Skip to content

NA 03.611.030Recalls & reported problems

Synthes GmbH · 1.5MM SETSCREW EXTRACTOR · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034713830), received between Nov 19, 2018 and Sep 18, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break2

Patient problems

  • No Patient Involvement2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2019Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING A ROUTINE INCOMING INSPECTION OF LOANER SET, THE SET SCREW EXTRACTOR WAS OBSERVED BROKEN. THERE WAS NO KNOWN HOSPITAL OR PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) 1.5 MM SET SCREW EXTRACTOR. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 9084315 ↗
  • MalfunctionNov 19, 2018Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2018, THE CENTRAL STERILE STAFF FOUND OUT THAT THE 1.5 MM SET SCREW EXTRACTOR WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) 1.5MM SETSCREW EXTRACTOR. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 8087128 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page