EXPEDIUM 299704140Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM TEAR-DROP HANDLE, RATCHETING · Class 1
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034467290), received between Aug 6, 2021 and Jul 16, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Contamination /Decontamination Problem2
- Device-Device Incompatibility2
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2024, THE DRIVER SHAFT WAS PLACED IN THE HANDLE AND IT COULD NOT BE LOCKED. THERE WAS NO CONSEQUENCE TO THE PATIENT. THE SURGERY WAS PERFORMED AS PLANNED. THE OTHER HANDLE IN THE SYSTEM WAS USED WITH NO DELAY. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM TEAR-DROP HANDLE, RATCHETING. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19754700 ↗MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED THIS REPORT IS FOR ONE (1) TEAR-DROP HANDLE, RATCHETING THIS IS REPORT 2 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18096890 ↗MalfunctionNov 8, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED ON WEDNESDAY, OCTOBER 18TH, SETS WERE DROPPED OFF FOR CASES ON FRIDAY, OCTOBER 20TH. UPON RECEIPT OF THESE SETS, THERE WERE MULTIPLE INSTRUMENTS THAT WERE DUBBED UNFIT TO BE USED IN SURGERY (PART NUMBERS OF EACH INSTRUMENT HAVE BEEN INCLUDED). SPD RAN A BOROSCOPE DOWN EACH OF THE CANNULATED INSTRUMENTS (WHICH WE AS REPS DO NOT HAVE ACCESS TO AT ANY OF OUR ACCOUNTS TO CHECK THESE OURSELVES), AND FOUND MULTIPLE INSTRUMENTS THAT FELL OUTSIDE OF THEIR GUIDELINES FOR SAFE USE IN SURGERY. SUBSEQUENTLY, I HAD TO RUN TO MULTIPLE DIFFERENT HOSPITALS AROUND DENVER TO BE ABLE TO MAKE THESE SETS COMPLETE AND USABLE. THIS PUT ME OUTSIDE OF THE GUIDELINES PUT IN PLACE BY THE VA IN TERMS OF TIMELINESS ON DROPPING OFF THE TRAYS. EVENT OCCURRED DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WERE PATIENT OUTCOMES OR CONSEQUENCES. ONE OF THE TWO CASES SCHEDULED HAD TO BE CANCELLED. THIS REPORT IS FOR ONE (1) TEAR-DROP HANDLE, RATCHETING. THIS IS REPORT 3 OF 10 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18096872 ↗MalfunctionAug 6, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, IT WAS FOUND THAT THE TEAR DROP HANDLE WAS UNABLE TO BE ATTACHED TO AN UNKNOWN INSERTER. THE ISSUE WAS DISCOVERED PREOPERATIVELY. THE SPINO-PELVIC FUSION WAS COMPLETED ON (B)(6) 2021 USING A REPLACEMENT DEVICE. THERE WAS NO SURGICAL DELAY. THIS REPORT IS FOR ONE (1) TEAR-DROP HANDLE, RATCHETING. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12287963 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page