Skip to content

NA 04.615.024Recalls & reported problems

Synthes GmbH · 3.5MM TI CANCELLOUS POLYAXIAL SCREW 24MM FOR 4.0MM RODS · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034739786), received between Oct 22, 2019 and Oct 22, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 22, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, WHILE REPLACING THE ROD, SURGEON WAS ATTEMPTING THE USE THE REDUCER IN ACCORDANCE WITH THE SURGICAL TECHNIQUE. THE ROD WAS APPROXIMATE 5MM PROUD OF THE TULIP HEAD AT C5 AND C6. UPON REDUCING THE ROD, THE TULIP HEAD DISENGAGED FROM THE SCREW SHANK AT C5. DR. (B)(6) THEN MOVED ON TO C6 WHERE THE SAME EVENT OCCURRED. PATIENT OUTCOME AND STATUS IS UNKNOWN. THIS COMPLAINT IS LINKED TO (B)(4). CONCOMITANT DEVICE REPORTED: UNKNOWN ROD (QUANTITY: 1). THIS COMPLAINT INVOLVES ONE (1) 3.5MM TI CANCELLOUS POLYAXIAL SCREW 24MM FOR 4.0MM ROD-STER. THIS IS 2 OF 2 FOR REPORT (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9219851 ↗

Other Orthosis, cervical pedicle screw spinal fixation devices

All Orthosis, cervical pedicle screw spinal fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page