Depuy Spine, LLC · VIPER 2 SYSTEM MULTI-PIECE SCREW EXTENSION CLOSED · Class 1

All models of VIPER (60)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10705034214320), received between Nov 1, 2019 and Jan 13, 2023 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility11
  • Break4
  • No Apparent Adverse Event1

Patient problems

  • No Consequences Or Impact To Patient10
  • No Clinical Signs, Symptoms or Conditions4
  • No Patient Involvement1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 13, 2023Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2022, THAT THE CUSTOMER COMPLAIN THAT THE THREE-PART ARTICLE CAN NO LONGER BE PROPERLY ASSEMBLED/ASSEMBLED. NO PATIENT INVOLVED. THIS REPORT IS FOR ONE (1) VIPER2 SCREW EXT, MULTI PIECE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16160028 ↗
  • MalfunctionJul 8, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM GERMANY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2022, THE INSERTER TWISTED OFF; SCREW EXTENSION SLEEVE COULD NOT BE FIXED TO THE SCREW. PROCEDURE WAS COMPLETED WITH NO DELAY. THERE WAS NO PATIENT HARM. THIS REPORT IS FOR A VIPER2 SCREW EXT, MULTI PIECE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14960941 ↗
  • MalfunctionNov 4, 2021Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2021, DURING INSERTION, THE SLEEVE LOOSENED SEVERAL TIMES FROM THE VIPER SCREW AND WAS REPLACED BY ANOTHER SLEEVE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THERE WAS NO PATIENT HARM. THIS REPORT IS FOR A VIPER 2 SYSTEM MULTI-PIECE SCREW EXTENSION CLOSED. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12751099 ↗
  • MalfunctionNov 3, 2021Break; No Apparent Adverse Event; Break; No Apparent Adverse Event

