ENDOPATH ATS45Recalls & reported problems

Ethicon Endo-Surgery, LLC · Endopath ETS-Flex45 Articulating Endoscopic Linear Cutter, No Reload · Class 2

All models of ENDOPATH (37)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Endopath endoscopic linear stapler is a surgical stapling tool used in minimally invasive surgery. Its 2,044 adverse event reports mostly describe the stapler failing to fire, being hard to open or close, and staples not forming properly, and some mention bleeding. The reports include 2 deaths and 109 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,044
Deaths reported
2
Injuries reported
110
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2,044 adverse event reports that list this exact device (DI 10705036001072), received between Jul 16, 2015 and Sep 21, 2026 — classified as 2 deaths, 110 injuries, 1,931 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (963 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 13 reportsJan 2024: 19 reportsFeb 2024: 29 reportsMar 2024: 17 reportsApr 2024: 16 reportsMay 2024: 10 reportsJun 2024: 13 reportsJul 2024: 16 reportsAug 2024: 7 reportsSep 2024: 30 reportsOct 2024: 15 reportsNov 2024: 13 reportsDec 2024: 15 reportsJan 2025: 10 reportsFeb 2025: 13 reportsMar 2025: 9 reportsApr 2025: 10 reportsMay 2025: 19 reportsJun 2025: 7 reportsJul 2025: 17 reportsAug 2025: 16 reportsSep 2025: 9 reportsOct 2025: 10 reportsNov 2025: 17 reportsDec 2025: 12 reportsJan 2026: 6 reportsFeb 2026: 10 reportsMar 2026: 9 reportsApr 2026: 9 reportsMay 2026: 8 reportsJun 2026: 5 reportsJul 2026: 1 reportsAug 2026: 4 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202630
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Fire963
  • Difficult to Open or Close359
  • Failure to Form Staple265
  • Firing Problem124
  • Failure to Cut118
  • Mechanical Jam105
  • Insufficient Device Problem Information101
  • Misfire87

Patient problems

  • No Clinical Signs, Symptoms or Conditions968
  • No Consequences Or Impact To Patient937
  • Hemorrhage/Blood Loss/Bleeding40
  • No Code Available37
  • Blood Loss25
  • Insufficient Information25
  • Failure to Anastomose15
  • Unspecified Tissue Injury12

Recent adverse event reports

The 25 most recent of 2,044 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Failure to Cut; Firing Problem

    It was reported that during an unknown procedure, the device attached the white cartridge and tried to fire, but the knife was too hard to move halfway through. The blue cartridge was used and the device could be fired. The cartridge colors were white and bule. Another cartridge was loaded into the same device to complete the case. There were no adverse consequences to the patient. Additional information requested.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50840830 ↗
  • MalfunctionSep 17, 2026Failure to Form Staple; No Apparent Adverse Event

    It was reported that during an unknown procedure, it was observed that the staple line did not close properly, causing bleeding in the intra-op, forcing the surgeon to perform cecal appendix raffia. The patient recovered from the procedure and is now well, with no further information. No additional information provided.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50804125 ↗
  • MalfunctionAug 17, 2026Difficult to Open or Close; Insufficient Device Problem Information

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50286745 ↗
  • MalfunctionAug 14, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50258426 ↗
  • MalfunctionAug 14, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50255749 ↗
  • InjuryAug 3, 2026Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    This complaint is from a literature source. The following literature cite has been reviewed:Tomotaka Shibata, Shinichi Sakuramoto, Hiroaki Mieno, Masayuki Nemoto, Nobue Futawatari, Keishi Yamashita, Natsuya Katada, Shirou Kikuchi; LAPAROSCOPY ASSISTED PROXIMAL GASTRECTOMY IN GASTRIC CANCER : SHORT TERM CLINICAL OUTCOME AND FUNCTIONAL STUDY; Surg Endosc (2012) 26:S249¿S430; (Pubmed cite not available).Laparoscopy assisted proximal gastrectomy (LAPG) has become prevalent for early gastric cancer (EGC) located in the upper stomach in Japan. However, standard reconstruction method is not defined. From May 2006 through June 2011, LAPG was performed for 50 EGC patients. Esophagogastrostomy with Liner stapler (LS) was performed in the initial 25 patients and the 25 remaining patients underwent that with circular stapler (CS). Postoperative esophago-gastro-reflux was assessed by 24 h pH monitoring (n = 25) and/or multichannel intraluminal impedance (MII/pH)(n = 11). Esophago-gastrostomy: Esophagus was laparoscopically dissected and both right and left crura of the diagram were exposed. Upper 3rd of the stomach was resected extra-corporeally with 5 cm minilaparotomy. Esophago-jejunostomy wa

