VIPER PRIME 286750150Recalls & reported problems

Medos International Sàrl · VIPER PRIME FENESTRATED ALIGNMENT GUIDE · Class 1

All models of VIPER PRIME (60)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034529776), received between Aug 20, 2025 and Dec 18, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility3
  • Contamination /Decontamination Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 18, 2025Contamination /Decontamination Problem; Device-Device Incompatibility; Contamination /Decontamination Problem; Device-Device Incompatibility

    IT WAS REPORTED THAT THIS WAS A POSTERIOR SPINAL FUSION(L4/5) PROCEDURE FOR SACRAL FRACTURE. DURING THE SURGERY, AFTER PLACING THE ALIGNMENT DEVICE ON THE SCREW, THE SURGEON CHECKED TO SEE IF THE OPEN CANNULA COULD PASS THROUGH, BUT THE OPEN CANNULA WOULD NOT GO THROUGH. THE SALES REP CHECKED THE ORIENTATION OF THE ALIGNMENT DEVICE SEVERAL TIMES AND REATTACHED IT, BUT IT STILL WOULD NOT GO THROUGH. THERE WAS NO PROBLEM WITH THE ORIENTATION OF THE SCREW, AND OF COURSE THE OPEN CANNULA WAS NOT BENT. THEREFORE, THE SURGEON TRIED CLEANING THE ALIGNMENT DEVICE WITH A GUIDE WIRE OUTSIDE THE OPERATING FIELD, AND ALTHOUGH THE GUIDE WIRE WENT THROUGH NORMALLY, THE OPEN CANNULA STILL WOULD NOT GO THROUGH. IT SEEMED THAT THE THICKER PART OF THE OPEN CANNULA WAS GETTING CAUGHT SOMEWHERE IN THE ALIGNMENT DEVICE, SO THE SURGEON REMOVED AND REINSERTED THE GUIDE WIRE TO CLEAN IT AGAIN, AND WHITE AND BLACK DEBRIS, SOME FOREIGN MATTER CAME OUT. IT'S POSSIBLE THAT CEMENT OR BLOOD FROM THE PREVIOUS OR EARLIER OPERATION HAD HARDENED AND STUCK TO THE INSIDE OF THE ALIGNMENT DEVICE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINS OF DELAY. PATIENT OUTCOME IS REPORTED AS STABLE. NO FURTHER INFORMATI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23843643 ↗
  • MalfunctionAug 20, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THIS WAS AN UNKNOWN SURGERY PERFORMED FOR THORACOLUMBAR VERTEBRAE ON (B)(6), 2025. THE SURGEON INSERTED THE ALIGNMENT DEVICE AND ATTEMPTED TO ATTACH AN OPEN CANNULA TO INSERT CEMENT, BUT IT DID NOT PENETRATE DEEP ENOUGH, AND THE CEMENT COULD NOT BE INSERTED. IT APPEARS THAT A WHITE SUBSTANCE LIKE CEMENT CAME OUT DURING THE OPERATION, AND IT IS LIKELY THAT CEMENT HAD BEEN STUCK INSIDE. WHEN THE SURGEON CHECKED AFTER THE OPERATION, HE FOUND THAT THE OPEN CANNULA HAD INDEED NOT REACHED ITS SPECIFIED SETTING POSITION. USING ANOTHER ALIGNMENT GUIDE, THE SURGERY WAS COMPLETED WITHOUT ANY PROBLEMS. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22851280 ↗
  • MalfunctionAug 20, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THIS WAS AN UNKNOWN SURGERY PERFORMED FOR THORACOLUMBAR VERTEBRAE ON (B)(6) 2025. THE SURGEON INSERTED THE ALIGNMENT DEVICE AND ATTEMPTED TO ATTACH AN OPEN CANNULA TO INSERT CEMENT, BUT IT DID NOT PENETRATE DEEP ENOUGH, AND THE CEMENT COULD NOT BE INSERTED. IT APPEARS THAT A WHITE SUBSTANCE LIKE CEMENT CAME OUT DURING THE OPERATION, AND IT IS LIKELY THAT CEMENT HAD BEEN STUCK INSIDE. WHEN THE SURGEON CHECKED AFTER THE OPERATION, HE FOUND THAT THE OPEN CANNULA HAD INDEED NOT REACHED ITS SPECIFIED SETTING POSITION. USING ANOTHER ALIGNMENT GUIDE, THE SURGERY WAS COMPLETED WITHOUT ANY PROBLEMS. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22851084 ↗