Skip to content

SYMPHONY 1020-40-328SRecalls & reported problems

Medos International Sàrl · SYMPHONY OCT SYSTEM POLYAXIAL SCREW REDUCTION CANNULATED Diameter 4.0 Rod, 4.0X28mm 3.5 4.0 · Class 2 · Implant

All models of SYMPHONY (113)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034520872), received between Dec 6, 2022 and Dec 6, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Deformation1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 6, 2022Material Deformation; Scratched Material; Material Deformation; Scratched Material

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A POSTERIOR CERVICAL FUSION (C2 RIGHT) FOR A C2 FRACTURE ON (B)(6) 2022. THE SURGEON TRIED TO REMOVE THE TAB BEFORE INSERTION BECAUSE THE TAB WAS INTERFERED. WHEN THE SURGEON TRIED TO REMOVE THE TAB OF THE TAB SCREW WITH A TAB BREAKER, THE SCREW MOVED WITH THE TAB BREAKER, AND THE TAB COULD NOT BE REMOVED. THE SURGEON PUT THE SCREW IN THE HOLE BEHIND THE COUNTER OF THE H2H AND REMOVED THE TAB. THEN, THE SCREW THREADS WERE SCRATCHED AND BENT AT THE SCREW NECK. A NEW SCREW WAS USED INSTEAD OF THE TAB SCREW IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE SYMPHONY OCT SYSTEM POLYAXIAL SCREW REDUCTION CANNULATED DIAMETER 4.0 ROD, 4.0X28MM 3.5 4.0. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15924248 ↗