    Patient problems: Failure to Anastomose

    MAUDE report 50115480 ↗
  • MalfunctionJul 27, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50037742 ↗
  • MalfunctionJun 19, 2026Failure to Fire; Material Integrity Problem

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED AND THE DEVICE MADE A STRANGE CLICKING SOUND. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25565771 ↗
  • MalfunctionJun 19, 2026Material Integrity Problem

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE RELEASE HANDLE ON THE DEVICE BROKE OFF AFTER THE DEVICE WAS SUCCESSFULLY TRIGGERED WITHIN THE ABDOMEN. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25565049 ↗
  • MalfunctionJun 8, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE DID NOT TRIGGER. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25428194 ↗
  • MalfunctionJun 2, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLER JAMMED DURING STAPLING. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25360679 ↗
  • MalfunctionJun 1, 2026Failure to Fire; Difficult to Open or Close

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE STAPLER HAS NOT FULLY FIRED, IT STOPPED IN THE MIDDLE OF THE FIRING CYCLE, THEN STOPPED GOING BACK AND FORTH. IT COULD THEN BE OPENED ONLY BY FORCE. THE STAPLES WERE NOT RELEASED IN THE ABDOMEN AND HAD TO BE REMOVED. A SECOND DEVICE HAD TO BE OPENED, WHICH THEN WORKED WITHOUT ANY PROBLEMS. NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25355686 ↗
  • MalfunctionMay 28, 2026Failure to Fire; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE STAPLER COULD BE CLOSED, BUT IT DID NOT FIRE PROPERLY. IT COULD BE ACTIVATED, BUT AFTER APPROXIMATELY 1 CM IT GOT STUCK. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? UNKNOWN, WAS PROCEDURE SUCCESSFULLY COMPLETED? UNKNOWN, WERE FRAGMENTS GENERATED? UNKNOWN, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? UNKNOWN, PATIENT STATUS/ OUTCOME / CONSEQUENCES YES, PATIENT CONSEQUENCE DESCRIPTION/WAS THERE A CLINICAL OUTCOME EXPERIENCED BY THE PATIENT (INFECTION, INFLAMMATION, ETC.)? UNKNOWN, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IF YES, DESCRIBE CUSTOMER REPORTED "PERFORATION AND BLEEDING".

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Insufficient Information

    MAUDE report 25316438 ↗
  • MalfunctionMay 14, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLER COULD NOT BE FULLY TRIGGERED. THE KNIFE ONLY ADVANCED A FEW MILLIMETERS AND THEN STOPPED. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25175612 ↗
  • MalfunctionMay 14, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE FIRED OUT. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25175122 ↗
  • MalfunctionMay 14, 2026Failure to Fire; Difficult to Open or Close; Noise, Audible

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE TRIGGER ON THE DEVICE DID NOT WORK CORRECTLY, THERE WAS A SOUND AS IF IT WOULD NOT CLOSE, AND THERE WERE PROBLEMS WITH THE KNIFE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25173438 ↗
  • MalfunctionMay 12, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE LOCKING OF THE DEVICE WENT WELL BUT THERE WERE NO STAPLES UPON SQUEEZING THE TRIGGER. THE DEVICE WAS BLOCKED. THERE WAS ALSO NO SUCCESS WITH ANOTHER MAGAZINE. A NEW STAPLER HAD TO BE OPENED AND THE ABLATION WAS SUCCESSFUL. HOWEVER, DUE TO THE CRUSHING AREAS, A COECUM WEDGE RESECTION INSTEAD OF A SIMPLE APPENDECTOMY HAD TO BE PERFORMED ON THE PATIENT. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25149713 ↗
  • MalfunctionMay 5, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE COULD NOT BE TRIGGERED. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25083394 ↗
  • MalfunctionMay 4, 2026Failure to Fire; Difficult to Open or Close; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDECTOMY PROCEDURE, IT WAS OBSERVED THAT THE STAPLER WAS MALFUNCTIONING AND THE JAWS WOULD NOT OPEN OR LOCKED-OUT. THEY COULDN'T PULL THE DEVICE FROM THE TISSUE SO THEY PULLED OUT THE DEVICE WITH THE TISSUE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury

