Skip to content

SYMPHONY 1020-40-114Recalls & reported problems

Medos International Sàrl · SYMPHONY OCT SYSTEM POLYAXIAL SCREW diameter 4.0 Rod, 4.0X14mm · Class 2 · Implant

All models of SYMPHONY (113)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034512242), received between Sep 23, 2020 and Apr 17, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2

Patient problems

  • Injury1
  • Insufficient Information1
  • No Code Available1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 17, 2024Break; Break

    PATIENT OUTCOME IS REPORTED AS FINE.

    Patient problems: Insufficient Information

    MAUDE report 19123761 ↗
  • InjurySep 23, 2020Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2020, WHILE PLACING THE C7 SCREW, THE SURGEON TAPPED WITH A 4.0 TAP AND AS HE WAS DOING THE FINAL TURN ON THE 4.0X14 SCREW THE SCREW SHEARED OFF BELOW THE POLY HEAD. THE SHANK WAS RECOVERED AND THEY SKIPPED THAT LEVEL. PATIENT HAD GIANT CELL NEOPLASM WITH A SOLID ANEURYSMAL BONE CYST. HE WAS TREATED WITH DENOSUMAB WHICH MAKES THE CYST HARD. THE BONE WAS EXTREMELY HARD AT THIS LEVEL. POSTERIOR CERVICAL C6-T2. THE SCREW SHANK WAS DRILLED OUT AND RECOVERED. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A TEN (10) MINUTE SURGICAL DELAY. PATIENT OUTCOME IS UNKNOWN. CONCOMITANT DEVICE REPORTED: SCREWDRIVER (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN); DRILL (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) 4.0 PLY SCRW 4.0X14. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Injury; No Code Available

    MAUDE report 10575044 ↗

Other Posterior cervical screw system devices

All Posterior cervical screw system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page