Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034720760), received between Apr 24, 2020 and Sep 25, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Compatibility Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K112068); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Compatibility Problem1
- Device-Device Incompatibility1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Compatibility Problem
It was reported while the implant was mounted on the implant holder, the holder was squeezed, according to the surgical technique. Then the knurled nut was turned on the thread to fix the position of the holder. Shortly before the nut was at the end position, it stopped turning and was jammed. It could no longer be loosened by hand. It did not come loose even with a clamp or forceps. A new implant had to be unpacked, which was then implanted with a clamp. The surgery could be completed successfully with a ten (10) minute delay and no harm to the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50899830 ↗MalfunctionMar 17, 2026Material Integrity Problem
IT WAS REPORTED DURING PREPARATION, IT WAS FOUND THAT THE LOCKNUT DOES NOT ROTATE CORRECTLY. THE THREAD APPEARS TO BE DEFECTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24617615 ↗MalfunctionApr 24, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, A POST-IMPLANTATION PROCEDURE WAS PERFORMED ON THE CAGE WITH CASPAR RESTRICTOR AND IMPLANT HOLDER. THE SURGEON WANTED TO REMOVE THE IMPLANT HOLDER FROM THE IMPLANT. WHILE DOING THIS, THE PLATE PART OF THE ZERO-P-VA REMAINED ON THE IMPLANT HOLDER AND THE PEEK PART REMAINED IN THE DISC POCKET. THE SURGEON WAS ABLE TO REMOVE THE PEEK-PART OF THE ZERO P VA WITH A PAIR OF TWEEZERS FROM THE DISC POCKET. THE SURGERY WAS SUCCESSFULLY COMPLETED WITH A DELAY OF THREE (3) MINUTES. NO ADVERSE PATIENT HARM WAS REPORTED. THIS REPORT IS FOR ONE (1) INSERTION DEVICE FOR ZERO-P VA. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9996224 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112068. One submission often covers many models, so check the product and lot details in each record.
- Z-0455-2014 · Sep 30, 2013 · TerminatedSynthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is a stand alone anterior cervical interbody fusion device indica
Other Intervertebral fusion device with integrated fixation, cervical devices
- NA 03.617.902Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical103 reports
- NA 03.617.900Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical77 reports
- Divergence® Anterior Cervical Fusion System G6626525Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical75 reports
- Divergence® Anterior Cervical Fusion System G6626526Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical49 reports
- NA 03.617.905Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical29 reports
- NA 03.617.904Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical27 reports
- Divergence® Anterior Cervical Fusion System G6626746Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical21 reports
- NA 03.647.972Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page