SynFrame 387.347Recalls & reported problems

Synthes GmbH · SYNFRAME RING CLAMP · Class 1

All models of SynFrame (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes SynFrame ring clamp has 131 adverse event reports on record, with no deaths and no injuries. The reports mostly describe the clamps that hold the blade holders slipping or losing grip on the frame, with some saying the clamps are worn and need replacing. Most say there was no impact on the patient. There are no recalls naming this device. No reports came in the last 12 months, compared with 6 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
131
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 131 adverse event reports that list this exact device (DI 10705034771137), received between Dec 12, 2016 and Jun 6, 2025 — classified as 131 malfunctions.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (35 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 8 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 6 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 6 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Operates Differently Than Expected35
  • Device-Device Incompatibility28
  • Mechanical Problem18
  • Failure To Adhere Or Bond15
  • Material Integrity Problem15
  • Compatibility Problem13
  • Unintended Movement9
  • Detachment of Device or Device Component6

Patient problems

  • No Consequences Or Impact To Patient58
  • No Known Impact Or Consequence To Patient36
  • No Clinical Signs, Symptoms or Conditions26
  • Insufficient Information14
  • No Patient Involvement7

Recent adverse event reports

The 25 most recent of 131 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 6, 2025Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT ALIF SURGERY ON (B)(6) 2025. THE SURGEON WAS USING THE SYNFRAME RETRACTOR, AND ALL THE CLAMPS THAT ATTACH THE BLADES HOLDERS TO THE RING WERE FAULTY AND KEPT SLIPPING. ALL ARE WORN ON THE BOTTOM LIP. THE CASE WAS ABLE TO BE COMPLETED WITH THE HOLDERS, HOWEVER THEY NEED TO BE REPLACED. THERE WAS A MINIMAL SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22158988 ↗
  • MalfunctionJun 6, 2025Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT ALIF SURGERY ON (B)(6) 2025. THE SURGEON WAS USING THE SYNFRAME RETRACTOR, AND ALL THE CLAMPS THAT ATTACH THE BLADES HOLDERS TO THE RING WERE FAULTY AND KEPT SLIPPING. ALL ARE WORN ON THE BOTTOM LIP. THE CASE WAS ABLE TO BE COMPLETED WITH THE HOLDERS; HOWEVER, THEY NEED TO BE REPLACED. THERE WAS A MINIMAL SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22158986 ↗
  • MalfunctionJun 6, 2025Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT ALIF SURGERY ON (B)(6) 2025. THE SURGEON WAS USING THE SYNFRAME RETRACTOR, AND ALL THE CLAMPS THAT ATTACH THE BLADES HOLDERS TO THE RING WERE FAULTY AND KEPT SLIPPING. ALL ARE WORN ON THE BOTTOM LIP. THE CASE WAS ABLE TO BE COMPLETED WITH THE HOLDERS, HOWEVER THEY NEED TO BE REPLACED. THERE WAS A MINIMAL SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22158985 ↗
  • MalfunctionJun 6, 2025Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT ALIF SURGERY ON (B)(6) 2025. THE SURGEON WAS USING THE SYNFRAME RETRACTOR, AND ALL THE CLAMPS THAT ATTACH THE BLADES HOLDERS TO THE RING WERE FAULTY AND KEPT SLIPPING. ALL ARE WORN ON THE BOTTOM LIP. THE CASE WAS ABLE TO BE COMPLETED WITH THE HOLDERS; HOWEVER THEY NEED TO BE REPLACED. THERE WAS A MINIMAL SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22158984 ↗
  • MalfunctionJun 6, 2025Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT ALIF SURGERY ON MAY 13, 2025. THE SURGEON WAS USING THE SYNFRAME RETRACTOR, AND ALL THE CLAMPS THAT ATTACH THE BLADES HOLDERS TO THE RING WERE FAULTY AND KEPT SLIPPING. ALL ARE WORN ON THE BOTTOM LIP. THE CASE WAS ABLE TO BE COMPLETED WITH THE HOLDERS, HOWEVER THEY NEED TO BE REPLACED. THERE WAS A MINIMAL SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22158983 ↗
  • MalfunctionJun 6, 2025Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT ALIF SURGERY ON (B)(6) 2025. THE SURGEON WAS USING THE SYNFRAME RETRACTOR, AND ALL THE CLAMPS THAT ATTACH THE BLADES HOLDERS TO THE RING WERE FAULTY AND KEPT SLIPPING. ALL ARE WORN ON THE BOTTOM LIP. THE CASE WAS ABLE TO BE COMPLETED WITH THE HOLDERS, HOWEVER THEY NEED TO BE REPLACED. THERE WAS A MINIMAL SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22158982 ↗
