Skip to content

Zero-P VA 04.647.227SRecalls & reported problems

Synthes GmbH · ZERO-P VA IMPLANT 7MM HEIGHT LORDOTIC/LARGE-STERILE · Class 2 · Implant

All models of Zero-P VA (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034751214), received between Apr 18, 2022 and Jan 7, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Compatibility Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K112068); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Compatibility Problem1
  • Detachment of Device or Device Component1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 7, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT ZERO-P VA CAGE LRG LORDOTIC H7 PEEK AND SCRDRIVER T8 SELF-HOLDING ANGL W/SLEEVE WERE INVOLVED IN AN INTRAOPERATIVE EVENT DURING A SPINAL FUSION PROCEDURE. DURING INSERTION OF THE INFERIOR SCREW, TWO ARTICULATING DRIVERS BROKE AND THE TITANIUM PORTION OF THE CAGE DETACHED. THE PIN AT THE ARTICULATING JOINT OF THE DRIVERS SHEARED OFF. THE CAGE WAS REMOVED AND REPLACED WITH A NEW ONE, AND A STRAIGHT SHAFT DRIVER WAS USED TO COMPLETE THE PROCEDURE. THE EVENT RESULTED IN A 10-MINUTE SURGICAL DELAY, BUT THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY ADVERSE PATIENT CONSEQUENCES. ALL DEVICE FRAGMENTS WERE REMOVED WITHOUT ADDITIONAL INTERVENTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23990033 ↗
  • MalfunctionApr 18, 2022Mechanical Problem; Compatibility Problem; Mechanical Problem; Compatibility Problem

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: THIS IS REPORT 1 OF 3 FOR (B)(4). IT WAS REPORTED BY THE SALES REP IN (B)(6) THAT DURING AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE FUSION CAGE DEVICE WAS UNSTABLE IN ITS ROTATION AND COULD NOT BE USED. ACCORDING TO THE REPORT, TWO SCREW DEVICES WERE WASTED AS A RESULT. ANOTHER LIKE DEVICES WERE USED TO COMPLETE THE PROCEDURE WITH A DELAY OF 10 MINUTES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) ZERO-P VA IMP 7MM HT LORDOTIC/LG-ST DEVICE. THIS COMPLAINT INVOLVES THREE (3) DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14137467 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112068. One submission often covers many models, so check the product and lot details in each record.

  • Z-0455-2014 · Sep 30, 2013 · Terminated
    Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is a stand alone anterior cervical interbody fusion device indica

Other Intervertebral fusion device with integrated fixation, cervical devices

All Intervertebral fusion device with integrated fixation, cervical devices →

Sources and updates