Skip to content

EXPEDIUM VERSE 199721870Recalls & reported problems

Medos International Sàrl · EXPEDIUM VERSE SPINE SYSTEM POLYAXIAL SCREW 5.5 8.0 x 70mm · Class 2 · Implant

All models of EXPEDIUM VERSE (90)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034441900), received between Jul 13, 2022 and Jul 13, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 13, 2022Material Twisted/Bent; Material Twisted/Bent

    DEVICE REPORT FROM GERMANY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2022, THE SURGEON HAD PLACED ALL SCREWS, INSERTED A TI ROD AND COCR ROD, AND BEGAN INSERTION OF THE CORRECTION CLOSER SCREWS FROM THE CAUDAL SIDE. ON THE SIDE WITH THE COCR ROD, THE CORRECTION SCREW PLUGS JUMPED IN THE THREAD ON THE 2ND EXP VERSE SCREW, DAMAGING THE THREAD IN THE TULIP OF THE SCREW. THE SAME EVENT OCCURRED WITH THE 4TH SCREW, BOTH SCREWS REQUIRED REPLACEMENT. THE ROD HAD TO BE REMOVED IN ORDER TO REPLACE THE SCREWS. AFTER THIS EVENT, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH REPLACEMENT PRODUCTS. THERE WAS A SURGICAL DELAY OF 30-40 MINUTES. PATIENT OUTCOME IS STABLE. THIS REPORT IS FOR A EXPEDIUM VERSE SPINE SYSTEM POLYAXIAL SCREW 5.5 8.0 X 70MM. THIS IS REPORT 1 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15004569 ↗

Other Orthosis, spondylolisthesis spinal fixation devices

All Orthosis, spondylolisthesis spinal fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page