  • MalfunctionSep 5, 2024Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE TEAM IS HAVING AN ISSUE WITH THE SYNFRAME CLAMPS WHERE THE CLAMPS ARE LOSING THEIR GRIP STRENGTH ON THE FRAME. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE SYNFRAME RING CLAMP FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 20144427 ↗
  • MalfunctionSep 5, 2024Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE TEAM IS HAVING AN ISSUE WITH THE SYNFRAME CLAMPS WHERE THE CLAMPS ARE LOSING THEIR GRIP STRENGTH ON THE FRAME. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE SYNFRAME RING CLAMP FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 20144401 ↗
  • MalfunctionSep 5, 2024Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE TEAM IS HAVING AN ISSUE WITH THE SYNFRAME CLAMPS WHERE THE CLAMPS ARE LOSING THEIR GRIP STRENGTH ON THE FRAME. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE SYNFRAME RING CLAMP FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 20144358 ↗
  • MalfunctionSep 5, 2024Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE TEAM IS HAVING AN ISSUE WITH THE SYNFRAME CLAMPS WHERE THE CLAMPS ARE LOSING THEIR GRIP STRENGTH ON THE FRAME. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE SYNFRAME RING CLAMP FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 20144271 ↗
  • MalfunctionSep 5, 2024Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE TEAM IS HAVING AN ISSUE WITH THE SYNFRAME CLAMPS WHERE THE CLAMPS ARE LOSING THEIR GRIP STRENGTH ON THE FRAME. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE SYNFRAME RING CLAMP FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 20144067 ↗
  • MalfunctionSep 5, 2024Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE TEAM IS HAVING AN ISSUE WITH THE SYNFRAME CLAMPS WHERE THE CLAMPS ARE LOSING THEIR GRIP STRENGTH ON THE FRAME. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE WERE PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE SYNFRAME RING CLAMP FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 20144028 ↗
  • MalfunctionJan 30, 2024Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6), 2024, A SURGEON PERFORMED AN ALIF PROCEDURE. SYNFRAME RING CLAMPS WERE PART OF THE SYNFRAME THAT HOLD THE RETRACTOR BLADES. THE RING CLAMPS WERE SLIPPING AFTER TIGHTENING DOWN THE RETRACTOR BLADES IN SITU UNDER PRESSURE. WE DO NOT TIME THE SURGICAL CASES; THE AVERAGE TIME FOR AN ALIF IS 60 MINUTES, AND THE XASE TOOK 60 MINUTES. THE SURGEON HAD TO READJUST THE RETRACTOR BLADES INVOLVING THE RING CLAMPS, AND HAN-HELD RETRACTION WAS UTILIZED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WEREN'T ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT. THIS REPORT IS FOR ONE (1)SYNFRAME RING CLAMP THIS IS REPORT 8 OF 8 (B)(4).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18608573 ↗
  • MalfunctionJan 30, 2024Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON JANUARY 15, 2024, A SURGEON PERFORMED AN ALIF PROCEDURE. SYNFRAME RING CLAMPS WERE PART OF THE SYNFRAME THAT HOLD THE RETRACTOR BLADES. THE RING CLAMPS WERE SLIPPING AFTER TIGHTENING DOWN THE RETRACTOR BLADES IN SITU UNDER PRESSURE. WE DO NOT TIME THE SURGICAL CASES; THE AVERAGE TIME FOR AN ALIF IS 60 MINUTES, AND THE XASE TOOK 60 MINUTES. THE SURGEON HAD TO READJUST THE RETRACTOR BLADES INVOLVING THE RING CLAMPS, AND HAN-HELD RETRACTION WAS UTILIZED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WEREN'T ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT. THIS REPORT IS FOR ONE (1)SYNFRAME RING CLAMP. THIS IS REPORT 6 OF 8 (B)(4).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18608560 ↗
  • MalfunctionJan 30, 2024Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6), 2024, A SURGEON PERFORMED AN ALIF PROCEDURE. SYNFRAME RING CLAMPS WERE PART OF THE SYNFRAME THAT HOLD THE RETRACTOR BLADES. THE RING CLAMPS WERE SLIPPING AFTER TIGHTENING DOWN THE RETRACTOR BLADES IN SITU UNDER PRESSURE. WE DO NOT TIME THE SURGICAL CASES; THE AVERAGE TIME FOR AN ALIF IS 60 MINUTES, AND THE XASE TOOK 60 MINUTES. THE SURGEON HAD TO READJUST THE RETRACTOR BLADES INVOLVING THE RING CLAMPS, AND HAN-HELD RETRACTION WAS UTILIZED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WEREN'T ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT. THIS REPORT IS FOR ONE (1)SYNFRAME RING CLAMP THIS IS REPORT 7 OF 8 (B)(4).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18608559 ↗
