Skip to content

NA 03.614.027Recalls & reported problems

Synthes GmbH · PERSUADER · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10705034714981), received between Jun 11, 2021 and May 17, 2025 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem5
  • Compatibility Problem3
  • Mechanical Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information1
  • Spinal Cord Injury1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 17, 2025Compatibility Problem; Material Integrity Problem; Compatibility Problem; Material Integrity Problem

    IT WAS REPORTED THE PERSUADER WAS DIFFICULT TO ASSEMBLE. WEAR AND TEAR WAS NOTICED; THE METAL WAS NOT ALIGNING. ANOTHER SET WAS OPENED. THERE WAS NO SURGICAL DELAY OR IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22041092 ↗
  • MalfunctionSep 20, 2024Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT ON (B)(6) 2024, THE PRONGS ON THE PERSUADER WERE SPLAYED TOO FAR APART AND COULD NOT GRASP THE SCREW HEADS. THERE WAS NO SURGICAL DELAY. THE PROCEDURE COMPLETED SUCCESSFULLY. THE PATIENT OUTCOME WAS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 20266726 ↗
  • MalfunctionNov 29, 2023Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN ISRAEL AS FOLLOWS: IT WAS REPORTED ON NOVEMBER 6, 2023, THAT DURING A POSTERIOR CERVICAL SURGERY WITH "SYNAPS" SET THE THREE ITEMS DID NOT WORK PROPERLY EVEN THOUGH THE SURGEON FOLLOWED THE TECHNIQUE CORRECTLY. SKU ROD INTRODUCT-NSTR DISINTEGRATED SEVERAL TIMES DURING USE. SKU HOLD-SL FAILED TO CONNECT TO THE SCREW PROPERLY AND IT APPEARS TO BE WORN AT ITS END. SKU BEND-PLIER F/R 3.5 IT WAS NOT POSSIBLE TO BEND THE ROD AS EXPECTED ACCORDING TO THE SURGEON. THERE WAS NO UNUSUAL USE OF THE PRODUCT AND THESE ARE PRODUCTS THAT BELONG TO A SET THAT THE SURGEON HAS BEEN USING FOR YEARS IN THIS HOSPITAL AND IN OTHER HOSPITALS. SKU. HOLD-SL ACCORDING TO WHAT IS VISIBLE, IT APPEARS TO BE DUE TO NATURAL WEAR AND TEAR AND THE OTHER ITEMS HAVE AN UNKNOWN CAUSE. THIS IS A SENIOR SURGEON WHO KNOWS THE TECHNIQUE AND WAS PROPERLY INSTRUCTED. THE SURGEON OVERCAME THE INCIDENT BY USING ADDITIONAL PRODUCTS IN THE SET. THERE WAS NO SURGICAL DELAY DUE TO THE REPORTED EVENT. THERE WERE NO PATIENT OUTCOMES OR CONSEQUENCES. THIS REPORT IS FOR ONE (1)ROD INTRODUCT-NSTR. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18231688 ↗
  • MalfunctionOct 4, 2023Mechanical Problem; Mechanical Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE SET WAS OPEN PRIOR TO THE CASE START, AND THE PERSUADER INSTRUMENT HAD FALLEN APART. THE SCREW THAT HOLDS THE SPRING MECHANISM OF THE INSTRUMENT HAD FALLEN OUT, AND THE SPRING MECHANISM FELL APART. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE PERSUADER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17870015 ↗
  • InjuryJul 10, 2023Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THAT THE DURING A POSTERIOR CERVICAL FIXATION USING SYNAPSE SYSTEM, THE SURGEON REQUESTED TO USE THE REDUCTION DEVICE. THE SCRUB HANDED THE INSTRUMENT TO THE SURGEON IN THE HORIZONTAL PLANE, THE SURGEON USED THE DEVICE AND AS HE LISTED IT OFF THE SCREW THE INNER BARREL FELL AND HIT THE EXPOSED SPINAL CORD, CAUSING THE PATIENT TO HAVE FURTHER INJURY. PATIENT STATUS/ OUTCOME : OPERATION COMPLETED, PATIENT REMAINED SEDATED AND IN ITU. THIS REPORT IS FOR ONE (1) ROD INTRODUCT-NSTR. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Spinal Cord Injury

    MAUDE report 17287514 ↗
  • MalfunctionFeb 13, 2023Mechanical Problem; Mechanical Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN NORWAY AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE CLIP¿S ATTACHMENT TO THE SLEEVES WAS NOT TIGHT ENOUGH. DURING ROD REDUCTION, THEY SLIPPED AND LOST THEIR ATTACHMENT ON THE SLEEVE AND THE SCREW HEAD. THEREFORE IT¿S NOT DOING ITS WORK ACCORDING TO ITS PURPOSE OF PRESSING THE ROD DOWN. THERE WAS A 15 MINUTE DELAY IN SURGERY AS A RESULT OF THE EVENT. THE PROCEDURE WAS COMPLETED WITH A SIMILAR PRODUCT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE PERSUADER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16367904 ↗
  • MalfunctionOct 31, 2022Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT DUE TO OVERUSE AND WEAR AND TEAR THE PERSUADER FROM THE SYNAPSE SET ISN'T REDUCING THE ROD CORRECTLY AS THE INNER SLEEVE GETS STUCK. ALSO THE OUTTER SLEEVE DOESN'T CORRECTLY ENGAGE WITH THE SCREW HEAD AS ITS LOOSE. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO ADDITIONAL TIME NEEDED IN THE SURGERY. NO OTHER INFO IS AVAILABLE. IS UNKNOWN IF THE PROCEDURE SUCCESSFULLY COMPLETED. PATIENT STATUS/ OUTCOME / CONSEQUENCES :NO. THIS REPORT IS FOR ONE (1) PERSUADER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15700623 ↗
  • MalfunctionOct 27, 2021Compatibility Problem; Compatibility Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT, THE INSTRUMENT WAS LOOSE AND SLACK, ATTACHES POORLY TO THE SCREW. IT SLIPS QUICKLY AND IS UNSTABLE. PROCEDURE WAS DELAYED FOR FIFTEEN(15)- TWENTY(20) MINUTES. THIS REPORT IS FOR ONE (1) PERSUADER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12704975 ↗
  • MalfunctionJun 11, 2021Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT THE PATIENT UNDERWENT FOR A SURGERY. DURING THE SURGERY, THE INSTRUMENT HANDLE WAS DISENGAGED FROM THE INNER AND OUT SHAFTS. THE SHAFTS ARE REMAINED ATTACHED TO THE PLOY AXIAL SCREW HEAD AND THE HANDLE COULD NOT BE REATTACHED. THE SURGERY WAS DELAYED TWO-MINUTES. IT WAS UNKNOWN IF THE SURGERY COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNKNOWN RODS: SYNAPSE (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1). UNKNOWN POLYAXIAL SCREWS: SYNAPSE (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) ROD INTRODUCT-NSTR. THIS REPORT IS 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11988011 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page