EXPEDIUM 299704145Recalls & reported problems

Medos International Sàrl · EXPEDIUM SPINE SYSTEM PENCIL HANDLE · Class 1

All models of EXPEDIUM (291)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10705034467245), received between Nov 8, 2023 and Nov 10, 2023 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 7 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem7

Patient problems

  • Insufficient Information7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED THIS REPORT IS FOR ONE (1) PENCIL HANDLE. THIS IS REPORT 4 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18110325 ↗
  • MalfunctionNov 9, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED. THIS REPORT IS FOR ONE (1)PENCIL HANDLE THIS IS REPORT 5 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18099480 ↗
  • MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED THIS REPORT IS FOR ONE (1)PENCIL HANDLE. THIS IS REPORT 4 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18096884 ↗
  • MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED THIS REPORT IS FOR ONE (1)PENCIL HANDLE THIS IS REPORT 8 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18096881 ↗
  • MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED. THIS REPORT IS FOR ONE (1)PENCIL HANDLE. THIS IS REPORT 9 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18096879 ↗
  • MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED THIS REPORT IS FOR ONE (1)PENCIL HANDLE THIS IS REPORT 7 OF 10 FOR COMPLAINT (B)(4)

    Patient problems: Insufficient Information

    MAUDE report 18096878 ↗
  • MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED ON WEDNESDAY, (B)(6) 2023, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED. THIS REPORT IS FOR ONE (1)PENCIL HANDLE THIS IS REPORT 6 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18096877 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page