ENDOPATH XCEL CB5LTRecalls & reported problems
Ethicon Endo-Surgery, LLC · ENDOPATH XCEL Universal Trocar Stablility Sleeves · Class 2
All models of ENDOPATH XCEL (25)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This is an Ethicon ENDOPATH XCEL trocar stability sleeve, a single-use port sleeve used in laparoscopic surgery. The 75 reports mostly describe packaging that arrives unsterile or damaged, plus sleeves that crack or break — including one that snapped in half while being tunneled under the skin and had to be removed by the surgical team. Reports mention 1 death and 4 injuries. No recalls name this device. Reports dropped sharply, from 42 in the prior 12 months to 5 in the last12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 75
- Deaths reported
- 1
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 75 adverse event reports that list this exact device (DI 10705036001348), received between Sep 19, 2013 and Mar 25, 2026 — classified as 1 death, 4 injuries, 70 malfunctions.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (50 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Delivered as Unsterile Product50
- Crack12
- Tear, Rip or Hole in Device Packaging5
- Break4
- Gas/Air Leak4
- Device Markings/Labelling Problem2
- Component or Accessory Incompatibility1
- Detachment Of Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions59
- No Consequences Or Impact To Patient9
- Insufficient Information5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 75 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2026Delivered as Unsterile Product; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE COSTUMER WAS COMPLAINING OF PACKAGING STERILITY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24691506 ↗MalfunctionMar 25, 2026Delivered as Unsterile Product; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE COSTUMER WAS COMPLAINING OF PACKAGING STERILITY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24691497 ↗MalfunctionMar 25, 2026Delivered as Unsterile Product; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE COSTUMER WAS COMPLAINING OF PACKAGING STERILITY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24691488 ↗MalfunctionMar 25, 2026Delivered as Unsterile Product; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE COSTUMER WAS COMPLAINING OF PACKAGING STERILITY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24691464 ↗MalfunctionMar 25, 2026Delivered as Unsterile Product; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE COSTUMER WAS COMPLAINING OF PACKAGING STERILITY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24691442 ↗MalfunctionOct 28, 2025Crack; Crack
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE A 5MM X 100MM UNIVERSAL TROCAR STABILITY SLEEVE (ENDOPATH XCEL REF (B)(4)) BROKE IN HALF WHILE THEY ARE TUNNELLING IT UNDER THE SKIN GOING TO THE NECK CAVITY. THE CONSULTANT SURGEON AND THE ASSISTANT SURGEON SUCCEED IN CAREFULLY REMOVING THE PART OF THE TROCAR THAT GOT STUCK UNDER THE SKIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23402153 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292453 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292449 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292447 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292416 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292344 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292335 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292332 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292268 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292266 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292264 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292262 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292257 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292254 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292093 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292084 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22289331 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22289330 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22289299 ↗MalfunctionJun 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE PACKAGING IS NOT STERILE. THERE IS A STERILITY ISSUE ON THE PACKAGING. THERE ARE BUMPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22289296 ↗
FDA clearance and classification
The FDA regulates product code GCJ (laparoscope, general & plastic surgery) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the ENDOPATH XCEL CB5LT
Has the ENDOPATH XCEL CB5LT been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036001348), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ENDOPATH XCEL CB5LT?
The FDA's MAUDE database lists 75 adverse event reports for this device, received between Sep 19, 2013 and Mar 25, 2026. In those reports, the most common specific device problems are “Delivered as Unsterile Product” (50), “Crack” (12) and “Tear, Rip or Hole in Device Packaging” (5). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ENDOPATH XCEL CB5LT?
1 report is classified as a death, 4 as injuries and 70 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ENDOPATH XCEL CB5LT, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ENDOPATH XCEL, model CB5LT). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,913 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,857 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,068 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall945 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery850 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery770 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall694 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery569 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.