VIPER 186731740Recalls & reported problems
Medos International Sàrl · VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 x 40mm · Class 2 · Implant
All models of VIPER (121)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 17
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 10705034352008), received between Apr 9, 2018 and Jan 9, 2024 — classified as 17 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem11
- Break6
- Device Slipped3
- Appropriate Device Problem Term/Code Not Available1
- Entrapment of Device1
- Material Twisted/Bent1
- Migration1
Patient problems
- Osteolysis8
- No Clinical Signs, Symptoms or Conditions5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Injury2
- Insufficient Information2
- Not Applicable2
- Pain2
- Foreign Body In Patient1
Recent adverse event reports
The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.
InjuryJan 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS AN UNKNOWN SURGERY ON (B)(6) 2018. AFTER THE SURGERY, THE EXTENDED RE-FUSION SURGERY WILL BE PERFORMED DUE TO THE ADJACENT INTERVERTEBRAL DISORDERS. IT IS CONFIRMED L4/5 ACHIEVED OSTEOSYNTHESIS. THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2023. THE SET SCREWS, RODS AND THE CROSSLINKS WERE REMOVED. PLIF WAS PERFORMED ON L1/2, AND TLIF WAS PERFORMED ON L2/3 AND EXTENDED. L1: EXPEDIUM DIAMETER5.0-40×2. L2 LEFT: DIAMETER5.0-45 RIGHT DIAMETER6.0-45. L2/3: CAGECONCORDE H8×W9×L23×ANGLE5. CROSSLINK: A4 ANDA5. L1-L5: ROD CONNECTION. REMARKS: (B)(4) ARE INVOLVED WITH THE SAME EVENT. (B)(4) (DEPUY SPINE): SCREWS. (B)(4) (SYNTHES SPINE): CAGES. THIS REPORT IS FOR ONE (1) 5.5 TI CORT FIX 7X40MM. THIS IS REPORT 10 OF 10 FOR COMPLAINT: (B)(4).
Patient problems: Osteolysis
MAUDE report 18476649 ↗InjuryJan 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED AN UNKNOWN SPINE SURGERY OCCURRED ON (B)(6), 2018. POST OPERATIVELY, THE EXTENDED RE-FUSION SURGERY WILL BE PERFORMED DUE TO THE ADJACENT INTERVERTEBRAL DISORDERS. IT IS CONFIRMED L4/5 ACHIEVED OSTEOSYNTHESIS. THE PLAN IS TLIF FOR L2/3, AND L5 MIGHT BE REMOVED DEPENDING ON INTRAOPERATIVE DECISION. NO FURTHER INFORMATION IS AVAILABLE. (B)(4) ARE LINKED TO (B)(4). (B)(4) (DEPUY SPINE): SCREWS. (B)(4) (SYNTHES SPINE): CAGES. THIS REPORT IS FOR ONE VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 40MM FOR (B)(4). THIS REPORT IS FOR ONE VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 40MM FOR (B)(4).
