SYMPHONY 2020-00-504Recalls & reported problems
Medos International Sàrl · SYMPHONY OCT SYSTEM AO HANDLE, TORQUE LIMITING 3.0Nm · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Medos Symphony AO torque-limiting handle is used in orthopedic surgery to tighten bone screws to a set torque. The 58 reports mostly describe handles reading out of calibration or failing the annual torque test, with a few mentioning rust and a ratcheting part stuck. No deaths or injuries are described, and most reports involve no patient impact. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 58
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 58 adverse event reports that list this exact device (DI 10705034532813), received between Apr 23, 2020 and Aug 18, 2026 — classified as 58 malfunctions.
The most frequently reported specific device problem is “Calibration Problem” (56 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Calibration Problem56
- Corroded2
- Device-Device Incompatibility1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions53
- Insufficient Information4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 58 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 18, 2026Calibration Problem
Patient problems: Insufficient Information
MAUDE report 50308243 ↗MalfunctionAug 18, 2026Calibration Problem
Patient problems: Insufficient Information
MAUDE report 50308241 ↗MalfunctionAug 18, 2026Calibration Problem
Patient problems: Insufficient Information
MAUDE report 50308229 ↗MalfunctionDec 4, 2025Calibration Problem; Calibration Problem
ADDITIONAL INFORMATION RECEIVED: THESE IMPACTED PRODUCTS ARE RESULT FROM ANNUAL TORQUE TEST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23715125 ↗MalfunctionDec 4, 2025Calibration Problem; Calibration Problem
ADDITIONAL INFORMATION RECEIVED: THESE IMPACTED PRODUCTS ARE RESULT FROM ANNUAL TORQUE TEST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23715098 ↗MalfunctionJun 26, 2025Corroded; Mechanical Jam; Corroded; Mechanical Jam
IT WAS REPORTED THAT THE HANDLES NEEDED REPLACING DUE TO RUST AND RATCHETING PORTION BEING STUCK. THERE WAS NO PATIENT INVOLVEMENT. ISSUE WAS NOTED PRE-OPERATIVELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22335990 ↗MalfunctionMay 16, 2025Calibration Problem; Calibration Problem
IT WAS REPORTED THAT AT A (B)(6) FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22038465 ↗MalfunctionMay 16, 2025Calibration Problem; Calibration Problem
IT WAS REPORTED THAT AT A JNJ FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22038430 ↗MalfunctionMay 16, 2025Calibration Problem; Calibration Problem
IT WAS REPORTED THAT PATIENT UNDERWENT SPINAL FUSION ON (B)(6) 2025. DURING FINAL TORQUE ON SYMPHONY IMPLANTS, THE SURGEON FELT A TOGGLE OF THE SCREWDRIVER SHAFT IN THE TORQUE DRIVER HANDLE. IT WAS REPLACED WITH A SECOND TORQUE DRIVER HANDLE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO GENERATION OF FRAGMENTS. THERE WAS NO PATIENT CONSEQUENCES. THERE WAS NO MEDICAL INTERVENTION.
