ECHELON FLEX PSEE60ARecalls & reported problems
Ethicon Endo-Surgery, LLC · ECHELON FLEX Powered Plus Articulating Endoscopic Linear Cutter , 60mm · Class 2
All models of ECHELON FLEX (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Echelon Flex endoscopic stapler is a surgical device that cuts and staples tissue. Its 3,278 reports mostly describe the stapler failing to fire, being hard to open or close, and failing to form staples. The reports include 13 deaths and 282 injuries. There is 1 closed recall, begun in June 2021, for packages labeled 60mm that contained devices of a different size.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 3,279
- Deaths reported
- 13
- Injuries reported
- 283
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3,279 adverse event reports that list this exact device (DI 10705036014607), received between Oct 25, 2017 and Sep 28, 2026 — classified as 13 deaths, 283 injuries, 2,983 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (1,055 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jun 3, 2021Class IITerminated
Mislabeled: Labeled on the package as 60mm contains 45mm devices
Z-2326-2021 · Root cause: Packaging change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Fire1,055
- Difficult to Open or Close993
- Failure to Form Staple575
- Mechanical Jam403
- Activation Failure246
- Adverse Event Without Identified Device or Use Problem193
- Material Integrity Problem136
- No Apparent Adverse Event129
Patient problems
- No Clinical Signs, Symptoms or Conditions2,232
- No Consequences Or Impact To Patient747
- Hemorrhage/Blood Loss/Bleeding181
- Failure to Anastomose77
- No Code Available74
- Blood Loss40
- Insufficient Information32
- Unspecified Tissue Injury23
Recent adverse event reports
The 25 most recent of 3,279 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Failure to Form Staple
It was reported that during an unspecified surgical procedure, it was observed that the staples were malformed after firing. Changed to a new one to complete the procedure. There was no patient consequence nor surgical delay reported. No additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50916746 ↗InjurySep 25, 2026Patient Device Interaction Problem
It was reported that a patient who had previously undergone a Sleeve Gastrectomy was readmitted on (b)(6). No additional information was provided. The device will be returned for evaluation.
Patient problems: Insufficient Information
MAUDE report 50898806 ↗MalfunctionSep 21, 2026Failure to Form Staple; No Apparent Adverse Event
It was reported that during an unknown procedure, it was observed that a few staples were formed with irregular shapes after firing; and anastomotic site was oozing. Changed another one to continue the surgery, the same problem happened again. Used suture to oversew. There was no patient consequence reported. No additional information could be provided.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 50841412 ↗InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem
It was reported that during a Sleeve Gastrectomy procedure, the patient leaked which forced the patient to be reoperated. Additional information requested.
Patient problems: Failure to Anastomose
MAUDE report 50828441 ↗InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem
It was reported that during a Gastric Sleeve procedure, the patient presented leakage complications, which forced the patient to be reoperated the day after surgery. Additional information requested.
Patient problems: Failure to Anastomose
MAUDE report 50826349 ↗MalfunctionSep 18, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50820429 ↗MalfunctionSep 17, 2026Failure to Form Staple; Failure to Cut
Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury
MAUDE report 50808531 ↗MalfunctionSep 17, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50806252 ↗MalfunctionSep 1, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50541461 ↗MalfunctionAug 31, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50519691 ↗MalfunctionAug 25, 2026Ejection Problem
It was reported that during an unknown procedure, it was observed that a piece of the stapler load broke and got left behind in the stapler. New stapler was opened. No patient harm was reported. No additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50431399 ↗MalfunctionAug 21, 2026Failure to Fire; Difficult to Open or Close
It was reported that during an unknown procedure of the Pancreas, it was observed that the device could not be fired; and they opened the jaw manually with big force. Changed to a new one to complete the procedure. There was no patient consequence nor surgical delay reported. No additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50374134 ↗MalfunctionAug 20, 2026Failure to Fire; Difficult to Open or Close; Mechanical Jam; Activation Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50351390 ↗MalfunctionAug 19, 2026Material Integrity Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50327379 ↗MalfunctionAug 14, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50259944 ↗MalfunctionAug 14, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50255232 ↗MalfunctionAug 14, 2026Failure to Fire; Difficult to Open or Close
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50254571 ↗MalfunctionAug 13, 2026Material Integrity Problem
It was reported that during an unknown procedure, it was observed that small metal pieces popped out after firing the device. Changed the new one to complete surgery. There was no patient consequence reported. No additional information can be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50235253 ↗MalfunctionAug 10, 2026Failure to Fire
It was reported that during an unspecified surgical procedure, it was observed that the device would not fire. Changed to a new one to complete the procedure. There was no patient consequence nor surgical delay reported. No additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50195330 ↗MalfunctionAug 4, 2026Failure to Form Staple
It was reported that during a Lung Resection procedure, it was observed that the staples were malformed after firing. Changed to a new one to complete the procedure. There was no patient consequence nor surgical delay reported. No additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50129735 ↗MalfunctionAug 3, 2026Failure to Cut; No Apparent Adverse Event; Insufficient Device Problem Information
Patient problems: Unspecified Tissue Injury
MAUDE report 50114495 ↗MalfunctionJul 27, 2026Failure to Fire; Firing Problem
t was reported that during a Bariatric procedure of gastroplasty by sleeve technique, it was observed that the device presented difficulty to continue the stapling with slowness in the mechanism. They changed the reload but the device did not complete the stapling as if the stapler lost the strength to staple. They had to open a new device to complete the procedure successfully with a delay of 20 minutes. There was no patient consequence. Additional information requested.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50046745 ↗MalfunctionJul 27, 2026Failure to Fire; Difficult to Open or Close
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50046462 ↗MalfunctionJul 27, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50038613 ↗MalfunctionJul 27, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50038292 ↗
FDA clearance and classification
The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the ECHELON FLEX PSEE60A
Has the ECHELON FLEX PSEE60A been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-2326-2021 (Class II), began on Jun 3, 2021: Mislabeled: Labeled on the package as 60mm contains 45mm devices. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the ECHELON FLEX PSEE60A?
The FDA's MAUDE database lists 3,279 adverse event reports for this device, received between Oct 25, 2017 and Sep 28, 2026. In those reports, the most common specific device problems are “Failure to Fire” (1,055), “Difficult to Open or Close” (993) and “Failure to Form Staple” (575); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (181), “Failure to Anastomose” (77) and “Blood Loss” (40). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ECHELON FLEX PSEE60A?
13 reports are classified as deaths, 283 as injuries and 2,983 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ECHELON FLEX PSEE60A, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ECHELON FLEX, model PSEE60A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,660 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,727 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,350 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,006 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,967 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,910 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,044 reports
- ECHELON FLEX PVE35AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE29 deaths1,874 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page