    MAUDE report 25072233 ↗
  • MalfunctionMay 4, 2026Failure to Fire; Difficult to Open or Close

    IT WAS REPORTED THAT DURING A LAP APPENDECTOMY PROCEDURE, IT WAS OBSERVED THAT THEY HAD DIFFICULTIES OPENING STAPLER BACK UP TO RELEASE THE TISSUE AND SPECIMEN. SURGEON INITIALLY COMMENTED THE FIRING DID NOT STAPLE FULLY ACROSS THE TISSUE IT WAS CLOSED ON, THEN VERBALIZED THAT IT STAPLED ACROSS A METAL LIGACLIP. SURGEON REQUESTED ANOTHER STAPLE LOAD. ONCE SCRUB NURSE PROVIDED THE SAME, SURGEON ATTEMPTED TO CLOSE STAPLER BEFORE INSERTING IT INTO THE PORT SITE AND COMMENTED HOW THE STAPLER WAS NOT CLOSING PROPERLY AND REQUESTED A NEW STAPLER. RESIDENT AND SURGEON REPORTED NO ISSUES WITH THE SECOND STAPLER AND COMMENTED THAT IT BOTH "SOUNDED AND FELT BETTER". UPON DEBRIEF OF THE INCIDENT, RESIDENT AND SURGEON STATED THE FIRST STAPLER "FELT FUNKY" PRIOR TO USING IT AND NOTED THAT EVEN INITIALLY THEY HAD DIFFICULTIES OPENING THE STAPLER. THEY STATED THE FIRST STAPLER HAD DIFFICULTIES BOTH OPENING AND CLOSING. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25069116 ↗
  • MalfunctionApr 30, 2026Difficult to Open or Close; Noise, Audible

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE TRIGGER DID NOT WORK CORRECTLY AS THERE WAS A SOUND AND IT WOULD NOT CLOSE. THEY ALSO HAD PROBLEMS WITH THE KNIFE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25047708 ↗
  • MalfunctionApr 28, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, IT WAS OBSERVED THAT THE DEVICE LOCKED OUT. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25020752 ↗
  • MalfunctionApr 27, 2026Difficult to Open or Close; Firing Problem

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WAS DIFFICULT TO FIRE AND GOT STUCK ON THE TISSUE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25000468 ↗
  • MalfunctionApr 23, 2026Failure to Fire

    IT WAS REPORTED THAT PREOPERATIVELY TO AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WAS BLOCKED BEFORE FIRING. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24967944 ↗
  • MalfunctionApr 16, 2026Failure to Fire

    IT WAS REPORTED THAT PREOPERATIVELY TO AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WAS BLOCKED BEFORE FIRING. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24904864 ↗

FDA clearance and classification

The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the ENDOPATH ATS45

Has the ENDOPATH ATS45 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705036001072), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ENDOPATH ATS45?

The FDA's MAUDE database lists 2,044 adverse event reports for this device, received between Jul 16, 2015 and Sep 21, 2026. In those reports, the most common specific device problems are “Failure to Fire” (963), “Difficult to Open or Close” (359) and “Failure to Form Staple” (265); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (40), “Blood Loss” (25) and “Failure to Anastomose” (15). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ENDOPATH ATS45?

2 reports are classified as deaths, 110 as injuries and 1,931 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the ENDOPATH ATS45, and is it still sold?

The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ENDOPATH, model ATS45). Its record shows it as in commercial distribution.