SKYLINE 286830400Recalls & reported problems
Depuy Spine, LLC · SKYLINE ANTERIOR CERVICAL PLATE SYSTEM STANDARD DRIVER · Class 1
All models of SKYLINE (8)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705034215419), received between Mar 18, 2024 and Mar 18, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Device-Device Incompatibility2
- Material Twisted/Bent2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2026Break
IT WAS REPORTED THAT INTRA-OPERATIVELY BOTH SKYLINE SCREWDRIVER ENDS BROKE AWAY FROM THE SHAFT RENDERING THEM UNUSABLE, THIS CAUSED A MINUTES DELAY IN THE SURGERY AS WE OPENED ANOTHER TRAY BUT HAD NO ADVERSE EFFECT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24631844 ↗MalfunctionMar 18, 2026Break
IT WAS REPORTED THAT INTRA-OPERATIVELY BOTH SKYLINE SCREWDRIVER ENDS BROKE AWAY FROM THE SHAFT RENDERING THEM UNUSABLE, THIS CAUSED A MINUTE DELAY IN THE SURGERY AS WE OPENED ANOTHER TRAY BUT HAD NO ADVERSE EFFECT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24626072 ↗MalfunctionMar 13, 2026Break
IT WAS REPORTED THAT INTRA-OPERATIVELY BOTH SKYLINE SCREWDRIVER ENDS BROKE AWAY FROM THE SHAFT RENDERING THEM UNUSABLE. THIS CAUSED A MINUTE DELAY IN THE SURGERY AS WE OPENED ANOTHER TRAY. THERE WAS NO ADVERSE EFFECT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24598102 ↗MalfunctionNov 17, 2025Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent
IT WAS REPORTED THAT DURING A PROCEDURE, IT WAS FOUND THAT INSTRUMENTS HAVE WORN DOWN AND STRIPPED DUE TO USAGE OVER TIME. THE DRILLS DID NOT CUT THROUGH BONE ADEQUATELY AND THE DRIVERS DID NOT HOLD A SCREW. THE COUPLING ON THE HANDLE WAS JARRED. ITEMS HAVE BEEN REMOVED FROM CIRCULATION. MULTIPLE SETS NEEDED TO BE OPENED TO FIND ALTERNATION INSTRUMENTS. THERE WAS A 10 MINUTES OF SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23572138 ↗MalfunctionMar 18, 2024Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED ON FEBRUARY 29, 2024, THAT THE SKYLINE WAS BROUGHT IN ON A LONG TERM LOAN. SURGEON WAS TRIALLING SKYLINE ANTERIOR CERVICAL PLATING SYSTEM IN RVI. THE X15 EXPANSION TIP SCREWDRIVER WOULD NOT RETAIN THE SCREWS AND THEREFORE COULD NOT BE USED. THE OTHER TWO NORMAL SCREWDRIVERS ALSO SEEMED TO BE STRIPPED AND WOULD NOT RETAIN THE SCREWS DESPITE SEVERAL EFFORTS. TRIED TO LOAD 4X DIFFERENT SCREWS AND HAD THE SAME ISSUE WITH EACH SCREW. SURGEON USED A PLATING SYSTEM FROM ANOTHER COMPANY WHICH WAS READILY AVAILABLE IN THEATRE. THERE SURGERY WAS SUCCESSFULLY COMPLETED. THERE WAS NO ADVERSE EVENT TO THE PATIENT. THIS REPORT IS FOR ONE (1)SKYLINE STANDARD DRIVER THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18923486 ↗
FDA clearance and classification
The FDA regulates product code HWR (driver, prosthesis) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SKYLINE 286830400
Has the SKYLINE 286830400 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705034215419), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SKYLINE 286830400?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Mar 18, 2024 and Mar 18, 2026. In those reports, the most common specific device problems are “Break” (3), “Device-Device Incompatibility” (2) and “Material Twisted/Bent” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SKYLINE 286830400?
None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SKYLINE 286830400 FDA approved?
Devices of this type (product code HWR) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy Spine, LLC.
Who makes the SKYLINE 286830400, and is it still sold?
The FDA's device database lists Depuy Spine, LLC as the company responsible for it (brand SKYLINE, model 286830400). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DRIVER, PROSTHESIS devices
- ATTUNE 2544-01-005Depuy (Ireland) · DRIVER, PROSTHESIS1,355 reports
- EXPEDIUM VERSE 299704230Medos International Sàrl · DRIVER, PROSTHESIS868 reports
- NA 9200-10-029Depuy (Ireland) · DRIVER, PROSTHESIS740 reports
- MOSS 279712600Depuy Spine, LLC · DRIVER, PROSTHESIS702 reports
- NA 2598-07-440Depuy (Ireland) · DRIVER, PROSTHESIS660 reports
- NA 2598-07-570Depuy (Ireland) · DRIVER, PROSTHESIS502 reports
- VIPER 286745550Depuy Spine, LLC · DRIVER, PROSTHESIS255 reports
- SUMMIT 2570-05-100Depuy (Ireland) · DRIVER, PROSTHESIS224 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page