 ↗
  • MalfunctionFeb 29, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2024. THAT THE CUSTOMER RECEIVED A DIRTY ITEM IN AN ORTHOKIT LOANER SET: ACCORDING TO THE PICTURES, THERE IS CONSIDERABLE SOILING ON THE HANDLE (DARK PARTICLES ON A BLUE BACKGROUND). NO PATIENT INVOLVEMENT. THERE WAS NO SURGERY DELAY.  IT IS NOTICED IN THE STERILIZATION AREA BEFORE OP- AFTER DELIVERY. THIS REPORT IS FOR ONE (1) VIPER2 RATCHTING MODULAR STRAI. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18806641 ↗
  • MalfunctionAug 28, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON 14-AUGUST-2023, THE BUTTON ON THE HANDLE IN QUESTION COULD NOT BE PRESSED WHEN IT WAS ATTEMPTED TO ATTACH A SHAFT. THE ISSUE WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR, THERE WAS NO REPORTED PATIENT OR PROCEDURE INTERACTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A VIPER 2 SYSTEM STRAIGHT HANDLE, CANNULATED AND RATCHETING. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17632872 ↗
  • MalfunctionNov 4, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THE HANDLE IN QUESTION WAS USED IN AN UNKNOWN SURGERY ON (B)(6) 2022. ON (B)(6) 2022, IT WAS FOUND THAT THE BUTTON PART OF THE HANDLE HAD BEEN DISASSEMBLED DURING CLEANSING AFTER THE SURGERY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE VIPER 2 SYSTEM STRAIGHT HANDLE, CANNULATED AND RATCHETING. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15735293 ↗
  • MalfunctionApr 30, 2021Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR SPINAL FUSION. THE SCRUB NURSE PREOPERATIVELY TRIED TO CONNECT THE HEAD ADJUSTER TO A TEARDROP HANDLE (279702120), WHICH FAILED. ALSO, TRIED TO CONNECT THE HEAD ADJUSTER WITH ANOTHER COUNTERPART HANDLES SUCH AS 279702140, 286715300, AND 299704135, WHICH ALL WAS FAILED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A REPLACEMENT WITHOUT SURGICAL DELAY. THE PATIENT OUTCOME WAS STABLE. THIS COMPLAINT INVOLVES FIVE (5) DEVICES. THIS REPORT IS FOR (1) VIPER2 RATCHETING MODULAR STRAI. THIS REPORT IS 4 OF 5 (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11756026 ↗
  • MalfunctionFeb 13, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2019, THE SCREW SLIPPED OFF PEDICLE AND WAS INSERTED INTO L4/5 DISC SPACE. BRAINLAB JAM SHIIDI AND 1.45MM GUIDEWIRE WERE ALSO USED. THE V2 RATCHETING T HANDLE WAS UNABLE TO BE REMOVED AFTER THE PROCEDURE. SURGICAL DELAY OF THIRTY MINUTES WAS REPORTED. THIS REPORT IS FOR ONE (1) VIPER2 STRAIGHT RATCHETING HANDLE, CANNULATED. THIS IS REPORT 3 OF 4 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9706249 ↗

FDA clearance and classification

The FDA regulates product code LXH (orthopedic manual surgical instrument) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VIPER 286715300

Has the VIPER 286715300 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034214023), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VIPER 286715300?

The FDA's MAUDE database lists 12 adverse event reports for this device, received between Feb 13, 2020 and Apr 20, 2026. In those reports, the most common specific device problems are “Break” (6), “Device-Device Incompatibility” (5) and “Contamination /Decontamination Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VIPER 286715300?

None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the VIPER 286715300 FDA approved?

Devices of this type (product code LXH) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy Spine, LLC.

Who makes the VIPER 286715300, and is it still sold?

The FDA's device database lists Depuy Spine, LLC as the company responsible for it (brand VIPER, model 286715300). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.