NA 03.820.153SRecalls & reported problems
Synthes (U.S.A.) LP · MILLING BIT HXC-STERILE · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034731827), received between Apr 5, 2018 and Apr 5, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 5, 2018Break; Break
IT WAS REPORT THAT THE PATIENT HAD ORIGINAL SURGERY ON (B)(6) 2018 FOR TREATMENT OF A SPINE CERVICAL ARTIFICIAL DISC REPLACEMENT (PDC) IMPLANT AT C5-C6 DISC LEVELS. PATIENT WAS IMPLANTED WITH ONE (1) LEVEL UNKNOWN PDC IMPLANT. INTER-OPERATIVELY, IT WAS REPORTED THAT A MILLING BIT HXC INSTRUMENT TIP BROKE DURING THE PROCEDURE. AFTER THE SURGEON HAD COMPLETED THE SUPERIOR KEEL CUT, THE MILLING BIT WAS PLACED IN THE INFERIOR KEEL PORT. THE BIT TIP BROKE AFTER MILLING WAS ONLY ABOUT 80% COMPLETE. NO HARM WAS REPORTED TO THE PATIENT. THE MILLING BIT WAS REMOVED BY REMOVING THE GUIDE AND THE KEEL CUT WAS COMPLETED USING A CHISEL. THE CASE PROCEEDED WITHOUT ISSUE. SURGERY WAS COMPLETED SUCCESSFULLY WITH NO TIME DELAY. PATIENT IS REPORTED IN STABLE CONDITION. REPORTER HAS NO MORE INFORMATION TO REPORT ON THIS EVENT. CONCOMITANT DEVICE UNKNOWN ARTIFICIAL DISC REPLACEMENT PDC (ITEM # UNKNOWN, LOT # UNKNOWN, QUANTITY 1 EACH). THIS REPORT IS FOR ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7401680 ↗
FDA clearance and classification
The FDA regulates product code GFG (bit, surgical) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 03.820.153S
Has the NA 03.820.153S been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705034731827), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 03.820.153S?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 5, 2018. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 03.820.153S?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 03.820.153S FDA approved?
Devices of this type (product code GFG) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes (U.S.A.) LP.
Who makes the NA 03.820.153S, and is it still sold?
The FDA's device database lists Synthes (U.S.A.) LP as the company responsible for it (brand NA, model 03.820.153S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BIT, SURGICAL devices
- Tornier Perform MWJ127Tornier, Inc. · BIT, SURGICAL119 reports
- NA 310.534Synthes GmbH · BIT, SURGICAL64 reports
- NA 4200355042Stryker Corporation · BIT, SURGICAL34 reports
- ACUFEX 72204044Smith & Nephew, Inc. · BIT, SURGICAL25 reports
- NA 03.168.011Synthes GmbH · BIT, SURGICAL20 reports
- ACUFEX 72204042Smith & Nephew, Inc. · BIT, SURGICAL19 reports
- ACUFEX 72204046Smith & Nephew, Inc. · BIT, SURGICAL16 reports
- ACUFEX 72204045Smith & Nephew, Inc. · BIT, SURGICAL14 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.