VIPER 286735730Recalls & reported problems

Depuy Spine, LLC · VIPER 2 SYSTEM REDUCTION THREADED POST 5.5 · Class 1

All models of VIPER (60)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10705034214665), received between Feb 1, 2021 and Aug 11, 2022 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility11
  • Device Slipped9
  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions15

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN NORWAY AS FOLLOWS: IT WAS REPORTED BY THE SALES REP IN NORWAY THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE SPLINTS INSIDE THE ARTICLE WAS GONE, AND IT IS NO LONGER STAYING IN PLACE ON THE REDUCTION DRIVER DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 FOR (B)(4). THIS REPORT IS FOR A THREADED POST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15212415 ↗
  • MalfunctionMay 23, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED BY THE SALES REP IN (B)(6) THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE SPLINTS INSIDE THE ARTICLE WAS GONE, AND IT IS NO LONGER STAYING IN PLACE ON THE REDUCTION DRIVER DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS COMPLAINT INVOLVES ONE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14471295 ↗
  • MalfunctionApr 25, 2022Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: THIS IS REPORT 5 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP IN (B)(6) THAT DURING AN INSPECTION, IT WAS OBSERVED THAT THE SMALL INNER SPRINGS ON THE IMPLANT INSERTER/EXTRACTOR DEVICE WERE LOOSE; AND THAT THE PEN DID NOT HOLD ON THE SHAFT ANY LONGER. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) REDUCTION THREADED POST DEVICE. THIS COMPLAINT INVOLVES 5 DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14193090 ↗
  • MalfunctionApr 25, 2022Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: THIS IS REPORT 2 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP IN GERMANY THAT DURING AN INSPECTION, IT WAS OBSERVED THAT THE SMALL INNER SPRINGS ON THE IMPLANT INSERTER/EXTRACTOR DEVICE WERE LOOSE; AND THAT THE PEN DID NOT HOLD ON THE SHAFT ANY LONGER. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) REDUCTION THREADED POST DEVICE. THIS COMPLAINT INVOLVES 5 DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14193087 ↗
  • MalfunctionApr 25, 2022Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: THIS IS REPORT 4 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP IN (B)(6) THAT DURING AN INSPECTION, IT WAS OBSERVED THAT THE SMALL INNER SPRINGS ON THE IMPLANT INSERTER/EXTRACTOR DEVICE WERE LOOSE; AND THAT THE PEN DID NOT HOLD ON THE SHAFT ANY LONGER. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) REDUCTION THREADED POST DEVICE. THIS COMPLAINT INVOLVES 5 DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14193086 ↗
  • MalfunctionApr 25, 2022Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: THIS IS REPORT 3 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP IN GERMANY THAT DURING AN INSPECTION, IT WAS OBSERVED THAT THE SMALL INNER SPRINGS ON THE IMPLANT INSERTER/EXTRACTOR DEVICE WERE LOOSE; AND THAT THE PEN DID NOT HOLD ON THE SHAFT ANY LONGER. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) REDUCTION THREADED POST DEVICE. THIS COMPLAINT INVOLVES 5 DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14193085 ↗
  • MalfunctionApr 25, 2022Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: THIS IS REPORT 1 OF 5 FOR (B)(4). IT WAS REPORTED BY THE SALES REP IN (B)(6) THAT DURING AN INSPECTION, IT WAS OBSERVED THAT THE SMALL INNER SPRINGS ON THE IMPLANT INSERTER/EXTRACTOR DEVICE WERE LOOSE; AND THAT THE PEN DID NOT HOLD ON THE SHAFT ANY LONGER. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) REDUCTION THREADED POST DEVICE. THIS COMPLAINT INVOLVES 5 DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14193082 ↗
  • MalfunctionFeb 25, 2022Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING KIT INSPECTION ON (B)(6) 2022, TWO THREADED POSTS DID NOT PROPERLY SECURE TO THE THREADED DRIVER. FOR ACCURACY, IT WAS ATTEMPTED WITH MULTIPLE DRIVERS. DIFFERENT THREADED POSTS ON THREADED DRIVERS WERE ALSO ATTEMPTED AND THOSE SECURED PROPERLY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR A VIPER REDUCTION THREADED POST. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13599231 ↗