Patient problems: Insufficient Information
MAUDE report 22036211 ↗MalfunctionSep 5, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON (B)(6) 2024, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED THAT ITEM WAS FOUND TO BE OUT OF DRAWING SPECIFIC. THE DRAWING SPECIFICATION TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. ALL INFORMATION REGARDING THIS EVENT HAS BEEN REPORTED. THIS REPORT IS FOR ONE (1) AO HANDLE TORQUE LIMITING3.0NM FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20149265 ↗MalfunctionSep 5, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED DURING ROUTINE INCOMING INSPECTION OF A LOANER SET AT THE HOSPITAL, IT WAS OBSERVED THAT THE 3.0 NM HANDLE, WAS FOUND TO BE OUT OF DRAWING SPECIFICATION; THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20148120 ↗MalfunctionSep 5, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON (B)(6) 2024, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET AT THE HOSPITAL, IT WAS OBSERVED THAT THE 3.0 NM, WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) AO HANDLE TORQUE LIMITING3.0NM THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20147649 ↗MalfunctionSep 4, 2024Calibration Problem; Calibration Problem
DURING ROUTINE INCOMING INSPECTION OF A LOANER SET AT THE HOSPITAL, IT WAS OBSERVED THAT 3.0 NM-HANDLE, WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE DRAWING SPECIFICATION THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. ALL INFORMATION REGARDING THIS EVENT HAS BEEN REPORTED. THIS IS REPORT 1 OF 1 AO HANDLE TORQUE LIMITING3.0NM FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20141185 ↗MalfunctionSep 4, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON (B)(6) 2024 DURING ROUTINE INCOMING INSPECTION OF A LOANER SET AT THE SITE, IT WAS OBSERVED THAT THE HANDLE HAD BEEN FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR A SYMPHONY OCT SYSTEM AO HANDLE, TORQUE LIMITING 3.0NM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20134465 ↗MalfunctionAug 29, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING ROUTINE INCOMING INSPECTION OF A LOANER SET AT THE HOSPITAL, IT WAS OBSERVED THAT 2020-00-504, 3.0 NM HANDLE, WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) AO HANDLE TORQUE LIMITING3.0NM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20103208 ↗MalfunctionAug 28, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON (B)(6) 2024 DURING ROUTINE INCOMING INSPECTION OF A LOANER SET AT THE HOSPITAL, IT WAS OBSERVED THAT THE HANDLE WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) AO HANDLE TORQUE LIMITING3.0NM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20091198 ↗MalfunctionAug 28, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT DURING ROUTINE INCOMING INSPECTION OF A LOANER SET AT THE HOSPITAL, IT WAS OBSERVED THAT 3.0 NM HANDLE, WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) AO HANDLE TORQUE LIMITING 3.0NM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20091053 ↗MalfunctionAug 28, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED ON JUL 31, 2024 DURING ROUTINE INCOMING INSPECTION OF A LOANER SET AT THE HOSPITAL, IT WAS OBSERVED THAT HANDLE WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) AO HANDLE TORQUE LIMITING3.0NM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20091014 ↗MalfunctionAug 28, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED THAT THE HANDLE, WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THE PRODUCT HAS BEEN QUARANTINED AND IS AVAILABLE AND IS BEING RETURNED TO THE REPAIR SITE. A REPLACEMENT DEVICE HAS BEEN ORDERED. NO REPLACEMENT PRODUCT IS REQUIRED, AND NO SHIPPING LABEL IS REQUIRED FOR THE PRODUCT RETURN. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE (1) AO HANDLE TORQUE LIMITING3.0NM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20089227 ↗MalfunctionAug 26, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON JULY 30, 2024, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED THAT THE N.M HANDLE, WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE DRAWING SPECIFICATION OF THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. ALL INFORMATION REGARDING THIS EVENT HAS BEEN REPORTED. THERE WAS NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) AO HANDLE TORQUE LIMITING3.0NM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20072206 ↗MalfunctionAug 23, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON (B)(6) 2024, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED THAT THE 3.0 NM HANDLE WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT INVOLVES ONE (1) SYMPHONY OCT SYSTEM AO HANDLE, TORQUE LIMITING 3.0NM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20055136 ↗MalfunctionAug 23, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON (B)(6) 2024 DURING ROUTINE INCOMING INSPECTION OF A LOANER SET AT THE SITE, IT WAS OBSERVED THAT 3.0 NM HANDLE, WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) AO HANDLE TORQUE LIMITING3.0NM. THIS IS REPORT 1 OF 1 FOR COMPLAINT(B)(4). PC-001662281
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20046724 ↗MalfunctionAug 22, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON JULY 26, 2024, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED ITEM WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) AO HANDLE TORQUE LIMITING3.0NM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20041422 ↗MalfunctionAug 21, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT DURING ROUTINE INCOMING INSPECTION OF A LOANER SET ON JULY 19, 2024, IT WAS OBSERVED THAT THE AO HANDLE TORQUE LIMITING3.0NM WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR AN AO HANDLE TORQUE LIMITING3.0NM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20016310 ↗MalfunctionAug 21, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT DURING ROUTINE INCOMING INSPECTION OF A LOANER SET ON (B)(6)2024, IT WAS OBSERVED THAT A 3.0 NM T-HANDLE WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR AN AO HANDLE TORQUE LIMITING3.0NM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20015478 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page