NA 03.647.971Recalls & reported problems
Synthes GmbH · REMOVAL SCREWDRIVER · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034720791), received between Oct 19, 2017 and Sep 13, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K112068); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Integrity Problem3
- Break2
- Activation, Positioning or Separation Problem1
- Component Missing1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Patient Involvement2
- Foreign Body In Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 13, 2024Activation, Positioning or Separation Problem; Material Integrity Problem; Activation, Positioning or Separation Problem; Material Integrity Problem
ADDITIONAL EVENT INFORMATION STATES THAT THE SCREW HAS BACKED OUT.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 20218088 ↗MalfunctionDec 20, 2023Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON NOVEMBER 28, 2023, THE INSTRUMENTS WERE TAKEN OFF THE TRAY BY THE STERILIZATION TEAM BECAUSE THEY APPEARED BLUNT. THIS REPORT INVOLVES ONE (1) REMOVAL SCREWDRIVER. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18364713 ↗MalfunctionDec 20, 2023Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON NOVEMBER 28, 2023, THE INSTRUMENTS WERE TAKEN OFF THE TRAY BY THE STERILIZATION TEAM BECAUSE THEY APPEARED BLUNT. THIS REPORT INVOLVES ONE (1) REMOVAL SCREWDRIVER. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18364637 ↗MalfunctionApr 18, 2023Material Twisted/Bent; Material Twisted/Bent
DEVICE REPORT FROM PORTUGAL REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED, THAT ON (B)(6) 2023, THE SET ONLY HAD ONE SCREWDRIVER SHAFT. ONE WAS MISSING. IT WAS LATER DETERMINED, THAT THE MISSING SCREWDRIVER SHAFT WAS PRESENT, BUT WAS STUCK IN THE REMOVAL SCREWDRIVER. UPON MANUFACTURER INVESTIGATION, IT WAS DETERMINED, THAT THE REMOVAL SCREWDRIVER WAS TWISTED FROM THE TIP. THIS REPORT IS FOR A REMOVAL SCREWDRIVER. THIS IS REPORT 2 OF 2 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16763348 ↗MalfunctionNov 13, 2018Break; Component Missing; Break; Component Missing
IT WAS REPORTED THAT ON (B)(6) 2018, DURING EVALUATION IN THE EVALUATION DEPARTMENT, IT WAS DISCOVERED THAT THE REMOVAL SCREWDRIVER HAD A BROKEN/MISSING TIP. THERE WERE NO PROCEDURE OR PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 8066667 ↗MalfunctionOct 19, 2017Break; Break
IT WAS REPORTED THAT BOTH THE REMOVAL SCREWDRIVER AND THE INNER SHAFT FOR THE REMOVAL SCREWDRIVER HAVE BROKEN TIPS. THE DEVICE ISSUES WERE FOUND DURING INSPECTION OF THE EVALUATION SET. THERE IS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) REMOVAL SCREWDRIVER THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Patient Involvement
MAUDE report 6962381 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112068. One submission often covers many models, so check the product and lot details in each record.
- Z-0455-2014 · Sep 30, 2013 · TerminatedSynthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is a stand alone anterior cervical interbody fusion device indica
Other Intervertebral fusion device with integrated fixation, cervical devices
- NA 03.617.902Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical103 reports
- NA 03.617.900Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical77 reports
- Divergence® Anterior Cervical Fusion System G6626525Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical75 reports
- Divergence® Anterior Cervical Fusion System G6626526Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical49 reports
- NA 03.617.905Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical29 reports
- NA 03.617.904Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical27 reports
- Divergence® Anterior Cervical Fusion System G6626746Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical21 reports
- NA 03.647.972Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page