VIPER 286715500Recalls & reported problems
Depuy Spine, LLC · VIPER 2 SYSTEM TAP, CANNULATED AND SELF-DRILLING 5mm 30-60mm Plus or Minus 0.24 Percent · Class 1
All models of VIPER (60)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705034214078), received between Jun 23, 2017 and Jun 17, 2024 — classified as 3 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Break” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break10
- Entrapment of Device4
- Contamination /Decontamination Problem2
Patient problems
- Foreign Body In Patient5
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information2
- No Known Impact Or Consequence To Patient2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT ON (B)(6) 2024, IT WAS CONFIRMED THAT THERE WAS RESIDUAL BLOOD IN THE HOLLOW PORTION OF THE PRODUCTS IN QUESTION. THEY WERE CLEANED AND STERILIZED AGAIN AT THE HOSPITAL FOR USE ON THE FOLLOWING DAY, WEDNESDAY, (B)(6) 2024. THE FACILITY HAS A VERY HIGH LEVEL OF INFECTION AWARENESS. IT IS LIKELY TO AFFECT PRODUCT ADOPTION. THIS REPORT INVOLVES ONE (1) VIPER 2 SYSTEM TAP, CANNULATED AND SELF-DRILLING 5MM 30-60MM PLUS OR MINUS 0.24 PERCENT. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 19546633 ↗MalfunctionJun 14, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024 THE DEVICE WAS BROKEN INTO TWO PIECES DURING THE SURGERY. ALL BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS NO SURGICAL DELAY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS REPORT IS FOR ONE (1) VIPER2 5MM SELF DRILLING TAP. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19542239 ↗MalfunctionMar 12, 2024Break; Entrapment of Device; Break; Entrapment of Device
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2024, TAP BROKE OFF IN SITU AND GOT STUCK IN THE PATIENT. THE SURGERY WAS COMPLETED SUCCESSFULLY AND THE PATIENT OUTCOME WAS GOOD. THERE WAS A SURGICAL DELAY BUT IT IS UNKNOWN HOW LONG IT WAS. THIS REPORT IS FOR ONE (1) VIPER2 5MM SELF DRILLING TAP, THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Foreign Body In Patient; Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18886981 ↗MalfunctionMar 7, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, DURING A PLIF FOR THE LUMBAR STENOSIS BEFORE STERILIZATION, STEIN WAS FOUND IN THE HOLLOW PART OF THE DEVICES. THE DEVICES WERE DELIVERED ON THE OPERATION DAY, AND THE SURGEON DETERMINED THAT IT WOULD TAKE TIME TO CLEAN. THEREFORE, OTHER SAME DEVICES WERE USED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) VIPER 2 SYSTEM TAP, CANNULATED AND SELF-DRILLING 5MM 30-60MM PLUS OR MINUS 0.24 PERCENT. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16496515 ↗MalfunctionOct 27, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN RUSSIA AS FOLLOWS: IT WAS REPORTED THAT ON OCTOBER 4, 2022, A TAP FROM THE VIPER 2 SET BROKE DURING THE OPERATION. A DELAY IN THE SURGERY OCCURRED DUE TO THE FACT THAT THE BROKEN PART OF THE TAP HAD TO BE REMOVED FROM THE VERTEBRA. THE DELAY WAS APPROXIMATELY 20 MINUTES. THE OPERATION WAS COMPLETED SUCCESSFULLY. THE TAP WAS REMOVED, THE SCREW WAS INSTALLED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE VIPER 2 SYSTEM TAP, CANNULATED AND SELF-DRILLING 5MM 30-60MM PLUS OR MINUS 0.24 PERCENT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15680393 ↗InjurySep 15, 2022Break; Entrapment of Device; Break; Entrapment of Device
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN RUSSIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE OPERATION WAS PERFORMED TO INSTALL THE CAGE AND VIPER 2 SCREWS. FOUR VIPER 2 SCREWS WERE INSTALLED. THE FIRST SCREW WAS INSTALLED USING A SIZE 5 TAP. WHILE INSTALLING THE SECOND SCREW, THE TAP BROKE IN THE PEDICLE OF THE VERTEBRA, AND THE REMOVAL OF THE FRAGMENTS FAILED. THE SCREW WAS INSTALLED TO THIS VERTEBRA AFTERWARD, BUT ALONG A DIFFERENT TRAJECTORY. THE REMAINING TWO SCREWS WERE INSTALLED USING A SIZE 4 TAP. THE PATIENT'S STATUS WAS REPORTED AS SATISFACTORY. THE SURGERY WAS COMPLETED SUCCESSFULLY. THIS REPORT INVOLVES ONE VIPER 2 SYSTEM TAP, CANNULATED AND SELF-DRILLING 5MM 30-60MM PLUS OR MINUS 0.24 PERCENT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 15427990 ↗InjuryMar 11, 2022Break; Entrapment of Device; Break; Entrapment of Device
