EXPEDIUM VERSE 299704220Recalls & reported problems

Medos International Sàrl · EXPEDIUM VERSE SPINE SYSTEM CORRECTION KEY INSERTER/TIGHTENER · Class 1

All models of EXPEDIUM VERSE (90)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Medos EXPEDIUM VERSE spine system correction key inserter/tightener is a surgical tool used to tighten and lock spinal implant parts. Reports mostly describe it wearing out during surgery, with frayed tips, a broken prong, or threads jumping so it would not load and assemble correctly, sometimes delaying or replacing the instrument mid-operation. There are 56 reports, no deaths and 2 injuries, and no recalls name this device. Reports came in 9 times in the last 12 months after none in the previous 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
56
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 56 adverse event reports that list this exact device (DI 10705034467498), received between Oct 24, 2017 and Apr 15, 2026 — classified as 2 injuries, 54 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (42 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 3 reportsJan 2024: 0 reportsFeb 2024: 12 reportsMar 2024: 5 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 8 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent42
  • Break8
  • Material Deformation4
  • Device-Device Incompatibility2
  • Contamination /Decontamination Problem1
  • Improper or Incorrect Procedure or Method1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions48
  • No Consequences Or Impact To Patient3
  • Injury2
  • Not Applicable2
  • Failure of Implant1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 56 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 15, 2026Break