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, INTRAOPERATIVELY THE INSTRUMENT FRACTURED AT THE WELDING SEAM WHEN INSERTING THE CAGE. NO FRAGMENT REMAINED IN THE PATIENT. A SECOND INSTRUMENT WAS USED FOR THE FINAL INSERTION OF THE CAGE. THERE WAS NO SIGNIFICANT DELAY AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY PATIENT HARM. THIS REPORT IS FOR ONE (1) VIPER 2 SYSTEM MULTI-PIECE SCREW EXTENSION CLOSED. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12749338 ↗
  • MalfunctionOct 16, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING MINIMALLY INVASIVE SPINAL FUSION PROCEDURE PERFORMED ON (B)(6) 2020, THE VIPER 2 TOWER FAILED TO GRIP ONTO THE TULIP OF THE SCREW AND CAME OFF DURING THE SURGERY. IT WAS REPLACED WITH ANOTHER ONE. THERE WAS SURGICAL DELAY OF FIVE (5) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCES REPORTED. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN) THIS REPORT IS FOR ONE (1) VIPER2 SCREW EXT, MULTI PIECE. THIS IS REPORT 8 OF 8 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10690799 ↗
  • MalfunctionOct 16, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING MINIMALLY INVASIVE SPINAL FUSION PROCEDURE PERFORMED ON (B)(6) 2020, THE VIPER 2 TOWER FAILED TO GRIP ONTO THE TULIP OF THE SCREW, AND CAME OFF DURING THE SURGERY. IT WAS REPLACED WITH ANOTHER ONE. THERE WAS SURGICAL DELAY OF FIVE (5) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCES REPORTED. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN ) THIS REPORT IS FOR ONE (1) VIPER2 SCREW EXT, MULTI PIECE. THIS IS REPORT 7 OF 8 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10690788 ↗
  • MalfunctionOct 16, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING MINIMALLY INVASIVE SPINAL FUSION PROCEDURE PERFORMED ON (B)(6) 2020, THE VIPER 2 TOWER FAILED TO GRIP ONTO THE TULIP OF THE SCREW AND CAME OFF DURING THE SURGERY. IT WAS REPLACED WITH ANOTHER ONE. THERE WAS SURGICAL DELAY OF FIVE (5) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCES REPORTED. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN ). THIS REPORT IS FOR ONE (1) VIPER2 SCREW EXT, MULTI PIECE. THIS IS REPORT 5 OF 8 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10690786 ↗
  • MalfunctionOct 16, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING MINIMALLY INVASIVE SPINAL FUSION PROCEDURE PERFORMED ON (B)(6) 2020, THE VIPER 2 TOWER FAILED TO GRIP ONTO THE TULIP OF THE SCREW AND CAME OFF DURING THE SURGERY. IT WAS REPLACED WITH ANOTHER ONE. THERE WAS SURGICAL DELAY OF FIVE (5) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCES REPORTED. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN ). THIS REPORT IS FOR ONE (1) VIPER2 SCREW EXT, MULTI PIECE. THIS IS REPORT 4 OF 8 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10690784 ↗
  • MalfunctionOct 16, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING MINIMALLY INVASIVE SPINAL FUSION PROCEDURE PERFORMED ON (B)(6) 2020, THE VIPER 2 TOWER FAILED TO GRIP ONTO THE TULIP OF THE SCREW AND CAME OFF DURING THE SURGERY. IT WAS REPLACED WITH ANOTHER ONE. THERE WAS SURGICAL DELAY OF FIVE (5) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCES REPORTED. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) VIPER2 SCREW EXT, MULTI PIECE. THIS IS REPORT 6 OF 8 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10690782 ↗
  • MalfunctionOct 16, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING MINIMALLY INVASIVE SPINAL FUSION PROCEDURE PERFORMED ON (B)(6) 2020, THE VIPER 2 TOWER FAILED TO GRIP ONTO THE TULIP OF THE SCREW AND CAME OFF DURING THE SURGERY. IT WAS REPLACED WITH ANOTHER ONE. THERE WAS SURGICAL DELAY OF FIVE (5) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCES REPORTED. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN ). THIS REPORT IS FOR ONE (1) VIPER2 SCREW EXT, MULTI PIECE. THIS IS REPORT 3 OF 8 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10690780 ↗
  • MalfunctionOct 16, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING MINIMALLY INVASIVE SPINAL FUSION PROCEDURE PERFORMED ON (B)(6) 2020, THE VIPER 2 TOWER FAILED TO GRIP ONTO THE TULIP OF THE SCREW, AND CAME OFF DURING THE SURGERY. IT WAS REPLACED WITH ANOTHER ONE. THERE WAS SURGICAL DELAY OF FIVE (5) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCES REPORTED. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) VIPER2 SCREW EXT, MULTI PIECE. THIS IS REPORT 2 OF 8 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10690768 ↗
  • MalfunctionOct 16, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING MINIMALLY INVASIVE SPINAL FUSION PROCEDURE PERFORMED ON (B)(6) 2020, THE VIPER 2 TOWER FAILED TO GRIP ONTO THE TULIP OF THE SCREW AND CAME OFF DURING THE SURGERY. IT WAS REPLACED WITH ANOTHER ONE. THERE WAS SURGICAL DELAY OF FIVE (5) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCES REPORTED. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN ). THIS REPORT IS FOR ONE (1) VIPER2 SCREW EXT, MULTI PIECE. THIS IS REPORT 1 OF 8 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10690571 ↗
  • MalfunctionFeb 13, 2020Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ONE "WING" BROKE ON THE INNER COMPONENT OF THE INSTRUMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 9707012 ↗
  • MalfunctionNov 1, 2019Break; Break

    THERE IS A FISSURE IN THE INSTRUMENT. SEEN IN CSSD. PROCEDURE WAS FINISHED SUCCESSFULLY WITHOUT SURGERY DELAY. THIS COMPLAINT INVOLVES TWO (2) DEVICES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9265999 ↗
  • MalfunctionNov 1, 2019Break; Break

    THERE IS A FISSURE IN THE INSTRUMENT. SEEN IN CSSD. PROCEDURE WAS FINISHED SUCCESSFULLY WITHOUT SURGERY DELAY. THIS COMPLAINT INVOLVES TWO (2) DEVICES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9265973 ↗

FDA clearance and classification

The FDA regulates product code LXH (orthopedic manual surgical instrument) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VIPER 286725225

Has the VIPER 286725225 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034214320), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VIPER 286725225?

The FDA's MAUDE database lists 15 adverse event reports for this device, received between Nov 1, 2019 and Jan 13, 2023. In those reports, the most common specific device problems are “Device-Device Incompatibility” (11) and “Break” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VIPER 286725225?

None of the reports are classified as deaths, 0 as injuries and 15 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the VIPER 286725225 FDA approved?

Devices of this type (product code LXH) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy Spine, LLC.

Who makes the VIPER 286725225, and is it still sold?

The FDA's device database lists Depuy Spine, LLC as the company responsible for it (brand VIPER, model 286725225). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page