NA 163005530Recalls & reported problems
Medos International Sàrl · CONCORDE Clear MIS DISCECTOMY DEVICE 5MM, 30 DEGREES · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034507378), received between Jun 14, 2018 and Mar 27, 2020 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break4
Patient problems
- No Consequences Or Impact To Patient4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 27, 2020Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN MEXICO AS FOLLOWS: IT WAS REPORTED THAT WHEN USED IN A PATIENT, IN DISC SPACE L4 L5, DURING SUCTION, IT WAS CONNECTED TO THE ASPIRATION TO PROCEED TO THE EXTRACTION OF THE NUCLEUS PULPOSUS DISK, IT WAS BROKEN IN THE VISOR OF THE UNION OF THE DIAMETERS OF THE PRODUCT. HOWEVER, IT COULD BE STILL USED IN SURGERY. THERE WERE NO OPERATIVE COMPLICATIONS OR RISK TO THE PATIENT. WE DO NOT HAVE THE DEVICE SINCE HAVING BEEN IN CONTACT WITH A PATIENT, THE DEVICE IS CONSIDERED CONTAMINATED AND MUST BE REMOVED BY BIOLOGICAL CONTROLS. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9892098 ↗MalfunctionDec 2, 2019Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT A PATIENT UNDERWENT A POSTERIOR LUMBAR INTERBODY FUSION (PLIF) PROCEDURE TO TREAT THE LUMBAR SPINAL CANAL STENOSIS (L4-5) ON (B)(6) 2019. DURING THE PROCEDURE THE SURGEON REMOVED DEGENERATED DISC WITH THE CONCORDE CLEAR 30 DEGREES IN QUESTION FROM THE RIGHT SIDE (SURGEON¿S SIDE) WITHOUT ISSUE. NEXT, THE SURGEON INSERTED IT FROM THE LEFT SIDE (SURGEON¿S SIDE). WHEN SURGEON STARTED DISC REMOVAL, IT BROKE WHERE THE TUBING¿S DIAMETER BECOMES LARGER TO 8MM (80MM FROM THE TIP GUARD). IT WAS CONFIRMED BY POSTOPERATIVE X-RAYS THAT NO FRAGMENT WAS LEFT IN THE PATIENT¿S BODY. PROCEDURE WAS COMPLETED SUCCESSFULLY. PATIENT'S OUTCOME IS REPORTED AS STABLE. NO FURTHER INFORMATION IS AVAILABLE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9396474 ↗MalfunctionDec 19, 2018Break; Break
IT WAS REPORTED THAT THE CONCORDE CLEAR TIP BROKE OFF. PATIENT CONSEQUENCE? :NO. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? : YES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8180811 ↗MalfunctionJun 14, 2018Break; Break
WHEN INTRODUCING THE DEVICE IN DISC SPACE, PERFORMING DISCECTOMY, IN A USUAL MANEUVER, THE DEVICE BROKE. I HAVE PHOTOS AND VIDEO! PATIENT CONSEQUENCE?: NO. ACTION TAKEN FOR PROCEDURE: PERFORMED THE REST OF PROCEDURE WITH TYPICAL INSTRUMENTS. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT?: YES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7602744 ↗
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page