    IT WAS REPORTED THAT VERSE KEY INSERTER TIGHTENER AND VERSE X25 INSERTER/TIGHTENER WERE INVOLVED IN AN EVENT. THE REPORTED FAILURES INCLUDED A BROKEN PRONG AT THE END OF THE VERSE KEY INSERTER TIGHTENER AND WORN TIPS ON THE VERSE X25 INSERTER/TIGHTENER, RESULTING IN INABILITY TO LOAD CORRECTLY AND ASSEMBLE MATING DEVICES. THE EVENT OCCURRED INTRAOPERATIVELY, CAUSING MINIMAL DELAY TO SURGERY, BUT THERE WERE NO ADVERSE INCIDENTS OR PATIENT CONSEQUENCES. ADDITIONAL INSTRUMENTS WERE USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24887979 ↗
  • MalfunctionDec 5, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE DEVICES WERE WORN OUT/THREADED JUMP INTRAOPERATIVELY AND ARE BEING REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23729941 ↗
  • MalfunctionDec 5, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE DEVICES WERE WORN OUT/THREADED JUMP INTRAOPERATIVELY AND ARE BEING REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23729733 ↗
  • MalfunctionDec 5, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE DEVICES WERE WORN OUT/THREADED JUMP INTRAOPERATIVELY AND ARE BEING REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23729730 ↗
  • MalfunctionDec 5, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE DEVICES WERE WORN OUT/THREADED JUMP INTRAOPERATIVELY AND ARE BEING REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23729647 ↗
  • MalfunctionDec 4, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED INTRA-OPERATIVELY THE INSTRUMENTS WERE DISCOVERED TO BE WORN OUT; THE THREADS JUMPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23718263 ↗
  • MalfunctionDec 4, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED INTRA-OPERATIVELY THE INSTRUMENTS WERE DISCOVERED TO BE WORN OUT; THE THREADS JUMPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23718170 ↗
  • MalfunctionDec 4, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED INTRA-OPERATIVELY THE INSTRUMENTS WERE DISCOVERED TO BE WORN OUT; THE THREADS JUMPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23717962 ↗
  • MalfunctionDec 4, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED INTRA-OPERATIVELY THE INSTRUMENTS WERE DISCOVERED TO BE WORN OUT; THE THREADS JUMPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23717938 ↗
  • MalfunctionApr 18, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2024 THE DRIVERS STRIPPED DURING FINAL TIGHTENING OF SET SCREW. TLIF INSERTERS BENT FROM OVER TIGHTENING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE VERSE KEY INSERTER TIGHTENER THIS IS REPORT 6 OF 6 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19139755 ↗
  • MalfunctionApr 18, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON MARCH 1, 2024 THE DRIVERS STRIPPED DURING FINAL TIGHTENING OF SET SCREW. TLIF INSERTERS BENT FROM OVER TIGHTENING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE VERSE KEY INSERTER TIGHTENER. THIS IS REPORT 2 OF 6 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19139654 ↗
  • MalfunctionApr 18, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2024 THE DRIVERS STRIPPED DURING FINAL TIGHTENING OF SET SCREW. TLIF INSERTERS BENT FROM OVER TIGHTENING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE VERSE KEY INSERTER TIGHTENER. THIS IS REPORT 1 OF 6 FOR COMPLAINT(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19139588 ↗
  • MalfunctionMar 1, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2024, THE TORQUE DRIVERS IN QUESTION WERE FOUND TO HAVE STRIPPED. THERE WAS NO REPORTED PATIENT OR PROCEDURE INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN EXPEDIUM VERSE SPINE SYSTEM CORRECTION KEY INSERTER/TIGHTENER. THIS IS REPORT 5 OF 5 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18818055 ↗
  • MalfunctionMar 1, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2024, THE TORQUE DRIVERS IN QUESTION WERE FOUND TO HAVE STRIPPED. THERE WAS NO REPORTED PATIENT OR PROCEDURE INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN EXPEDIUM VERSE SPINE SYSTEM CORRECTION KEY INSERTER/TIGHTENER. THIS IS REPORT 3 OF 5 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18818054 ↗
  • MalfunctionMar 1, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2024, THE TORQUE DRIVERS IN QUESTION WERE FOUND TO HAVE STRIPPED. THERE WAS NO REPORTED PATIENT OR PROCEDURE INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN EXPEDIUM VERSE SPINE SYSTEM CORRECTION KEY INSERTER/TIGHTENER. THIS IS REPORT 4 OF 5 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18818053 ↗
  • MalfunctionMar 1, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2024, THE TORQUE DRIVERS IN QUESTION WERE FOUND TO HAVE STRIPPED. THERE WAS NO REPORTED PATIENT OR PROCEDURE INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN EXPEDIUM VERSE SPINE SYSTEM CORRECTION KEY INSERTER/TIGHTENER. THIS IS REPORT 2 OF 5 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18818052 ↗
  • MalfunctionMar 1, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2024, THE TORQUE DRIVERS IN QUESTION WERE FOUND TO HAVE STRIPPED. THERE WAS NO REPORTED PATIENT OR PROCEDURE INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN EXPEDIUM VERSE SPINE SYSTEM CORRECTION KEY INSERTER/TIGHTENER. THIS IS REPORT 1 OF 5 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18818046 ↗
  • MalfunctionFeb 29, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2024, THE REPORTER NOTIFIED THAT THEY WERE RUNNING OUT OF TORQUE DRIVERS DUE TO THEM BEING STRIPPED. THERE WAS NO PATIENT INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) VERSE KEY INSERTER TIGHTENER. THIS IS REPORT 10 OF 10 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18812570 ↗
  • MalfunctionFeb 29, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2024, THE REPORTER NOTIFIED THAT THEY WERE RUNNING OUT OF TORQUE DRIVERS DUE TO THEM BEING STRIPPED. THERE WAS NO PATIENT INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) VERSE KEY INSERTER TIGHTENER THIS IS REPORT 9 OF 10 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18812557 ↗
  • MalfunctionFeb 29, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2024, THE REPORTER NOTIFIED THAT THEY WERE RUNNING OUT OF TORQUE DRIVERS DUE TO THEM BEING STRIPPED. THERE WAS NO PATIENT INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) VERSE KEY INSERTER TIGHTENER THIS IS REPORT 8 OF 10 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18812525 ↗
  • MalfunctionFeb 29, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2024, THE REPORTER NOTIFIED THAT THEY WERE RUNNING OUT OF TORQUE DRIVERS DUE TO THEM BEING STRIPPED. THERE WAS NO PATIENT INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) VERSE KEY INSERTER TIGHTENER. THIS IS REPORT 7 OF 10 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18812497 ↗
  • MalfunctionFeb 29, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2024, THE REPORTER NOTIFIED THAT THEY WERE RUNNING OUT OF TORQUE DRIVERS DUE TO THEM BEING STRIPPED. THERE WAS NO PATIENT INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) VERSE KEY INSERTER TIGHTENER. THIS IS REPORT 6 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18812475 ↗
  • MalfunctionFeb 29, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6)2024, THE REPORTER NOTIFIED THAT THEY WERE RUNNING OUT OF TORQUE DRIVERS DUE TO THEM BEING STRIPPED. THERE WAS NO PATIENT INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) VERSE KEY INSERTER TIGHTENER THIS IS REPORT 5 OF 10 FOR COMPLAINT (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18812464 ↗
  • MalfunctionFeb 29, 2024Material Twisted/Bent; Material Twisted/Bent

    THIS IS REPORT 2 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18812433 ↗
  • MalfunctionFeb 29, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON FEBRUARY 14 2024, THE REPORTER NOTIFIED THAT THEY WERE RUNNING OUT OF TORQUE DRIVERS DUE TO THEM BEING STRIPPED. THERE WAS NO PATIENT INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) VERSE KEY INSERTER TIGHTENER THIS IS REPORT 3 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18812282 ↗

Other DRIVER, PROSTHESIS devices

All DRIVER, PROSTHESIS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page