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN TAIWAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, SURGEON USED AN SURGICAL INSTRUMENT (SCREW TAP) TO FORM A THREAD ON THE PATIENT'S BONE, THE PATIENT'S BONE WAS TOO HARD. THE FRONT END OF TAP WAS BROKEN INTO THE PATIENT'S L5 VERTEBRAL BODY. THE SURGEON COULD NOT REMOVE IT AFTER REPEATED ATTEMPTS AND IT REMAINED IN PATIENT'S BODY. THIS REPORT IS FOR ONE (1) VIPER2 5MM SELF DRILLING TAP. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 13736226 ↗InjuryFeb 25, 2021Break; Entrapment of Device; Break; Entrapment of Device
MalfunctionJun 17, 2019Break; Break
THE FOLLOWING WAS REPORTED: UNKNOWN DEFECT OF INSTRUMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8705005 ↗MalfunctionMay 8, 2019Break; Break
IT WAS REPORTED THAT THE PRODUCT BROKE DURING PROCEDURE. TO REMOVE THE BROKEN PART PHYSICIAN SPEND 2 ADDITIONAL HOURS. NO ADDITIONAL DETAIL AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8591946 ↗MalfunctionJan 14, 2019Break; Break
IT WAS REPORTED THE SURGEON WAS TAPPING OVER A WIRE INTO VERY SCLEROTIC BONE ON THE CONCAVE SIDE OF A SCOLIOTIC CURVE. THE TAP BEGAN "SQUEAK" AND BECAME HARD TO TURN. SUDDENLY IT BEGAN TO SPIN FREELY AS WE NOTICED THAT IT HAD BROKEN IN THE BONE. THE SURGEON THEN "RETRIEVE" THE WIRE FROM THE BONE AND DETERMINED THAT IT WOULD BE MORE DETRIMENTAL TO RETRIEVE THE BROKEN COMPONENT THAN IT WOULD BE TO SIMPLY LEAVE IT IN. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? YES. ACTION TAKEN WHEN EVENT OCCURRED? NONE - DETERMINED TO BE MORE INVASIVE TO RETRIEVE PIECE. WAS PROCEDURE SUCCESSFULLY COMPLETED? YES. WERE FRAGMENTS GENERATED? YES. IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? UNKNOWN. PATIENT STATUS/ OUTCOME / CONSEQUENCES? UNKNOWN. WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED? UNKNOWN. IS THE PATIENT PART OF A CLINICAL STUDY? UNKNOWN.
Patient problems: Foreign Body In Patient
MAUDE report 8244937 ↗MalfunctionJun 23, 2017Break; Break
THERE ARE 100+ ITEMS PLACED ON DC ORTHOKIT QUARANTINE DUE TO BREAKAGES, FUNCTIONAL TESTS FAILURES. THIRTY EIGHT PERTAIN TO DEPUY SPINE PRODUCTS. IN ALL CASES NO DATA AVAILABLE IF ANY INJURES OR ADVERSE EVENTS. NO INFORMATION ABOUT SURGERY, DELAYS, PATIENT RELATED DATA. ONLY PRODUCT CODE/LOT DESCRIPTION AND BREAKAGE DEFINITION IS AVAILABLE. (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6663995 ↗
FDA clearance and classification
The FDA regulates product code HWX (tap, bone) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VIPER 286715500
Has the VIPER 286715500 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705034214078), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the VIPER 286715500?
The FDA's MAUDE database lists 12 adverse event reports for this device, received between Jun 23, 2017 and Jun 17, 2024. In those reports, the most common specific device problems are “Break” (10), “Entrapment of Device” (4) and “Contamination /Decontamination Problem” (2); the most common patient problem is “Foreign Body In Patient” (5). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VIPER 286715500?
None of the reports are classified as deaths, 3 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the VIPER 286715500 FDA approved?
Devices of this type (product code HWX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy Spine, LLC.
Who makes the VIPER 286715500, and is it still sold?
The FDA's device database lists Depuy Spine, LLC as the company responsible for it (brand VIPER, model 286715500). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other TAP, BONE devices
- NA 311.43Synthes GmbH · TAP, BONE172 reports
- NA 311.01Synthes GmbH · TAP, BONE106 reports
- NA 311.431Synthes GmbH · TAP, BONE69 reports
- NA 03.424.124Synthes GmbH · TAP, BONE1 recall7 reports
- N/A 876-474Medtronic Sofamor Danek, Inc. · TAP, BONE1 recall0 reports
- A.L.P.S.® 8241-50-170Biomet Orthopedics, LLC · TAP, BONE1 recall0 reports
- NA 03.424.126Synthes GmbH · TAP, BONE1 recall0 reports
- NA 311.44Synthes GmbH · TAP, BONE41 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.