SYMPHONY 1020-35-316SRecalls & reported problems

Medos International Sàrl · SYMPHONY OCT SYSTEM POLYAXIAL SCREW REDUCTION CANNULATED Diameter 4.0 Rod, 3.5X16mm 3.5 4.0 · Class 2 · Implant

All models of SYMPHONY (113)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034527819), received between Jun 20, 2024 and Aug 1, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Compatibility Problem1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 1, 2024Compatibility Problem; Material Deformation; Compatibility Problem; Material Deformation

    IT WAS REPORTED THAT ON (B)(6) 2024, DURING A POSTERIOR CERVICAL FUSION (C5) FOR CERVICAL MYELOPATHY THE SCREW IN QUESTION COULD NOT BE ATTACHED STRAIGHT TO THE SCREWDRIVER. THE SCREW WAS REPLACED WITH A NEW ONE, AS IT WAS STILL SLIGHTLY BENT EVEN AFTER REPEATED ATTEMPTS. THE OTHER 11 SCREWS HAD NO PROBLEM, AND ONLY ONE OF THEM WAS BENT NO MATTER HOW MANY TIMES THE SURGEON TRIED TO REATTACH IT. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH 30 MINUTES SURGICAL DELAY. THIS REPORT INVOLVES ONE (1) SYMPHONY OCT SYSTEM POLYAXIAL SCREW REDUCTION CANNULATED DIAMETER 4.0 ROD, 3.5X16MM 3.5 4.0. THIS IS REPORT 1 OF 1 FOR (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19882657 ↗
  • MalfunctionJun 20, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT PATIENT UNDERWENT A C5-C6 POSTERIOR FUSION USING THE SYMPHONY POSTERIOR CERVICAL SYSTEM. THE SURGEON OPTED TO USE THE LONG TAB SCREWS. THE PROCEDURE WENT SMOOTHLY, UNTIL PART OF THE FINAL STEP WHERE THE TABS HAVE TO BE REMOVED. THE SYMPHONY TAB BREAKER WAS AVAILABLE, BUT THERE WAS HUGE RESISTANCE TO THE TAB SNAPPING. AS THE SURGEON WAS APPLYING FORCE TO THE TAB BREAKER THE WHOLE CONSTRUCT AND NECK WERE VISIBLY TWISTING WITH HIS MOVEMENT. SURGEON SWAPPED TO USING A PAIR OF NIBBLERS. THE TAB STILL WERE RESISTANT TO SNAPPING. THE TABS WERE REMOVED BY USING A ROD HOLDER ON THE ROD IN SITU TO STABILISE WHILE FORCE WAS APPLIED TO SNAP THE TABS. IT WASN¿T IDEAL AND THE ROD WAS EXPOSED, AND SWOLLEN, SO THERE WAS A HUGE RISK TO FURTHER DAMAGE. THERE WAS NO REPORTED PATIENT COMPLICATIONS. PROCEDURE WAS COMPLETED WITH NO DELAY. THIS REPORT IS FOR A 4.0 CAN/RED SCRW 3.5X16. THIS IS REPORT 3 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19575106 ↗
  • MalfunctionJun 20, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT PATIENT UNDERWENT A C5-C6 POSTERIOR FUSION USING THE SYMPHONY POSTERIOR CERVICAL SYSTEM. THE SURGEON OPTED TO USE THE LONG TAB SCREWS. THE PROCEDURE WENT SMOOTHLY, UNTIL PART OF THE FINAL STEP WHERE THE TABS HAVE TO BE REMOVED. THE SYMPHONY TAB BREAKER WAS AVAILABLE, BUT THERE WAS HUGE RESISTANCE TO THE TAB SNAPPING. AS THE SURGEON WAS APPLYING FORCE TO THE TAB BREAKER THE WHOLE CONSTRUCT AND NECK WERE VISIBLY TWISTING WITH HIS MOVEMENT. SURGEON SWAPPED TO USING A PAIR OF NIBBLERS. THE TAB STILL WERE RESISTANT TO SNAPPING. THE TABS WERE REMOVED BY USING A ROD HOLDER ON THE ROD IN SITU TO STABILISE WHILE FORCE WAS APPLIED TO SNAP THE TABS. IT WASN¿T IDEAL AND THE ROD WAS EXPOSED, AND SWOLLEN, SO THERE WAS A HUGE RISK TO FURTHER DAMAGE. THERE WAS NO REPORTED PATIENT COMPLICATIONS. PROCEDURE WAS COMPLETED WITH NO DELAY. THIS REPORT IS FOR A 4.0 CAN/RED SCRW 3.5X16. THIS IS REPORT 1 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19575096 ↗