EXPEDIUM 299704145Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM PENCIL HANDLE · Class 1
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10705034467245), received between Nov 8, 2023 and Nov 10, 2023 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Contamination /Decontamination Problem7
Patient problems
- Insufficient Information7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED THIS REPORT IS FOR ONE (1) PENCIL HANDLE. THIS IS REPORT 4 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18110325 ↗MalfunctionNov 9, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED. THIS REPORT IS FOR ONE (1)PENCIL HANDLE THIS IS REPORT 5 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18099480 ↗MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED THIS REPORT IS FOR ONE (1)PENCIL HANDLE. THIS IS REPORT 4 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18096884 ↗MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED THIS REPORT IS FOR ONE (1)PENCIL HANDLE THIS IS REPORT 8 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18096881 ↗MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED. THIS REPORT IS FOR ONE (1)PENCIL HANDLE. THIS IS REPORT 9 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18096879 ↗MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED THIS REPORT IS FOR ONE (1)PENCIL HANDLE THIS IS REPORT 7 OF 10 FOR COMPLAINT (B)(4)
Patient problems: Insufficient Information
MAUDE report 18096878 ↗MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED ON WEDNESDAY, (B)(6) 2023, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED. THIS REPORT IS FOR ONE (1)PENCIL HANDLE THIS IS REPORT 6 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18096877 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page