NA 04.614.516Recalls & reported problems
Synthes GmbH · TI TOP LOADING TRANSCONNECTOR MEDIUM · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034739304), received between Mar 13, 2020 and Jun 22, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device Slipped” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Slipped1
- Device-Device Incompatibility1
- Material Integrity Problem1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 22, 2022Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018 THE PATIENT UNDERWENT THE PRIMARY PROCEDURE. TRANSVERSE CONNECTORS WERE IMPLANTED. ON (B)(6) 2022 A REMOVAL PROCEDURE WAS PERFORMED. WHEN EITHER 04.614.515S (LOT # H726297) OR 04.614.516S (LOT # H715939) WAS REMOVED DURING THE PROCEDURE, THE CONNECTOR AND A SETSCREW STUCK TOGETHER, AND THE SURGEON HAD DIFFICULTY IN DETACHING THEM. THEN, THE CONNECTING AREA ON THE CONNECTOR BROKE. A FRAGMENT WAS REMOVED FROM THE LESION, AND THE PROCEDURE WAS COMPLETED MORE THAN 30-MINUTE SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. CONCOMITANT DEVICE REPORTED: UNK - LOCKING/SET SCREWS (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1). THIS COMPLAINT INVOLVES ONE (1) DEVICE TI TOPLOADING TRANSCONNECTOR MED-STERILE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 14778060 ↗MalfunctionMar 13, 2020Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2020, THE SURGEON WAS PERFORMING C2-6 POSTERIOR CERVICAL FUSION PROCEDURE USING THE 4.0 SYNAPSE SYSTEM. NEAR THE END OF THE CASE, AFTER THE SURGEON APPLIED TWO TRANSCONNECTOR LOCKING SET SCREWS AND EIGHT (8) TITANIUM LOCKING SCREWS TO THE REMAINING SCREWS, THE DOCTOR THEN FINALLY TIGHTENED ALL TEN (10) SET SCREWS. THE PHYSICIAN ATTEMPTED TO SEAT A MEDIUM SYNAPSE TOP LOADING TRANSCONNECTOR TO THE TRANSCONNECTOR SET SCREWS. HE THEN APPLIED A 7.5MM TRANSCONNECTOR LOCKING NUT TO THE LEFT SIDE OF THE TRANSCONNECTOR BUT WHEN HE ATTEMPTED TO APPLY AND SEAT THE RIGHT SIDE THE TRANSCONNECTOR WOULD NOT SEAT. HE LOOSENED THE NUT ON THE LEFT SIDE TO SEE IF THAT WOULD HELP, BUT IT WOULD NOT SEAT. HE TRIED THIS A NUMBER OF TIMES, BUT IT CLEARLY WOULDN¿T SEAT PROPERLY. THEY USED MEDIUM ANGLED TOP LOADING TRANSCONNECTOR BUT HAD THE SAME RESULT. WHEN HE WAS REMOVING THE 7.5MM BUT IT WOULD NOT RELEASE AND HE SAID IT FELT AS IF THE HEAD OF THE PEDICLE SCREW SPLAYED OPEN. HE DECIDED TO REMOVE THE TRANSCONNECTOR NUTS, LOCKING SCREWS AND THE TRANSCONNECTOR AND SIMPLY USE JUST THE STANDARD TITANIUM LOCKING SCREWS. HE WANTS TO KNOW WHY THE TRANSCONNECTOR WOULD NOT SEAT AND I
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9832540 ↗
Other Orthosis, cervical pedicle screw spinal fixation devices
- INFINITY™ Occipitocervical Upper Thoracic System 3604026Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation2 recalls1 report
- CONNECTOR 79-2003Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 360PT4022Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- CONNECTOR 79-2002Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604532Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604024Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- OASYS 48551000Stryker Corporation · Orthosis, cervical pedicle screw spinal fixation45 reports
- YUKON OCT Spinal System 7601-05030VB Spine LLC · Orthosis, cervical pedicle screw spinal fixation15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page