Patient problems: Osteolysis
MAUDE report 18475062 ↗MalfunctionNov 8, 2023Device Slipped; Device Slipped
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN TAIWAN AS FOLLOWS: IT WAS REPORTED (B)(6) 2023, THAT THE HEAD FELL OFF HALFWAY THROUGH THE LOCK, AND THE SLEEVE HELD THE NAIL HEAD AND FELL OFF (THE SLEEVE IS STILL MOVING). CHANGE NEW PRODUCT. YES, THERE WAS PATIENT INVOLVEMENT. THE OPERATION WENT WELL, NO ADVERSE REACTION. THERE WERE NO PATIENT OUTCOME OR CONSEQUENCES THIS REPORT IS FOR ONE (1) 5.5 TI CORT FIX 7X40MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18095036 ↗InjurySep 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT AN UNKNOWN SURGERY WAS PERFORMED ON (B)(6) 2023. ON (B)(6) 2023, A REOPERATION VIA POSTERIOR FUSION (L3 ¿ S2) WAS PERFORMED TO EXTEND FIXATION. THE REVISION SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. THIS REPORT INVOLVES ONE VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 40MM. THIS REPORT IS RELATED TO (B)(4), WHICH CAPTURES THE RODS INVOLVED IN THIS EVENT. THIS REPORT IS RELATED TO (B)(4), WHICH REPORTS A DEVICE MALFUNCTION THAT OCCURRED DURING THE REVISION SURGERY. THIS IS REPORT 1 OF 13 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 17818363 ↗InjurySep 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT AN UNKNOWN SURGERY WAS PERFORMED ON (B)(6) 2023. ON (B)(6) 2023, A REOPERATION VIA POSTERIOR FUSION (L3 ¿ S2) WAS PERFORMED TO EXTEND FIXATION. THE REVISION SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. THIS REPORT INVOLVES ONE VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 40MM. THIS REPORT IS RELATED TO (B)(4), WHICH CAPTURES THE RODS INVOLVED IN THIS EVENT. THIS REPORT IS RELATED TO (B)(4), WHICH REPORTS A DEVICE MALFUNCTION THAT OCCURRED DURING THE REVISION SURGERY. THIS IS REPORT 2 OF 13 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 17818347 ↗InjurySep 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
(B)(4) AND (B)(4) ARE RELATED TO EACH OTHER. (B)(4) HAS BEEN CREATED TO CAPTURE THE ADDITIONAL IPS.
Patient problems: Osteolysis
MAUDE report 17790133 ↗InjurySep 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
(B)(4) ARE RELATED TO EACH OTHER. (B)(4) HAS BEEN CREATED TO CAPTURE THE ADDITIONAL IPS.
Patient problems: Osteolysis
MAUDE report 17790132 ↗InjurySep 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
(B)(4) ARE RELATED TO EACH OTHER. (B)(4) HAS BEEN CREATED TO CAPTURE THE ADDITIONAL IPS.
Patient problems: Osteolysis
MAUDE report 17790131 ↗InjuryMar 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT UNDERWENT A SPINAL FUSION (L3/4) ON (B)(6) 2014. AFTER SURGERY, ADJACENT INTERVERTEBRAL DISORDER OCCURRED AT L2/3 AND L4/5. ON (B)(6) 2023, A REOPERATION WAS PERFORMED. IN THE REOPERATION, SET SCREWS AND RODS OF L3/4 WERE REMOVED, AND SCREWS WERE ADDED TO L2 AND L5. TLIF TREATING L2/3 AND TLIF TREATING L4/5 WERE DONE. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. THERE IS NO PROBLEM IN THE FUSION SITE TREATED IN THE INITIAL SURGERY IN 2024. THIS REPORT INVOLVES ONE VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 40MM. THIS IS REPORT 4 OF 7 FOR (B)(4).
Patient problems: Osteolysis
MAUDE report 16644042 ↗InjuryMar 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT UNDERWENT A SPINAL FUSION (L3/4) ON (B)(6) 2014. AFTER SURGERY, ADJACENT INTERVERTEBRAL DISORDER OCCURRED AT L2/3 AND L4/5. ON (B)(6) 2023, A REOPERATION WAS PERFORMED. IN THE REOPERATION, SET SCREWS AND RODS OF L3/4 WERE REMOVED, AND SCREWS WERE ADDED TO L2 AND L5. TLIF TREATING L2/3 AND TLIF TREATING L4/5 WERE DONE. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. THERE IS NO PROBLEM IN THE FUSION SITE TREATED IN THE INITIAL SURGERY IN 2024. THIS REPORT INVOLVES ONE VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 40MM. THIS IS REPORT 1 OF 7 FOR (B)(4).
