NA 03.616.043Recalls & reported problems
Synthes GmbH · LOCKING HOLDING SLEEVE-LONG FOR MATRIX · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10705034716343), received between Mar 19, 2018 and Jan 31, 2023 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Integrity Problem4
- Break3
- Compatibility Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 31, 2023Material Integrity Problem; Material Integrity Problem
THIS REPORT IS RELATED TO (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16274788 ↗MalfunctionAug 24, 2022Compatibility Problem; Material Integrity Problem; Compatibility Problem; Material Integrity Problem
IT WAS REPORTED ON (B)(6) 2022, THAT THE STRIPPED DRIVER SLEEVE THREADED AND IT WAS DIFFICULT IN ENGAGING HEAD OF SCREW. THERE WAS NO PATIENT INVOLVEMENT. CONCOMITANT DEVICE REPORTED: UNK. SCREWS: TRAUMA (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: UNKNOWN). THIS REPORT IS FOR ONE (1) LOCKING HOLDING SLEEVE-LONG FOR MATRIX. THIS IS REPORT 2 OF 2 FOR COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15281980 ↗MalfunctionAug 24, 2022Compatibility Problem; Material Integrity Problem; Compatibility Problem; Material Integrity Problem
IT WAS REPORTED ON (B)(6) 2022, THAT THE STRIPPED DRIVER SLEEVE THREADED AND IT WAS DIFFICULT IN ENGAGING HEAD OF SCREW. THERE WAS NO PATIENT INVOLVEMENT. CONCOMITANT DEVICE REPORTED: UNK - SCREWS: TRAUMA(PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: UNKNOWN) THIS REPORT IS FOR ONE (1) LOCKING HOLDING SLEEVE-LONG FOR MATRIX. THIS IS REPORT 1 OF 2 FOR COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15281972 ↗MalfunctionAug 1, 2022Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN HONG KONG AS FOLLOWS: IT WAS REPORTED THAT THE PEDICLE SCREW DRIVER SLEEVE OF MATRIX DEGEN WAS FOUND DAMAGED AFTER PATIENT USED IN SPINE OPERATION. THERE IS NO FURTHER INFORMATION. THIS REPORT IS FOR ONE (1) RETAIN-SLEEVE LOCKING LONG. THIS IS REPORT 1 OF 1 FOR COMPLAINT: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15139285 ↗MalfunctionApr 29, 2019Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING A TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) PROCEDURE DUE TO COLLAPSE ON THE DISC SPACE ON (B)(6) 2019, THE TIP OF THE LONG LOCKING RETAINING SLEEVE WAS BROKEN AFTER AN UNKNOWN SCREW WAS INSERTED UPON CHECKING ON AN X-RAY. THE BROKEN FRAGMENT WAS SEEN ON THE SCREW HEAD BUT WAS EASILY REMOVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. PATIENT CONDITION WAS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREW (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) LOCKING HOLDING SLEEVE-LONG FOR MATRIX. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8560354 ↗MalfunctionMar 5, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, THE LOCKING HOLDING SLEEVE-LONG FOR MATRIX BROKE DURING SCREW INSERTION. FRAGMENTS WERE GENERATED AND REMOVED EASILY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8391102 ↗MalfunctionMar 19, 2018Break; Break
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED DURING A PROCEDURE FOR LUMBAR SPINAL CANAL STENOSIS AT T10-L4, ON (B)(6) 2018, THE TIP OF THE LOCKING HOLDING SLEEVE WAS BROKEN. ANOTHER OF THE SAME DEVICE WAS READILY AVAILABLE AND WAS USED TO COMPLETE THE PROCEDURE SAFELY WITH A DELAY OF APPROXIMATELY 5 MINUTES. NO ADVERSE CONSEQUENCE TO THE PATIENT WAS REPORTED. THIS REPORT IS FOR ONE (1) LOCKING HOLDING SLEEVE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7349684 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.