  • MalfunctionFeb 25, 2022Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING KIT INSPECTION ON (B)(6) 2022, TWO THREADED POSTS DID NOT PROPERLY SECURE TO THE THREADED DRIVER. FOR ACCURACY, IT WAS ATTEMPTED WITH MULTIPLE DRIVERS. DIFFERENT THREADED POSTS ON THREADED DRIVERS WERE ALSO ATTEMPTED AND THOSE SECURED PROPERLY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR A VIPER REDUCTION THREADED POST. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13599190 ↗
  • MalfunctionSep 14, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2021, DURING PRE-OP FOR A PROCEDURE, THE EXTENSIONS HAD BEEN CRACKED. THE TIP OF THE POLY DRIVER SHAFT WAS BROKEN AND UNABLE TO ENGAGE INTO THE SCREW. ALSO, THE THREADED POST OF THE X25 DRIVER IS NOT STAYING AND CONTINUES TO SLIDE OFF. THERE WERE NO PATIENT CONSEQUENCES. THE PROCEDURE OUTCOME IS UNKNOWN. CONCOMITANT DEVICE REPORTED. UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QTY UNKNOWN). THIS COMPLAINT INVOLVES FIVE (5) DEVICES. THIS REPORT IS FOR (1) REDUCTION THREADED POST. THIS REPORT IS 2 OF 5 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12470900 ↗
  • MalfunctionMar 4, 2021Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING THE CLEANING PROCEDURE IT WAS NOTICED THAT REDUCTION THREADED POST DID NOT SNAP INTO PLACE WHEN PLACED OVER REDUCTION SHAFT. THERE WAS NO SURGERY OR PATIENT IMPACT. THIS REPORT INVOLVES ONE (1) VIPER 2 SYSTEM REDUCTION THREADED POST 5.5. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11414660 ↗
  • MalfunctionFeb 1, 2021Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, FOUR (4) REDUCTION THREADED POSTS WERE LOOSE AND APPEARED TO HAVE LOST THE TIGHT SEAL THEY NORMALLY PROVIDE WHEN PLACED ON THE REDUCTION DRIVERS. THE ISSUE WAS NOTICED DURING TRAINING. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) REDUCTION THREADED POST. THIS IS REPORT 4 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11259876 ↗
  • MalfunctionFeb 1, 2021Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, FOUR (4) REDUCTION THREADED POSTS WERE LOOSE AND APPEARED TO HAVE LOST THE TIGHT SEAL THEY NORMALLY PROVIDE WHEN PLACED ON THE REDUCTION DRIVERS. THE ISSUE WAS NOTICED DURING TRAINING. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) REDUCTION THREADED POST. THIS IS REPORT 2 OF 4 FOR (B)(4). .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11259861 ↗
  • MalfunctionFeb 1, 2021Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, FOUR (4) REDUCTION THREADED POSTS WERE LOOSE AND APPEARED TO HAVE LOST THE TIGHT SEAL THEY NORMALLY PROVIDE WHEN PLACED ON THE REDUCTION DRIVERS. THE ISSUE WAS NOTICED DURING TRAINING. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) REDUCTION THREADED POST. THIS IS REPORT 3 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11259853 ↗
  • MalfunctionFeb 1, 2021Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, FOUR (4) REDUCTION THREADED POSTS WERE LOOSE AND APPEARED TO HAVE LOST THE TIGHT SEAL THEY NORMALLY PROVIDE WHEN PLACED ON THE REDUCTION DRIVERS. THE ISSUE WAS NOTICED DURING TRAINING. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) REDUCTION THREADED POST. THIS IS REPORT 1 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11259800 ↗

FDA clearance and classification

The FDA regulates product code LXH (orthopedic manual surgical instrument) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VIPER 286735730

Has the VIPER 286735730 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034214665), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VIPER 286735730?

The FDA's MAUDE database lists 15 adverse event reports for this device, received between Feb 1, 2021 and Aug 11, 2022. In those reports, the most common specific device problems are “Device-Device Incompatibility” (11), “Device Slipped” (9) and “Break” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VIPER 286735730?

None of the reports are classified as deaths, 0 as injuries and 15 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the VIPER 286735730 FDA approved?

Devices of this type (product code LXH) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy Spine, LLC.

Who makes the VIPER 286735730, and is it still sold?

The FDA's device database lists Depuy Spine, LLC as the company responsible for it (brand VIPER, model 286735730). Its record shows it as in commercial distribution.