Patient problems: Osteolysis
MAUDE report 16644033 ↗InjurySep 12, 2022Device Slipped; Device Slipped
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A PLIF PROCEDURES (L4/5) FOR TREATING ASD ON (B)(6) 2022. THE EVENT TOOK PLACE IN SEQUENCE AS FOLLOWS: THERE WAS A POSSIBILITY THAT THE SCREW 7.0×40 (UNK), THAT INSERTED TO L5 IN THIS SURGERY WAS LOOSE, SO IT WAS REMOVED. THE SURGEON TRIED TO REPLACE IT WITH THE SCREW 8.0×40 AND TRIED TO INSERT IT BUT WAS UNABLE TO INSERT IT DUE TO RESISTANCE. THE TIP OF THE SCREWDRIVER IN QUESTION WAS BROKEN. THE TIP OF THE SCREWDRIVER GOT CAUGHT IN THE CORE OF THE SCREW 8.0×40, AND THE SURGEON HAD DIFFICULTY REMOVING THE SCREW 8.0×40, BUT HE MANAGED TO DO SO AND THE REMOVAL WAS COMPLETED. THE SURGEON COMMENTED THAT THE REASON FOR THE DIFFICULTY IN INSERTING THE SCREW 8.0×40¿WAS THAT THE PATIENT'S BONE WAS QUITE HARD, AND THAT WHEN HE DID NOT TAP THE SCREW 8.0×40 TO BREAK THROUGH ALTHOUGH HE FELT RESISTANCE DURING INSERTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) 5.5 TI CORT FIX 7X40MM. THIS IS REPORT 3 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15401513 ↗MalfunctionAug 19, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS AN UNKNOWN PROCEDURE PERFORMED ON (B)(6) 2022. THE PATIENT WAS X-RAYED AFTER SCREW INSERTION. IT WAS CONFIRMED THAT THE SCREW IN QUESTION HAD COME OFF THE LESION. WHEN THEY TRIED TO RETRACT THE SCREW WITH THE DRIVER SHAFT, THE TIP OF THE DRIVER SHAFT BROKE OFF. A FRAGMENT WHICH HAD STUCK TO THE SCREW WAS REMOVED FROM THE LESION. THE PROCEDURE WAS COMPLETED LESS THAN 30-MINUTE SURGICAL DELAY. THE PATIENT¿S BONE WAS QUITE HARD, SO THE SURGEON HAD DIFFICULTY IN THE SCREW INSERTION. PATIENT STATUS/ OUTCOME: THE PATIENT WAS X-RAYED, AND IT WAS CONFIRMED THAT NO FRAGMENT HAD BEEN LEFT IN THE LESION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1)5.5 TI CORT FIX 7X40MM. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15262386 ↗MalfunctionJul 22, 2022Migration; Migration
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A PLIF (L4-5) FOR TREATING SPONDYLOLISTHESIS ON (B)(6) 2022 THE 7.0X 40MM SCREW (UNK) CUT OUT DURING SCREW TIGHTENING (WIPER MOTION) WITH THE TORQUE WRENCH IN QUESTION. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT (7.0X 45MM SCREW) LESS THAN 30-MINUTE SURGICAL DELAY. HE COMMENTED THAT THE EVENT MIGHT HAVE BEEN TRIGGERED BY MALFUNCTIONING TORQUE-VALUE SETTING ON THE TORQUE WRENCH. IT WAS BECAUSE AN EXCESSIVE LOAD HAD BEEN UNEXPECTEDLY PUT ONTO THE SCREW. WHEN THE SURGEON WAS APPLYING THE REPLACEMENT, HE ALSO FELT STRONG LOAD EVEN BEFORE TORQUE CAME INTO ACTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) 5.5 TI CORT FIX 7X40MM. THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15085949 ↗InjuryFeb 23, 2022Break; Break
IT WAS REPORTED THAT ON (B)(6) 2022, PATIENT UNDERWENT REVISION SURGERY DUE TO TWO BROKEN EXPEDIUM CORTICAL SCREWS AT THE S1 LEVEL. PATIENT AND PROCEDURE OUTCOME WERE UNKNOWN. PATIENT HAD A PREVIOUS L3-S1 FUSION, WHICH WAS EXTENDED UP TO L2 AND ADDED SAI AFTER REPLACEMENT OF THE BROKEN SCREWS. PROCEDURE OUTCOME AND PATIENT STATUS ARE UNKNOWN. THIS REPORT IS FOR AN UNKNOWN EXPEDIUM SCREW. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13584861 ↗InjuryFeb 23, 2022Break; Break
IT WAS REPORTED THAT ON (B)(6) 2022, PATIENT UNDERWENT REVISION SURGERY DUE TO TWO BROKEN EXPEDIUM CORTICAL SCREWS AT THE S1 LEVEL. PATIENT AND PROCEDURE OUTCOME WERE UNKNOWN. PATIENT HAD A PREVIOUS L3-S1 FUSION, WHICH WAS EXTENDED UP TO L2 AND ADDED SAI AFTER REPLACEMENT OF THE BROKEN SCREWS. PROCEDURE OUTCOME AND PATIENT STATUS ARE UNKNOWN. THIS REPORT IS FOR AN UNKNOWN EXPEDIUM SCREW. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13584849 ↗InjuryFeb 16, 2022Device Slipped; Device Slipped