  • MalfunctionJan 30, 2024Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6), 2024, A SURGEON PERFORMED AN ALIF PROCEDURE. SYNFRAME RING CLAMPS WERE PART OF THE SYNFRAME THAT HOLD THE RETRACTOR BLADES. THE RING CLAMPS WERE SLIPPING AFTER TIGHTENING DOWN THE RETRACTOR BLADES IN SITU UNDER PRESSURE. WE DO NOT TIME THE SURGICAL CASES; THE AVERAGE TIME FOR AN ALIF IS 60 MINUTES, AND THE XASE TOOK 60 MINUTES. THE SURGEON HAD TO READJUST THE RETRACTOR BLADES INVOLVING THE RING CLAMPS, AND HAN-HELD RETRACTION WAS UTILIZED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WEREN'T ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT. THIS REPORT IS FOR ONE (1)SYNFRAME RING CLAMP THIS IS REPORT 5 OF 8 (B)(4).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18608558 ↗
  • MalfunctionJan 30, 2024Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6) 2024, A SURGEON PERFORMED AN ALIF PROCEDURE. SYNFRAME RING CLAMPS WERE PART OF THE SYNFRAME THAT HOLD THE RETRACTOR BLADES. THE RING CLAMPS WERE SLIPPING AFTER TIGHTENING DOWN THE RETRACTOR BLADES IN SITU UNDER PRESSURE. WE DO NOT TIME THE SURGICAL CASES; THE AVERAGE TIME FOR AN ALIF IS 60 MINUTES, AND THE XASE TOOK 60 MINUTES. THE SURGEON HAD TO READJUST THE RETRACTOR BLADES INVOLVING THE RING CLAMPS, AND HAN-HELD RETRACTION WAS UTILIZED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WEREN'T ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT. THIS REPORT IS FOR ONE (1)SYNFRAME RING CLAMP. THIS IS REPORT 4 OF 8 (B)(4).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18608555 ↗
  • MalfunctionJan 30, 2024Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6) 2024, A SURGEON PERFORMED AN ALIF PROCEDURE. SYNFRAME RING CLAMPS WERE PART OF THE SYNFRAME THAT HOLD THE RETRACTOR BLADES. THE RING CLAMPS WERE SLIPPING AFTER TIGHTENING DOWN THE RETRACTOR BLADES IN SITU UNDER PRESSURE. WE DO NOT TIME THE SURGICAL CASES; THE AVERAGE TIME FOR AN ALIF IS 60 MINUTES, AND THE XASE TOOK 60 MINUTES. THE SURGEON HAD TO READJUST THE RETRACTOR BLADES INVOLVING THE RING CLAMPS, AND HAN-HELD RETRACTION WAS UTILIZED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WEREN'T ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT. THIS REPORT IS FOR ONE (1)SYNFRAME RING CLAMP. THIS IS REPORT 3 OF 8 (B)(4).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18608551 ↗
  • MalfunctionJan 30, 2024Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6) 2024, A SURGEON PERFORMED AN ALIF PROCEDURE. SYNFRAME RING CLAMPS WERE PART OF THE SYNFRAME THAT HOLD THE RETRACTOR BLADES. THE RING CLAMPS WERE SLIPPING AFTER TIGHTENING DOWN THE RETRACTOR BLADES IN SITU UNDER PRESSURE. WE DO NOT TIME THE SURGICAL CASES; THE AVERAGE TIME FOR AN ALIF IS 60 MINUTES, AND THE XASE TOOK 60 MINUTES. THE SURGEON HAD TO READJUST THE RETRACTOR BLADES INVOLVING THE RING CLAMPS, AND HAN-HELD RETRACTION WAS UTILIZED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WEREN'T ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT. THIS REPORT IS FOR ONE (1)SYNFRAME RING CLAMP. THIS IS REPORT 2 OF 8 (B)(4).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18608540 ↗
  • MalfunctionJan 30, 2024Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6) 2024, A SURGEON PERFORMED AN ALIF PROCEDURE. SYNFRAME RING CLAMPS WERE PART OF THE SYNFRAME THAT HOLD THE RETRACTOR BLADES. THE RING CLAMPS WERE SLIPPING AFTER TIGHTENING DOWN THE RETRACTOR BLADES IN SITU UNDER PRESSURE. WE DO NOT TIME THE SURGICAL CASES; THE AVERAGE TIME FOR AN ALIF IS 60 MINUTES, AND THE XASE TOOK 60 MINUTES. THE SURGEON HAD TO READJUST THE RETRACTOR BLADES INVOLVING THE RING CLAMPS, AND HAN-HELD RETRACTION WAS UTILIZED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WEREN'T ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT. THIS REPORT IS FOR ONE (1)SYNFRAME RING CLAMP. THIS IS REPORT 1 OF 8 (B)(4).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18608538 ↗
  • MalfunctionSep 11, 2023Compatibility Problem; Compatibility Problem

