Skip to content

SPOTLIGHT 292904201Recalls & reported problems

Depuy Spine, LLC · SPOTLIGHT ACCESS SYSTEM STRAIGHT MICROPITUITARY 2mm · Class 1

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034225678), received between Sep 1, 2017 and Dec 29, 2025 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Device Embedded In Tissue or Plaque1
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 29, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THE KEY IS DAMAGED.

    Patient problems: Insufficient Information

    MAUDE report 23906878 ↗
  • MalfunctionJan 8, 2025Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE TWO PITUITARIES WERE BROKEN AND ONE FINAL TIGHTENER WAS STRIPPED DURING A CASE. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PATIENT CONSEQUENCES. THE FRAGMENTS WERE NOT GENERATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21102972 ↗
  • MalfunctionNov 15, 2024Break; Break

    IT WAS REPORTED THAT THE DOCTOR WAS PERFORMING A TLIF PROCEDURE ON AN UNKNOWN DATE AND DURING THE DISSECTION WAS TRYING TO BITE OFF A BONE SPUR. THE PITUITARY SNAPPED IN HALF. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20693681 ↗
  • InjuryFeb 22, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2021, THE SURGEON WAS PERFORMING A MICRODISCECTOMY THROUGH A SPOTLIGHT PORT. WHILE USING THE MICROPITUITARY, THE JAW BROKE. THE PIECE WAS DEEP IN THE PATIENT AND HAD TO BE RETRIEVED USING VARIOUS TOOLS AND THE USE OF X-RAY GUIDANCE. FRAGMENTS WERE GENERATED AND REMOVED. THERE WAS A SURGICAL DELAY OF TWENTY (20) MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT HARM/CONSEQUENCE. THIS REPORT IS FOR ONE (1) SPOT 2MM STRAIGHT MICROPITUITA. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11359123 ↗
  • MalfunctionOct 29, 2019Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED THAT THE SPOT STRAIGHT WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9251528 ↗
  • MalfunctionSep 1, 2017Break; Break

    INSTRUMENT BROKE DURING SURGERY. TIP LEFT IN PATIENT. PER COMPLAINT FORM: SURGEON BROKE OFF TIP OF 2MM PITUITARY FROM SPOTLIGHT INSTRUMENT SET IN DISC SPACE OF PATIENT DURING MIS MICRODISCECTOMY SURGERY. TINY PIECE WAS UNABLE TO BE REMOVED AND THUS LEFT IN PATIENT. X-RAY CONFIRMED PIECE LEFT IN L4-L5 DISC SPACE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 6839816 ↗

Other RONGEUR devices

All RONGEUR devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.