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT UNDERWENT A REMOVAL PROCEDURE DUE TO SEVERE PUS IN THE LESION. DURING THE REMOVAL PROCEDURE THE SETSCREW CHIPPED. THE SURGEON REMOVED TWO FRAGMENTS FROM THE PATIENT¿S LESION. NO MORE FRAGMENTS WERE LEFT IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY. THE PATIENT RECEIVED THE IMPLANTS ON (B)(6) 2015 AND THE BONE FUSION HAD NOT BEEN ATTAINED AT THE TIME OF THE REMOVAL SURGERY. THIS REPORT INVOLVES ONE (1) VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 40MM. THIS IS REPORT 4 OF 5 FOR (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture; Purulent Discharge
MAUDE report 13543387 ↗InjuryDec 8, 2021Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT, PROCEDURE WAS REVISED DUE TO BACK PAIN AND CONNECTION DEGENERATION. THIS REPORT IS FOR (1) 5.5 TI CORT FIX 7X40MM. THIS IS REPORT 5 OF 7 FOR COMPLAINT (B)(4).
Patient problems: Osteolysis; Pain
MAUDE report 12961911 ↗InjuryAug 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT HAD POSTERIOR SPINAL FUSION SURGERY DONE ON (B)(6) 2021. THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2021 AS THE PATIENT WAS HAVING COMPLICATION FROM A RIGHT SIDE S1 PEDICLE SCREW. THE PATIENT WAS EXPERIENCING RIGHT SIDE LEG PAIN. THE SCREW WAS REMOVED AND NEW IMPLANTS WERE PUT IN. PROCEDURE OUTCOME IS UNKNOWN. THIS REPORT IS FOR ONE (1) VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 40MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Pain
MAUDE report 12378820 ↗MalfunctionJul 28, 2021Material Twisted/Bent; Material Twisted/Bent
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR A PLIF AT L5 TREATING SPINAL CANAL STENOSIS. DURING THE SURGERY, THE SCREWDRIVER WAS NOT ABLE TO BE ATTACHED TO THE SCREW. SOME DEFORMITY WAS FOUND INSIDE THE SCREWHEAD. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT LESS THAN 30-MINUTE SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR (1) UNK ¿ PLATES THIS REPORT IS 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12242573 ↗MalfunctionJul 19, 2021Break; Entrapment of Device; Break; Entrapment of Device
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT PATIENT UNDERWENT REVISION SURGERY ON (B)(6) 2021, DUE TO A BROKEN SCREW. AS IT WAS DIFFICULT TO REMOVE THE FRAGMENT, THE SURGEON LEFT IT IN THE PATIENT¿S BODY. BONE FUSION HAD BEEN OBTAINED. PROCEDURE WAS COMPLETED WITHOUT DELAY. THIS REPORT IS FOR A 5.5 TITANIUM (TI) CORTICAL FIX 7X40MM SCREW. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 12192870 ↗InjuryApr 9, 2018Break; Break
ON (B)(6) 2014, THE SPINAL SURGERY WAS PERFORMED. THE FIXED ARE WAS L5-S. ON (B)(6) 2018, THE REOPERATION WAS PERFORMED. DURING THE PRE-OPERATIVE MEETING, THE SURGEON PLANNED THAT THE S1 IMPLANTS WOULD BE REPLACED WITH L4-5 VERSE IMPLANTS AFTER THE S1 IMPLANTS WOULD BE REMOVED. WHEN THE SURGEON CHECKED THE X-RAY IMAGES DURING THE PRE-OPERATIVE MEETING, NO BROKEN IMPLANTS WERE FOUND, THE OPERATION WAS PERFORMED AFTER THE MEETING ON THE SAME DAY (ON (B)(6) 2018). AS A USUAL PROCEDURE, THE EXPEDIUM L5-S SET SCREWS AND RODS WERE REMOVED. WHEN THE SURGEON THEN TRIED TO REMOVE THE REPORTED CORTICAL SCREWS, IT WAS FOUND THAT THE REPORTED CORTEX SCREWS WERE BROKEN. THE REPORTED CORTICAL SCREWS ON BOTH RIGHT S1 AND LEFT S1 WERE AFFECTED. THE SURGERY WAS COMPLETED WITH A 60-MINUTE-DELAY, AND THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. ((B)(4)).
