SYMPHONY 2020-00-501Recalls & reported problems
Medos International Sàrl · SYMPHONY OCT SYSTEM AO HANDLE, MEDIUM WITH RATCHET · Class 1
All models of SYMPHONY (113)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034532615), received between Jun 7, 2021 and Jan 9, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility4
- Corroded2
- Contamination /Decontamination Problem1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 9, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2024, THE HANDLE STOPPED WORKING DURING SCREW INSERTION. THERE WAS NO SURGERY DELAY DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21118102 ↗MalfunctionAug 22, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT BEFORE SURGERY ON (B)(6) 2024, WHEN THE AO HANDLES IN QUESTION WERE CLEANED AND STERILIZED IN THE CENTRAL SUPPLY ROOM, A BLOOD-LIKE DIRT WAS FOUND ON THEM, AND THEIR USE WAS STOPPED. ANOTHER HANDLE WAS USED AND THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. THIS REPORT INVOLVES ONE SYMPHONY OCT SYSTEM AO HANDLE, MEDIUM WITH RATCHET THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20044074 ↗MalfunctionMay 21, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2024, THE SYMPHONY SCREW DRIVER HANDLE AND DRIVER WERE DIFFICULT TO DISCONNECT. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) AO HANDLE MEDIUM W RATCHET. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19366968 ↗MalfunctionMar 24, 2023Mechanical Problem; Mechanical Problem
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN PORTUGAL AS FOLLOWS: IT WAS REPORTED THAT THE RATCHET HANDLE DIDN'T WORK (ROTATE) AS WAS SUPPOSED. IT WAS WORKING AS A REGULAR HANDLE. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS IDENTIFIED THAT THERE APPEARS TO BE FOREIGN SUBSTANCE BETWEEN THE PULL RELEASE AO COUPLING AND THE RATCHET MECHANISM. THIS REPORT IS FOR ONE (1) AO HANDLE MEDIUM W RATCHET THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16612462 ↗MalfunctionJun 7, 2021Corroded; Device-Device Incompatibility; Corroded; Device-Device Incompatibility
THIS IS REPORT 3 OF 5 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11950377 ↗MalfunctionJun 7, 2021Corroded; Device-Device Incompatibility; Corroded; Device-Device Incompatibility
THIS IS REPORT 1 OF 5 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11950308 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.