NA 03.647.994Recalls & reported problems
Synthes GmbH · 2.5MM AWL WITH SLEEVE VARIABLE ANGLE · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034720876), received between Nov 26, 2019 and May 7, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K112068); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 7, 2021Break; Break
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE THAT TWO (2) 2.5 MM AWLS WITH SLEEVES HAD BROKEN SPRINGS. DAMAGED SPRINGS WERE NOTED DURING REPROCESSING THEREFOR THERE WAS NO SURGICAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) 2.5MM AWL WITH SLEEVE VARIABLE ANGLE. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11791105 ↗MalfunctionMay 7, 2021Break; Break
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE THAT TWO (2) 2.5 MM AWLS WITH SLEEVES HAD BROKEN SPRINGS. DAMAGED SPRINGS WERE NOTED DURING REPROCESSING THEREFOR THERE WAS NO SURGICAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) 2.5MM AWL WITH SLEEVE VARIABLE ANGLE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11791040 ↗MalfunctionOct 15, 2020Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT DURING THE INSPECTION OF 2.5 AWL, IT WAS DISCOVERED THE SPRING SLEEVE WOULDN'T THREAD PROPERLY, AND WAS DAMAGED WHEN TRYING TO RE-ENGAGE INTO THE SLEEVE. THERE WAS NO PATIENT CONSEQUENCES, OR SURGICAL DELAY INVOLVED. THIS COMPLAINT INVOLVES ONE(1) DEVICE. THIS REPORT IS FOR (1) MMSI DRIVER SHAFT FOR RED SCREW. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10684479 ↗MalfunctionNov 26, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, DURING INSPECTION OF SET AT LOANERS DEPARTMENT. THE AWL WITH SLEEVE VARIABLE ANGLE WAS BROKEN INTO THREE (3) PIECES. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 9375738 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112068. One submission often covers many models, so check the product and lot details in each record.
- Z-0455-2014 · Sep 30, 2013 · TerminatedSynthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is a stand alone anterior cervical interbody fusion device indica
Other Intervertebral fusion device with integrated fixation, cervical devices
- NA 03.617.902Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical103 reports
- NA 03.617.900Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical77 reports
- Divergence® Anterior Cervical Fusion System G6626525Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical75 reports
- Divergence® Anterior Cervical Fusion System G6626526Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical49 reports
- NA 03.617.905Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical29 reports
- NA 03.617.904Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical27 reports
- Divergence® Anterior Cervical Fusion System G6626746Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with integrated fixation, cervical21 reports
- NA 03.647.972Synthes GmbH · Intervertebral fusion device with integrated fixation, cervical17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page