    THE PATIENT WAS REPORTED TO BE RECOVERING WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17719405 ↗
  • MalfunctionSep 11, 2023Compatibility Problem; Compatibility Problem

    THE PATIENT WAS REPORTED TO BE RECOVERING WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17719373 ↗
  • MalfunctionJul 25, 2023Mechanical Problem; Mechanical Problem

    DEVICE REPORT FROM SYNTHESE REPORTS AN EVENT IN THE UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, DURING A CASE THE TWO INSTRUMENTS WOULD NOT TIGHTEN. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH GOOD OUTCOME. THIS REPORT INVOLVES ONE (1) SYNFRAME RING CLAMP. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17390484 ↗
  • MalfunctionJul 25, 2023Mechanical Problem; Mechanical Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON (B)(4)2023, DURING A CASE THE TWO INSTRUMENTS WOULD NOT TIGHTEN. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH GOOD OUTCOME. THIS REPORT INVOLVES ONE (1) SYNFRAME RING CLAMP. THIS IS REPORT 1 OF 2 FOR (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17390440 ↗
  • MalfunctionJan 31, 2023Mechanical Problem; Mechanical Problem

    DEVICE REPORT FROM NORWAY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2023, THE LOCKING BUTTON ON THE SYNFRAME COULD BE PUSHED FROM SIDE TO SIDE, BUT THE TOOTH WHEELS COULD NOT BE MOVED SO THAT THE POSITION OF THE WOUND HOOKS COULD FASTEN ON. A SIMILAR PRODUCT WAS USED INSTEAD. PROCEDURE WAS COMPLETED WITH NO DELAY. THIS REPORT IS FOR A SYNFRAME RING CLAMP. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16274150 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →