Skip to content

NA 698335325Recalls & reported problems

Depuy Spine, LLC · MODIFIED INSTRUMENT BENDER TIP, A 5 DEGREES · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034371061), received between Nov 2, 2022 and Jun 28, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 28, 2024Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED DURING SURGERY ON (B)(6) 2024, THE TIP OF INSTRUMENT BROKE DURING SURGERY. IT IS A "ROD BENDER" TIP AND IT BROKE DURING THE ROD BENDING PROCESS. NO INCIDENT OR FRAGMENTS REPORTED. NO SURGICAL DELAY. THE TIP OF A SIMILAR INSTRUMENT IS ALSO BENT OUT OF SHAPE FOLLOWING REPEATED USE OVER THE YEARS. ANOTHER "TIP" WAS USED. THE BROKEN/DAMAGED ONES ARE 5 DEGREE TIPS, SO SURGERY WAS COMPLETED USING 10 DEGREE TIPS. PROCEDURE WAS COMPLETED SUCCESSFULLY. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19637689 ↗
  • MalfunctionNov 2, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2022, THE SURGEON WAS CORRECTING SCOLIOSIS WITH THE CUSTOM CORONAL BENDERS FROM THE EXPEDIUM 5.5 SYSTEM, AND ONE OF THE PIECES BROKE INSIDE THE PATIENT. THE BROKEN PIECE WAS RETRIEVED AND THE PATIENT WAS NOT IMPACTED. THE SURGERY WAS COMPLETED IN A TIMELY MANNER AND NO DELAYS OCCURRED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE MODIFIED INSTRUMENT BENDER TIP, A 5 DEGREES. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15714499 ↗

Other BENDER devices

All BENDER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page