SYMPHONY 1020-35-312SRecalls & reported problems
Medos International Sàrl · SYMPHONY OCT SYSTEM POLYAXIAL SCREW REDUCTION CANNULATED Diameter 4.0 Rod, 3.5X12mm 3.5 4.0 · Class 2 · Implant
All models of SYMPHONY (113)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034527796), received between Jun 20, 2024 and Jun 20, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 20, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PATIENT UNDERWENT A C5-C6 POSTERIOR FUSION USING THE SYMPHONY POSTERIOR CERVICAL SYSTEM. THE SURGEON OPTED TO USE THE LONG TAB SCREWS. THE PROCEDURE WENT SMOOTHLY, UNTIL PART OF THE FINAL STEP WHERE THE TABS HAVE TO BE REMOVED. THE SYMPHONY TAB BREAKER WAS AVAILABLE, BUT THERE WAS HUGE RESISTANCE TO THE TAB SNAPPING. AS THE SURGEON WAS APPLYING FORCE TO THE TAB BREAKER THE WHOLE CONSTRUCT AND NECK WERE VISIBLY TWISTING WITH HIS MOVEMENT. SURGEON SWAPPED TO USING A PAIR OF NIBBLERS. THE TAB STILL WERE RESISTANT TO SNAPPING. THE TABS WERE REMOVED BY USING A ROD HOLDER ON THE ROD IN SITU TO STABILISE WHILE FORCE WAS APPLIED TO SNAP THE TABS. IT WASN¿T IDEAL AND THE ROD WAS EXPOSED, AND SWOLLEN, SO THERE WAS A HUGE RISK TO FURTHER DAMAGE. THERE WAS NO REPORTED PATIENT COMPLICATIONS. PROCEDURE WAS COMPLETED WITH NO DELAY. THIS REPORT IS FOR A 4.0 CAN/RED SCRW 3.5X12. THIS IS REPORT 4 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19575109 ↗MalfunctionJun 20, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT PATIENT UNDERWENT A C5-C6 POSTERIOR FUSION USING THE SYMPHONY POSTERIOR CERVICAL SYSTEM. THE SURGEON OPTED TO USE THE LONG TAB SCREWS. THE PROCEDURE WENT SMOOTHLY, UNTIL PART OF THE FINAL STEP WHERE THE TABS HAVE TO BE REMOVED. THE SYMPHONY TAB BREAKER WAS AVAILABLE, BUT THERE WAS HUGE RESISTANCE TO THE TAB SNAPPING. AS THE SURGEON WAS APPLYING FORCE TO THE TAB BREAKER THE WHOLE CONSTRUCT AND NECK WERE VISIBLY TWISTING WITH HIS MOVEMENT. SURGEON SWAPPED TO USING A PAIR OF NIBBLERS. THE TAB STILL WERE RESISTANT TO SNAPPING. THE TABS WERE REMOVED BY USING A ROD HOLDER ON THE ROD IN SITU TO STABILISE WHILE FORCE WAS APPLIED TO SNAP THE TABS. IT WASN¿T IDEAL AND THE ROD WAS EXPOSED, AND SWOLLEN, SO THERE WAS A HUGE RISK TO FURTHER DAMAGE. THERE WAS NO REPORTED PATIENT COMPLICATIONS. PROCEDURE WAS COMPLETED WITH NO DELAY. THIS REPORT IS FOR A 4.0 CAN/RED SCRW 3.5X12. THIS IS REPORT 2 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19575094 ↗
Other Posterior cervical screw system devices
- INFINITY™ Occipitocervical Upper Thoracic System 3604026Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation2 recalls1 report
- CONNECTOR 79-2003Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 360PT4022Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- CONNECTOR 79-2002Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604532Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604024Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- OASYS 48551000Stryker Corporation · Orthosis, cervical pedicle screw spinal fixation45 reports
- YUKON OCT Spinal System 7601-05030VB Spine LLC · Orthosis, cervical pedicle screw spinal fixation15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page