SPOTLIGHT 292900610Recalls & reported problems
Depuy Spine, LLC · SPOTLIGHT ACCESS SYSTEM TABLE CLAMP · Class 1
All models of SPOTLIGHT (16)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10705034225357), received between Jun 15, 2020 and Apr 27, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Device Slipped2
- Device-Device Incompatibility2
- Detachment of Device or Device Component1
- Material Twisted/Bent1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 27, 2026Break
IT WAS REPORTED THAT 2MM PITUITARY MOUTH WAS LOOSE. CLAMP SPRING BROKE. THERE WAS NO SURGERY DELAYED DUE TO THE REPORTED EVENT. PROCEDURE SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT STATUS/ OUTCOME / CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25001042 ↗MalfunctionJan 8, 2026Detachment of Device or Device Component
IT WAS REPORTED ON AN UNKNOWN DATE THAT SCREW ON BOTTOM PART OF CLAMP POPPED OFF. SURGERY WAS NOT DELAYED. PROCEDURE SUCCESSFULLY COMPLETED. NO FRAGMENTS WERE GENERATED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24002637 ↗MalfunctionJan 24, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON JANUARY 4 2024, THE CLAMP IN QUESTION HAD A DEFECTIVE "WING SCREW" WHICH PREVENTED IT FROM BEING TIGHTENED OR LOOSENED TO THE OPERATING TABLE. IT WAS NOTICED AS THE NURSES WERE OPENING THE SETS SO THERE WAS NO DELAY IN SURGERY. A DIFFERENT SET FROM THE COMPETITOR HAD TO BE USED INSTEAD OF THIS ONE. THIS REPORT IS FOR ONE (1) SPOT TABLE CLAMP THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4)
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18570302 ↗MalfunctionJul 18, 2022Break; Break
IT WAS REPORTED THAT ON (B)(6) 2022, AFTER THE CASE THE RIGID ARM CLAMP WAS LEFT ON THE BED. ONE OF THE KNOB WAS BROKEN ACCIDENTALLY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE SPOT TABLE CLAMP THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15040724 ↗MalfunctionMar 2, 2022Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT ON (B)(6) 2022, THE TIGHTENING CLAMP PHALANGE WAS STRIPPED. THERE WERE NO PATIENT CONSEQUENCES. PROCEDURE OUTCOME IS UNKNOWN. THIS REPORT IS FOR A TIGHTENING CLAMP. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13652471 ↗MalfunctionJun 4, 2021Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam
IT WAS REPORTED THAT ON (B)(6) 2021, DURING DISCECTOMY PROCEDURE THE NURSING STAFF WENT TO REMOVE THE SPOTLIGHT TABLE CLAMP FROM THE OPERATING ROOM OPEN JACKSON TABLE AND COULD NOT ADVANCE THE THREADING MECHANISM COUNTER CLOCKWISE TO RELEASE THE CLAMP FROM THE ATTACHMENT. FOLLOWING THE CASE, THE NURSING STAFF USED VICE GRIPS TO RELEASE THE STRIPPED THREADED MECHANISM AND REMOVE THE CLAMP FROM THE JACKSON TABLE ATTACHMENT. NURSING STAFF REMOVED THE BAR ATTACHMENT FROM THE TABLE SO THEY COULD FLIP THE PATIENT AND PROCEDURE WITH WAKING UP. THE INCISION WAS CLOSED BUT THE PATIENT WAS STILL POSITIONED ON THE TABLE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. NO FRAGMENTS GENERATED IN CLOSE PROXIMITY TO THE SURGICAL SITE, ALL WERE GENERATED OFF OF THE STERILE FIELD AND NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) SPOTLIGHT ACCESS SYSTEM TABLE CLAMP. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11942342 ↗MalfunctionMar 3, 2021Device Slipped; Device Slipped
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021 DURING A HERNIA EXTIRPATION TO TREAT A LUMBAR SPINAL CANAL STENOSIS, THE CLAMP WAS WOBBLIER THAN USUAL. THIS REPORT IS FOR ONE (1) SPOTLIGHT ACCESS SYSTEM TABLE CLAMP. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11404741 ↗MalfunctionNov 11, 2020Device Slipped; Device Slipped
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2020, WHILE FIXING THE RIGID ARM TO THE BED, THE CLAMP WAS SPINNING AND WOULDN¿T TIGHTEN. THE FIXATION THAT CLAMPS ONTO THE BED WAS ALSO LOOSE DESPITE TIGHTENING AS MUCH AS POSSIBLE. A NON-SYNTHES PRODUCT WAS USED INSTEAD. PROCEDURE WAS COMPLETED WITH A FIFTEEN (15) MINUTE DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR A SPOT TABLE CLAMP. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10822647 ↗MalfunctionJun 15, 2020Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, AT THE END OF THE CASE, THE CLAMP BROKE WHEN THE SCRUB TEAM WAS ATTEMPTING TO UNSCREW THE CLAMP HANDLE. THE PROCEDURE HAD BEEN COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT INVOLVEMENT, THE PATIENT HAD BEEN REMOVED FROM THE THEATER PRIOR TO THE CLAMP BREAKING. THIS REPORT INVOLVES ONE (1) SPOT TABLE CLAMP. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10154097 ↗
FDA clearance and classification
The FDA regulates product code FWZ (operating room accessories table tray) under 21 CFR 878.4950 as a Class I device. 21 CFR 878.4950 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SPOTLIGHT 292900610
Has the SPOTLIGHT 292900610 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705034225357), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SPOTLIGHT 292900610?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Jun 15, 2020 and Apr 27, 2026. In those reports, the most common specific device problems are “Break” (3), “Device Slipped” (2) and “Device-Device Incompatibility” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SPOTLIGHT 292900610?
None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the SPOTLIGHT 292900610, and is it still sold?
The FDA's device database lists Depuy Spine, LLC as the company responsible for it (brand SPOTLIGHT, model 292900610). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Operating Room Accessories Table Tray devices
- MAYFIELD® A2101Integra Lifesciences Corporation · Operating Room Accessories Table Tray100 reports
- NA 210080Mako Surgical Corp. · Operating Room Accessories Table Tray74 reports
- ALLEN 104002Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall8 reports
- N/A 7210569Smith & Nephew, Inc. · Accessories, operating-room, table (kit)57 reports
- ALLEN 107004Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall5 reports
- ALLEN 108001Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall3 reports
- NA 7210551Smith & Nephew, Inc. · Accessories, operating-room, table (kit)41 reports
- MAYFIELD® A1018Integra Lifesciences Corporation · Operating Room Accessories Table Tray36 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page