SKYLINE 286850000Recalls & reported problems

Depuy Spine, LLC · SKYLINE ANTERIOR CERVICAL PLATE SYSTEM TEMPORARY FIXATION PIN - INSERTER · Class 1

All models of SKYLINE (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705034215464), received between May 19, 2020 and Dec 12, 2025 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Break” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break8
  • Device-Device Incompatibility4
  • Contamination /Decontamination Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    STILL CONTAMINATED INSTRUMENTS AFTER REPROCESSING. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23789218 ↗
  • MalfunctionJan 8, 2025Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024. WHEN THE OPERATION WAS CARRIED OUT, IT WAS DISCOVERED THAT THE INSERTER WAS BROKEN INTO TWO PARTS. THE INNER SHAFT WAS NO LONGER IN CONTACT WITH THE HANDLE. THE OPERATION COULD BE CARRIED OUT AS USUAL. THE OPERATION TIME WAS NOT EXTENDED BY THIS EVENT. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21106027 ↗
  • MalfunctionJan 8, 2025Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024. WHEN THE OPERATION WAS CARRIED OUT, IT WAS DISCOVERED THAT THE INSERTER WAS BROKEN INTO TWO PARTS. THE INNER SHAFT WAS NO LONGER IN CONTACT WITH THE HANDLE. THE OPERATION COULD BE CARRIED OUT AS USUAL. THE OPERATION TIME WAS NOT EXTENDED BY THIS EVENT. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21106026 ↗
  • MalfunctionApr 26, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE FIXATION PIN INSERTER IS NOT RETAINING THE PINS. UNKNOWN IF SURGERY WAS DELAYED. PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR ONE (1) SKY TFP INSERTER THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19197266 ↗
  • MalfunctionFeb 13, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    SHAFT BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18694308 ↗
  • MalfunctionFeb 13, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    SHAFT BROKE OFF.

    Patient problems: Insufficient Information

    MAUDE report 18694307 ↗
  • MalfunctionAug 10, 2023Break; Break

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT ON JULY 25, 2023, THE TEMPORARY FIXATION PIN INSERTER FROM OUR SKYLINE SIEVE HAS APPARENTLY BEEN COMPLETELY BROKEN APART AT OUR ZSVA AND DISMANTLED INTO THREE PARTS. NO PATIENT HARM. THIS REPORT IS FOR ONE (1) SKYLINE ANTERIOR CERVICAL PLATE SYSTEM TEMPORARY FIXATION PIN - INSERTER. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17503292 ↗
  • MalfunctionAug 10, 2023Break; Break

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE TEMPORARY FIXATION PIN INSERTER FROM OUR SKYLINE SIEVE HAS APPARENTLY BEEN COMPLETELY BROKEN APART AT OUR ZSVA AND DISMANTLED INTO THREE PARTS. NO PATIENT HARM. THIS REPORT IS FOR ONE (1) SKYLINE ANTERIOR CERVICAL PLATE SYSTEM TEMPORARY FIXATION PIN - INSERTER. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17503289 ↗
  • MalfunctionOct 4, 2022Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2022, DURING SURGERY, IT WAS NOTICED THAT THE TEMPORARY PIN DRIVER BROKE. THE SURGERY WAS COMPLETED SUCCESSFULLY. NO FRAGMENTS WERE RETAINED IN THE PATIENT'S BODY. THIS REPORT INVOLVES ONE SKYLINE ANTERIOR CERVICAL PLATE SYSTEM TEMPORARY FIXATION PIN - INSERTER . THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15536231 ↗
  • MalfunctionFeb 9, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE DEVICE WAS DIRTY. THE DEVICE IS PART OF LOAN SET AND IT WAS DECONTAMINATED IN MIELE MACHINE IN LOC DC. THE SURGERY WAS EXTENDED SINCE THE INSTRUMENT WAS NOT OPERATIONAL. THERE WAS A CHANGE IN THE SURGICAL METHOD AS IT WAS NOT POSSIBLE TO USE THE INSTRUMENT TO PLACE THE TEMPORARY FIXATION PINS. IT WAS NECESSARY TO USE GUIDES AND DRILLS TO ACHIEVE THE SAME PURPOSE, BUT THIS ACTION INVOLVED USING MORE INSTRUMENTS IN THE LITTLE SPACE AVAILABLE AND GENERATED MORE BONE INVASION. THERE WAS NO PATIENT IMPACT. THIS REPORT IS FOR ONE (1) SKYLINE ANTERIOR CERVICAL PLATE SYSTEM TEMPORARY FIXATION PIN - INSERTER THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13502061 ↗
  • MalfunctionAug 25, 2020Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, THE PIN INSERTER BROKE OFF INTRAOPERATIVELY. THE SURGERY WAS SUCCESSFULLY COMPLETED. THERE IS NO FURTHER INFORMATION AVAILABLE. CONCOMITANT DEVICE REPORTED: UNKNOWN PLATE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) SKY TFP INSERTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10450906 ↗
  • MalfunctionMay 19, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE THAT THE TEMPORARY FIXATION PIN INSERTER CANNOT BE DISASSEMBLED. THE INSTRUCTIONS FOR USE (IFU) SUGGESTS A PH VALUE OF 7 - 9 TO CLEAN THE INSTRUMENT BUT THE CUSTOMER NEEDS AN APPROVAL TO CLEAN THE INSTRUMENT WITH A PH VALUE OF GREATER THAN (>) 10. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) SKY TFP INSERTER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 10071711 ↗

FDA clearance and classification

The FDA regulates product code HWR (driver, prosthesis) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the SKYLINE 286850000

Has the SKYLINE 286850000 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034215464), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SKYLINE 286850000?

The FDA's MAUDE database lists 12 adverse event reports for this device, received between May 19, 2020 and Dec 12, 2025. In those reports, the most common specific device problems are “Break” (8), “Device-Device Incompatibility” (4) and “Contamination /Decontamination Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SKYLINE 286850000?

None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SKYLINE 286850000 FDA approved?

Devices of this type (product code HWR) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy Spine, LLC.

Who makes the SKYLINE 286850000, and is it still sold?

The FDA's device database lists Depuy Spine, LLC as the company responsible for it (brand SKYLINE, model 286850000). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.