Patient problems: Injury; Not Applicable
MAUDE report 7408274 ↗InjuryApr 9, 2018Break; Break
ON (B)(6) 2014, THE SPINAL SURGERY WAS PERFORMED. THE FIXED ARE WAS L5-S. ON (B)(6) 2018, THE REOPERATION WAS PERFORMED. DURING THE PRE-OPERATIVE MEETING, THE SURGEON PLANNED THAT THE S1 IMPLANTS WOULD BE REPLACED WITH L4-5 VERSE IMPLANTS AFTER THE S1 IMPLANTS WOULD BE REMOVED. WHEN THE SURGEON CHECKED THE X-RAY IMAGES DURING THE PRE-OPERATIVE MEETING, NO BROKEN IMPLANTS WERE FOUND, THE OPERATION WAS PERFORMED AFTER THE MEETING ON THE SAME DAY (ON (B)(6) 2018). AS AN USUAL PROCEDURE, THE EXPEDIUM L5-S SET SCREWS AND RODS WERE REMOVED. WHEN THE SURGEON THEN TRIED TO REMOVE THE REPORTED CORTICAL SCREWS, IT WAS FOUND THAT THE REPORTED CORTEX SCREWS WERE BROKEN. THE REPORTED CORTICAL SCREWS ON BOTH RIGHT S1 AND LEFT S1 WERE AFFECTED. THE SURGERY WAS COMPLETED WITH A 60-MINUTE-DELAY, AND THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. (THIS (B)(4) IS RELATED TO (B)(4).
Patient problems: Injury; Not Applicable
MAUDE report 7408163 ↗
FDA clearance and classification
Cleared through Special 510(k) K110216 on Apr 1, 2011 (decision: Substantially Equivalent). Submitted by Depuy Spine, Inc. as “Expedium System, Viper System”. The FDA review took 66 days. FDA record ↗
The FDA regulates product code MNH (orthosis, spondylolisthesis spinal fixation) under 21 CFR 888.3070 as a Class II device. 21 CFR 888.3070 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VIPER 186731740
Has the VIPER 186731740 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705034352008), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the VIPER 186731740?
The FDA's MAUDE database lists 22 adverse event reports for this device, received between Apr 9, 2018 and Jan 9, 2024. In those reports, the most common specific device problems are “Break” (6), “Device Slipped” (3) and “Entrapment of Device” (1); the most common patient problems are “Osteolysis” (8), “Injury” (2) and “Not Applicable” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VIPER 186731740?
None of the reports are classified as deaths, 17 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the VIPER 186731740 FDA approved?
It was cleared through Special 510(k) K110216 on Apr 1, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the VIPER 186731740 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the VIPER 186731740, and is it still sold?
The FDA's device database lists Medos International Sàrl as the company responsible for it (brand VIPER, model 186731740). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Orthosis, spondylolisthesis spinal fixation devices
- CD HORIZON® Spinal System 7540020Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation257 reports
- MANTIS Redux 48289999Stryker Corporation · Orthosis, spondylolisthesis spinal fixation76 reports
- EXPEDIUM TOP NOTCH 179771555Depuy Spine, LLC · Orthosis, spondylolisthesis spinal fixation41 reports
- EXPEDIUM 179712650Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation31 reports
- XIA 4 5 48130000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation29 reports
- EXPEDIUM VERSE 199721530Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4.5 System - XIA CT 48850000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation27 reports
- CD HORIZON® Spinal System 7576545Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation26 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page