Matrix 03.632.087Recalls & reported problems
Synthes GmbH · MATRIX DISTRACTOR RACK · Class 1
All models of Matrix (80)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034719290), received between Jan 30, 2017 and Oct 1, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Compatibility Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Compatibility Problem2
- Break1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 1, 2024Compatibility Problem; Compatibility Problem
IT WAS REPORTED THAT ON (B)(6) 2024 , WHEN THE SURGEON WAS USING THE RETRACTOR THE SCREW WITHIN THE 2 PART TOOTHED DISTRACTOR WOULD NOT TIGHTEN COMPLETELY. THERE WAS NO PATIENT INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED AT THIS TIME. THIS REPORT INVOLVES ONE TOOTHED RACK RETRACTOR F/MATRIX.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20345577 ↗MalfunctionOct 5, 2023Compatibility Problem; Compatibility Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT THE LOCKING MECHANISM ON THE MATRIX DISTRACTOR KEEPS SLIPPING AND WILL NOT LOCK IN PLACE. THE SURGEON COULD NOT KEEP THE SCREWS DISTRACTED. THERE WAS NO SURGICAL DELAY DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT OUTCOMES OR CONSEQUENCES. THIS REPORT IS FOR ONE (1) MATRIX DISTRACTOR RACK. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17875892 ↗MalfunctionMay 19, 2021Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT IN THE STERILIZATION DEPARTMENT IT WAS FOUND THAT A SPRING HAD BROKEN OFF AND THE SPREADER WAS THEREFORE BROKEN. THE SURGERY AND PATIENT OUTCOME WERE UNKNOWN. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) MATRIX DISTRACTOR RACK. THIS REPORT IS 1 OF 1 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11850324 ↗MalfunctionJan 30, 2017Break; Break
IT WAS REPORTED THAT ON (B)(6) 2017, PATIENT UNDERWENT INITIAL TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) PROCEDURE. DURING THE PROCEDURE, PATIENT WAS IMPLANTED AT L5-S1 DISC LEVEL WITH MATRIX PEDICLE SCREWS, RODS, CAPS AND OPAL 10X11X24 TLIF SPACE AT LEFT SIDE. WHEN DISTRACTOR INSTRUMENT WAS FULLY DISTRACTED, SURGEON NOTICED ONE OF THE POSTS OF THE MATRIX RACK DISTRACTOR INSTRUMENT IN THE L5 DISC LEVEL HAD FRACTURED AT THE TOP. THE SURGEON REMOVED THE DISTRACTOR ON THE LEFT AND THE BROKEN PIECE OF THE DISTRACTOR ALONG THE L5 TUBE. WHILE REMOVING BOTH THREADED RODS, THE RACK DISTRACTOR INSTRUMENT CAME UNATTACHED FROM THE LEFT L5-S1 BONE SCREWS. ALSO, UPON THREADING THE TUBE ONTO THE S1 TULIP HEAD SCREW IMPLANT, THE TULIP HEAD SCREW POPPED OFF BUT DID NOT BREAK. THE TULIP HEAD SCREW WAS NOT ABLE TO BE REPLACED ON THE ROD BECAUSE THE INNER SADDLE ON TOP OF THE SCREW WAS LOCKED IN THE DOWN POSITION. ENTIRE TULIP HEAD SCREW WAS REMOVED AND A NEW TI MATRIX TULIP HEAD SCREW WAS READILY AVAILABLE TO COMPLETE THE PROCEDURE. THE RACK FROM OPPOSITE SIDE WAS PLACED DOWN THE RODS AND EXPANDED TO ALLOW FOR TLIF INSERTION. PATIENT WAS CLOSED UP AND DRAIN INSERTED. THERE WAS EIGHT (8) MINUTES
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6286580 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page