Skip to content

VIPER 286705340Recalls & reported problems

Medos International Sàrl · VIPER SYSTEM NITINOL GUIDEWIRE, LONG AND DISPOSABLE 1.45mm · Class 1

All models of VIPER (121)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034507873), received between Sep 23, 2019 and Nov 11, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Entrapment of Device1

Patient problems

  • Foreign Body In Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 11, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT THE ITEM BROKE OFF INSIDE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 23526527 ↗
  • InjurySep 23, 2019Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2019, DURING SPINAL FUSION AND PEDICLE PREPARATION FOR LEFT S1 SCREW INSERTION. THE K-WIRE KINKED DURING SCREW INSERTION. REPORTEDLY, NITINOL K-WIRE WAS PLACED INTO THE PEDICLE PATHWAY. AN 8 X 45 VIPER CORTICAL FIX SCREW AND CANNULATED SCREW DRIVER WERE THEN PASSED OVER THE NITINOL K-WIRE. WHILE REMOVING THE K-WIRE THROUGH THE DISTAL PORTION OF THE SCREW DRIVER ASSEMBLY, THE SERRATED/ETCHED TIP (APPROXIMATELY 15MM) OF THE K-WIRE SNAPPED OFF INSITU, PAST THE ANTERIOR VORTEX OF THE S1 VERTEBRAL BODY. THE FRAGMENT WAS NOT RETRIEVABLE AND FLUOROSCOPY WAS TAKEN TO CONFIRM. SURGICAL DELAY OF 60 MINUTES. FRAGMENTS WAS GENERATED BUT UNRETRIEVABLE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED AND IT IS UNKNOWN IF THERE WAS NO PATIENT CONSEQUENCES. CONCOMITANT DEVICES REPORTED: VIPER CORTICAL FIX SCREW , (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# UNKNOWN); CANNULATED SCREWDRIVER, (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY# UNKNOWN). THIS COMPLAINT INVOLVES ONE (1) DEVICE.

    Patient problems: Foreign Body In Patient

    MAUDE report 9106519 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates